Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine

PURPOSE: To investigate the set‐up error and consequent dosimetric change in HexaPOD evo RT 6D couch under image‐guided intensity‐modulated radiotherapy (IG‐IMRT) for primary malignant tumor of the cervical spine. METHODS: Ten cases with primary malignant tumor of the cervical spine were treated wit...

Descripción completa

Detalles Bibliográficos
Autores principales: Jiang, Ping, Zhang, Xile, Wei, Shuhua, Zhao, Tiandi, Wang, Junjie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7170283/
https://www.ncbi.nlm.nih.gov/pubmed/32170991
http://dx.doi.org/10.1002/acm2.12840
_version_ 1783523862787391488
author Jiang, Ping
Zhang, Xile
Wei, Shuhua
Zhao, Tiandi
Wang, Junjie
author_facet Jiang, Ping
Zhang, Xile
Wei, Shuhua
Zhao, Tiandi
Wang, Junjie
author_sort Jiang, Ping
collection PubMed
description PURPOSE: To investigate the set‐up error and consequent dosimetric change in HexaPOD evo RT 6D couch under image‐guided intensity‐modulated radiotherapy (IG‐IMRT) for primary malignant tumor of the cervical spine. METHODS: Ten cases with primary malignant tumor of the cervical spine were treated with intensity‐modulated radiotherapy (IMRT) in our hospital from August 2013 to November 2014. The X‐ray volumetric images (XVI) were scanned and obtained by cone‐beam CT (CBCT). The six directions (6D) of set‐up errors of translation and rotation were obtained by planned CT image registration. HexaPOD evo RT 6D couch made online correction of the set‐up error, and then the CBCT was conducted to obtain the residual error. RESULTS: We performed set‐up error and dosimetric analysis. First, for the set‐up error analysis, the average error in three translation directions of 6D set‐up error of the primary tumor of the cervical spine was <2 mm, whereas the single maximum error (absolute value) is 7.0 mm. Among average errors of rotation direction, Rotation X (RX) direction 0.67° ± 0.04°, Rotation Y (RY) direction 1.06° ± 0.06°, Rotation Z (RZ) direction 0.78° ± 0.05°; and the single maximum error in three rotation directions were 2.8°, 3.8°, and 2.9°, respectively. On three directions (X, Y, Z axis), the extended distance from clinical target volume (CTV) to planning target volume (PTV) was 3.45, 3.17, and 3.90 mm by calculating, respectively. Then, for the dosimetric analysis, the parameters, including plan sum PTV D98 and D95, planning gross tumor volume D98 and D95, V100% of the plan sum were significantly lower than the treatment plan. Moreover, Dmax of the spinal cord was significantly higher than the treatment plan. CONCLUSION: 6D set‐up error correction system should be used for accurate position calibration of precise radiotherapy for patients with malignant tumor of the cervical spine.
format Online
Article
Text
id pubmed-7170283
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-71702832020-04-21 Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine Jiang, Ping Zhang, Xile Wei, Shuhua Zhao, Tiandi Wang, Junjie J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: To investigate the set‐up error and consequent dosimetric change in HexaPOD evo RT 6D couch under image‐guided intensity‐modulated radiotherapy (IG‐IMRT) for primary malignant tumor of the cervical spine. METHODS: Ten cases with primary malignant tumor of the cervical spine were treated with intensity‐modulated radiotherapy (IMRT) in our hospital from August 2013 to November 2014. The X‐ray volumetric images (XVI) were scanned and obtained by cone‐beam CT (CBCT). The six directions (6D) of set‐up errors of translation and rotation were obtained by planned CT image registration. HexaPOD evo RT 6D couch made online correction of the set‐up error, and then the CBCT was conducted to obtain the residual error. RESULTS: We performed set‐up error and dosimetric analysis. First, for the set‐up error analysis, the average error in three translation directions of 6D set‐up error of the primary tumor of the cervical spine was <2 mm, whereas the single maximum error (absolute value) is 7.0 mm. Among average errors of rotation direction, Rotation X (RX) direction 0.67° ± 0.04°, Rotation Y (RY) direction 1.06° ± 0.06°, Rotation Z (RZ) direction 0.78° ± 0.05°; and the single maximum error in three rotation directions were 2.8°, 3.8°, and 2.9°, respectively. On three directions (X, Y, Z axis), the extended distance from clinical target volume (CTV) to planning target volume (PTV) was 3.45, 3.17, and 3.90 mm by calculating, respectively. Then, for the dosimetric analysis, the parameters, including plan sum PTV D98 and D95, planning gross tumor volume D98 and D95, V100% of the plan sum were significantly lower than the treatment plan. Moreover, Dmax of the spinal cord was significantly higher than the treatment plan. CONCLUSION: 6D set‐up error correction system should be used for accurate position calibration of precise radiotherapy for patients with malignant tumor of the cervical spine. John Wiley and Sons Inc. 2020-03-14 /pmc/articles/PMC7170283/ /pubmed/32170991 http://dx.doi.org/10.1002/acm2.12840 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Jiang, Ping
Zhang, Xile
Wei, Shuhua
Zhao, Tiandi
Wang, Junjie
Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
title Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
title_full Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
title_fullStr Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
title_full_unstemmed Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
title_short Set‐up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
title_sort set‐up error and dosimetric analysis of hexapod evo rt 6d couch combined with cone beam ct image‐guided intensity‐modulated radiotherapy for primary malignant tumor of the cervical spine
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7170283/
https://www.ncbi.nlm.nih.gov/pubmed/32170991
http://dx.doi.org/10.1002/acm2.12840
work_keys_str_mv AT jiangping setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine
AT zhangxile setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine
AT weishuhua setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine
AT zhaotiandi setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine
AT wangjunjie setuperroranddosimetricanalysisofhexapodevort6dcouchcombinedwithconebeamctimageguidedintensitymodulatedradiotherapyforprimarymalignanttumorofthecervicalspine