An accelerometer-based navigation system provides acetabular cup orientation accuracy comparable to that of computed tomography-based navigation during total hip arthroplasty in the supine position
BACKGROUND: Inadequate acetabular component orientation is associated with postoperative impingement, dislocation, and accelerated polyethylene wear. Computed tomography (CT)-based navigation systems provide accuracy for total hip arthroplasty (THA) but are not available in all facilities. Accelerom...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7161129/ https://www.ncbi.nlm.nih.gov/pubmed/32295628 http://dx.doi.org/10.1186/s13018-020-01673-y |
Sumario: | BACKGROUND: Inadequate acetabular component orientation is associated with postoperative impingement, dislocation, and accelerated polyethylene wear. Computed tomography (CT)-based navigation systems provide accuracy for total hip arthroplasty (THA) but are not available in all facilities. Accelerometer-based navigation systems are inexpensive, but their accuracy remains undetermined. This study compares the accuracy of cup orientation in THA using CT-based and accelerometer-based navigation systems. METHODS: This retrospective study included 35 consecutive patients (11 males, 24 females; mean age, 65 years) who underwent primary cementless THA via an anterolateral approach in the supine position. Both CT-based and accelerometer-based navigation systems were used simultaneously. The accuracy of cup orientation was compared between the two systems using postoperative CT. RESULTS: The accuracy of cup inclination was 2.7° ± 2.0° in the CT-based group and 3.3° ± 2.4° in the accelerometer-based group. The accuracy of cup anteversion was 2.8° ± 2.6° in the CT-based group and 3.4° ± 2.2° in the accelerometer-based group. No significant difference was observed in cup inclination (p = 0.29) or cup anteversion (p = 0.34) between CT-based and accelerometer-based navigation. CONCLUSIONS: The accuracy of cup positioning did not differ significantly between CT-based and accelerometer-based navigation systems. |
---|