AIMS: Ramp testing in the postoperative period can be used to optimize left ventricular assist device (LVAD) speed for optimal left ventricular (LV) unloading. We tested the hypothesis that a non‐invasive echocardiographic ramp test post‐HeartMate 3 implantation improves LV unloading immediately aft...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7160500/ https://www.ncbi.nlm.nih.gov/pubmed/32037731 http://dx.doi.org/10.1002/ehf2.12638 |
Sumario: | AIMS: Ramp testing in the postoperative period can be used to optimize left ventricular assist device (LVAD) speed for optimal left ventricular (LV) unloading. We tested the hypothesis that a non‐invasive echocardiographic ramp test post‐HeartMate 3 implantation improves LV unloading immediately after and 1–3 months after as compared with before the test. We also tested a secondary hypothesis that speed adjustments during echocardiography‐guided ramp testing do not worsen right ventricular (RV) function immediately after and 1–3 months after. METHODS AND RESULTS: We retrospectively reviewed data from patients who underwent an echocardiographic ramp test. A total of 14 out of 19 patients were clinically stable and were enrolled. Adequate LV unloading was defined as no more than mild mitral regurgitation, and intermittent aortic valve (AV) opening or closed AV, and reduction of left ventricular end‐diastolic diameter (LVEDD); and for the follow‐up measurement, decreased NT‐proBNP. Median (interquartile range) time from implantation to ramp test was 27 (16; 56) days, and median time from ramp test to follow‐up echocardiography was 55 (47; 102) days. Median LVAD speed achieved during ramp testing was 5550 (5375; 6025) revolutions per minute (rpm), and median final LVAD speed was 5200 (5000; 5425) rpm. Ramp testing resulted in final LVAD speed increase in 11 (79%) patients and a median net change of 200 (200; 300) rpm. Speed adjustments after ramp testing resulted in improved LVAD unloading that was achieved in additional 3 (21%) patients who were not originally optimized. RV function did not worsen significantly during ramp testing or at final LVAD speed. CONCLUSIONS: The echocardiographic ramp test allowed LVAD speed adjustment and optimization and improved LV unloading during ramp testing and at final speed with no evidence of worsening of RV function. |
---|