Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements
Open-ended coaxial probes are widely used to gather dielectric properties of biological tissues. Due to the lack of an agreed data acquisition protocol, several environmental conditions can cause inaccuracies when comparing dielectric data. In this work, the effect of a different measurement probe-t...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7181017/ https://www.ncbi.nlm.nih.gov/pubmed/32268598 http://dx.doi.org/10.3390/s20072060 |
_version_ | 1783525955416883200 |
---|---|
author | Maenhout, Gertjan Markovic, Tomislav Ocket, Ilja Nauwelaers, Bart |
author_facet | Maenhout, Gertjan Markovic, Tomislav Ocket, Ilja Nauwelaers, Bart |
author_sort | Maenhout, Gertjan |
collection | PubMed |
description | Open-ended coaxial probes are widely used to gather dielectric properties of biological tissues. Due to the lack of an agreed data acquisition protocol, several environmental conditions can cause inaccuracies when comparing dielectric data. In this work, the effect of a different measurement probe-to-tissue contact pressure was monitored in the frequency range from 0.5 to 20 GHz. Therefore, we constructed a controlled lifting platform with an integrated pressure sensor to exert a constant pressure on the tissue sample during the dielectric measurement. In the pressure range from 7.74 kPa to 77.4 kPa, we observed a linear correlation of [Formula: see text] % and [Formula: see text] % per kPa for, respectively, the relative real and imaginary complex permittivity. These values are statistically significant compared with the reported measurement uncertainty. Following the literature in different biology-related disciplines regarding pressure-induced variability in measurements, we hypothesize that these changes originate from squeezing out the interstitial and extracellular fluid. This process locally increases the concentration of membranes, cellular organelles, and proteins in the sensed volume. Finally, we suggest moving towards a standardized probe-to-tissue contact pressure, since the literature has already demonstrated that reprobing at the same pressure can produce repeatable data within a 1% uncertainty interval. |
format | Online Article Text |
id | pubmed-7181017 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-71810172020-04-30 Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements Maenhout, Gertjan Markovic, Tomislav Ocket, Ilja Nauwelaers, Bart Sensors (Basel) Article Open-ended coaxial probes are widely used to gather dielectric properties of biological tissues. Due to the lack of an agreed data acquisition protocol, several environmental conditions can cause inaccuracies when comparing dielectric data. In this work, the effect of a different measurement probe-to-tissue contact pressure was monitored in the frequency range from 0.5 to 20 GHz. Therefore, we constructed a controlled lifting platform with an integrated pressure sensor to exert a constant pressure on the tissue sample during the dielectric measurement. In the pressure range from 7.74 kPa to 77.4 kPa, we observed a linear correlation of [Formula: see text] % and [Formula: see text] % per kPa for, respectively, the relative real and imaginary complex permittivity. These values are statistically significant compared with the reported measurement uncertainty. Following the literature in different biology-related disciplines regarding pressure-induced variability in measurements, we hypothesize that these changes originate from squeezing out the interstitial and extracellular fluid. This process locally increases the concentration of membranes, cellular organelles, and proteins in the sensed volume. Finally, we suggest moving towards a standardized probe-to-tissue contact pressure, since the literature has already demonstrated that reprobing at the same pressure can produce repeatable data within a 1% uncertainty interval. MDPI 2020-04-06 /pmc/articles/PMC7181017/ /pubmed/32268598 http://dx.doi.org/10.3390/s20072060 Text en © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Maenhout, Gertjan Markovic, Tomislav Ocket, Ilja Nauwelaers, Bart Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements |
title | Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements |
title_full | Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements |
title_fullStr | Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements |
title_full_unstemmed | Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements |
title_short | Effect of Open-Ended Coaxial Probe-to-Tissue Contact Pressure on Dielectric Measurements |
title_sort | effect of open-ended coaxial probe-to-tissue contact pressure on dielectric measurements |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7181017/ https://www.ncbi.nlm.nih.gov/pubmed/32268598 http://dx.doi.org/10.3390/s20072060 |
work_keys_str_mv | AT maenhoutgertjan effectofopenendedcoaxialprobetotissuecontactpressureondielectricmeasurements AT markovictomislav effectofopenendedcoaxialprobetotissuecontactpressureondielectricmeasurements AT ocketilja effectofopenendedcoaxialprobetotissuecontactpressureondielectricmeasurements AT nauwelaersbart effectofopenendedcoaxialprobetotissuecontactpressureondielectricmeasurements |