Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
PURPOSE: To investigate the set‐up error and consequent dosimetric change in HexaPOD evo RT 6D couch under image‐guided intensity‐modulated radiotherapy (IG‐IMRT) for primary malignant tumor of the cervical spine. METHODS: Ten cases with primary malignant tumor of the cervical spine were treated wit...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7170283/ https://www.ncbi.nlm.nih.gov/pubmed/32170991 http://dx.doi.org/10.1002/acm2.12840 |
_version_ | 1783523862787391488 |
---|---|
author | Jiang, Ping Zhang, Xile Wei, Shuhua Zhao, Tiandi Wang, Junjie |
author_facet | Jiang, Ping Zhang, Xile Wei, Shuhua Zhao, Tiandi Wang, Junjie |
author_sort | Jiang, Ping |
collection | PubMed |
description | PURPOSE: To investigate the set‐up error and consequent dosimetric change in HexaPOD evo RT 6D couch under image‐guided intensity‐modulated radiotherapy (IG‐IMRT) for primary malignant tumor of the cervical spine. METHODS: Ten cases with primary malignant tumor of the cervical spine were treated with intensity‐modulated radiotherapy (IMRT) in our hospital from August 2013 to November 2014. The X‐ray volumetric images (XVI) were scanned and obtained by cone‐beam CT (CBCT). The six directions (6D) of set‐up errors of translation and rotation were obtained by planned CT image registration. HexaPOD evo RT 6D couch made online correction of the set‐up error, and then the CBCT was conducted to obtain the residual error. RESULTS: We performed set‐up error and dosimetric analysis. First, for the set‐up error analysis, the average error in three translation directions of 6D set‐up error of the primary tumor of the cervical spine was <2 mm, whereas the single maximum error (absolute value) is 7.0 mm. Among average errors of rotation direction, Rotation X (RX) direction 0.67° ± 0.04°, Rotation Y (RY) direction 1.06° ± 0.06°, Rotation Z (RZ) direction 0.78° ± 0.05°; and the single maximum error in three rotation directions were 2.8°, 3.8°, and 2.9°, respectively. On three directions (X, Y, Z axis), the extended distance from clinical target volume (CTV) to planning target volume (PTV) was 3.45, 3.17, and 3.90 mm by calculating, respectively. Then, for the dosimetric analysis, the parameters, including plan sum PTV D98 and D95, planning gross tumor volume D98 and D95, V100% of the plan sum were significantly lower than the treatment plan. Moreover, Dmax of the spinal cord was significantly higher than the treatment plan. CONCLUSION: 6D set‐up error correction system should be used for accurate position calibration of precise radiotherapy for patients with malignant tumor of the cervical spine. |
format | Online Article Text |
id | pubmed-7170283 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-71702832020-04-21 Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine Jiang, Ping Zhang, Xile Wei, Shuhua Zhao, Tiandi Wang, Junjie J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: To investigate the set‐up error and consequent dosimetric change in HexaPOD evo RT 6D couch under image‐guided intensity‐modulated radiotherapy (IG‐IMRT) for primary malignant tumor of the cervical spine. METHODS: Ten cases with primary malignant tumor of the cervical spine were treated with intensity‐modulated radiotherapy (IMRT) in our hospital from August 2013 to November 2014. The X‐ray volumetric images (XVI) were scanned and obtained by cone‐beam CT (CBCT). The six directions (6D) of set‐up errors of translation and rotation were obtained by planned CT image registration. HexaPOD evo RT 6D couch made online correction of the set‐up error, and then the CBCT was conducted to obtain the residual error. RESULTS: We performed set‐up error and dosimetric analysis. First, for the set‐up error analysis, the average error in three translation directions of 6D set‐up error of the primary tumor of the cervical spine was <2 mm, whereas the single maximum error (absolute value) is 7.0 mm. Among average errors of rotation direction, Rotation X (RX) direction 0.67° ± 0.04°, Rotation Y (RY) direction 1.06° ± 0.06°, Rotation Z (RZ) direction 0.78° ± 0.05°; and the single maximum error in three rotation directions were 2.8°, 3.8°, and 2.9°, respectively. On three directions (X, Y, Z axis), the extended distance from clinical target volume (CTV) to planning target volume (PTV) was 3.45, 3.17, and 3.90 mm by calculating, respectively. Then, for the dosimetric analysis, the parameters, including plan sum PTV D98 and D95, planning gross tumor volume D98 and D95, V100% of the plan sum were significantly lower than the treatment plan. Moreover, Dmax of the spinal cord was significantly higher than the treatment plan. CONCLUSION: 6D set‐up error correction system should be used for accurate position calibration of precise radiotherapy for patients with malignant tumor of the cervical spine. John Wiley and Sons Inc. 2020-03-14 /pmc/articles/PMC7170283/ /pubmed/32170991 http://dx.doi.org/10.1002/acm2.12840 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Jiang, Ping Zhang, Xile Wei, Shuhua Zhao, Tiandi Wang, Junjie Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
title | Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
title_full | Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
title_fullStr | Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
title_full_unstemmed | Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
title_short | Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
title_sort | set‐up error and dosimetric analysis of hexapod evo rt 6d couch combined with cone beam ct image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7170283/ https://www.ncbi.nlm.nih.gov/pubmed/32170991 http://dx.doi.org/10.1002/acm2.12840 |
work_keys_str_mv | AT jiangping setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine AT zhangxile setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine AT weishuhua setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine AT zhaotiandi setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine AT wangjunjie setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine |