Retrieval Analysis of Modern Knee Tumor Megaendoprosthesis Shows Considerable Volumetric Metal Wear Generated at the Rotating Hinge
Frequently occurring damage, as well as elevated blood metal ion levels, are reported in relation to a tumor and revision system for total knee arthroplasty (TKA), which applies a rotating hinge coupling with a metal-on-metal (MoM) articulation. As the patient collective for this specific system is...
Autores principales: | Bormann, Therese, Jäger, Sebastian, Kretzer, J. Philippe, Nebel, Laura, Clarius, Lucas, Omlor, Georg, Bitsch, Rudi, Lehner, Burkhard |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7178100/ https://www.ncbi.nlm.nih.gov/pubmed/32224971 http://dx.doi.org/10.3390/ma13071519 |
Ejemplares similares
-
Structural Failure of a Modern Knee Tumor Megaendoprosthesis
por: Koch, Katrin, et al.
Publicado: (2017) -
Hip Spacers with a Metal-on-Cement Articulation Did Not Show Significant Surface Alterations of the Metal Femoral Head in Two-Stage Revision for Periprosthetic Joint Infection
por: Lunz, Andre, et al.
Publicado: (2020) -
Failure of Polyethylene Inlays in Cementless Total Hip Arthroplasty: A Retrieval Analysis
por: Mueller, Ulrike, et al.
Publicado: (2016) -
Analysis of Carbon Fiber Reinforced PEEK Hinge Mechanism Articulation Components in a Rotating Hinge Knee Design: A Comparison of In Vitro and Retrieval Findings
por: Schierjott, Ronja A., et al.
Publicado: (2016) -
Posterior Dislocation of the Hinge-Post Extension in a Rotating Hinge Total Knee Prosthesis
por: Manzano, Givenchy, et al.