An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator?
We describe the case of a patient who received cardiac resynchronization pacemaker therapy (CRT-P) incorporating bridge protection via a wearable cardioverter-defibrillator (WCD) while awaiting final evaluation for permanent placement of an implantable CRT device with defibrillation capabilities. In...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MediaSphere Medical
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7192131/ https://www.ncbi.nlm.nih.gov/pubmed/32368370 http://dx.doi.org/10.19102/icrm.2020.110201 |
_version_ | 1783527968375570432 |
---|---|
author | Tobb, Kardie Corbisiero, Raffaele Fish, Frank Kazemian, Pedram Muller, David |
author_facet | Tobb, Kardie Corbisiero, Raffaele Fish, Frank Kazemian, Pedram Muller, David |
author_sort | Tobb, Kardie |
collection | PubMed |
description | We describe the case of a patient who received cardiac resynchronization pacemaker therapy (CRT-P) incorporating bridge protection via a wearable cardioverter-defibrillator (WCD) while awaiting final evaluation for permanent placement of an implantable CRT device with defibrillation capabilities. In an attempt to mitigate symptoms of heart failure, reprogramming of the CRT-P with a V–V offset caused an unusual interaction and alarm from the WCD. This case highlights a unique interaction not previously reported and is clinically relevant to patients awaiting final evaluation for implantable defibrillator system placement. |
format | Online Article Text |
id | pubmed-7192131 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | MediaSphere Medical |
record_format | MEDLINE/PubMed |
spelling | pubmed-71921312020-05-04 An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? Tobb, Kardie Corbisiero, Raffaele Fish, Frank Kazemian, Pedram Muller, David J Innov Card Rhythm Manag Case Report We describe the case of a patient who received cardiac resynchronization pacemaker therapy (CRT-P) incorporating bridge protection via a wearable cardioverter-defibrillator (WCD) while awaiting final evaluation for permanent placement of an implantable CRT device with defibrillation capabilities. In an attempt to mitigate symptoms of heart failure, reprogramming of the CRT-P with a V–V offset caused an unusual interaction and alarm from the WCD. This case highlights a unique interaction not previously reported and is clinically relevant to patients awaiting final evaluation for implantable defibrillator system placement. MediaSphere Medical 2020-02-15 /pmc/articles/PMC7192131/ /pubmed/32368370 http://dx.doi.org/10.19102/icrm.2020.110201 Text en Copyright: © 2020 Innovations in Cardiac Rhythm Management http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Tobb, Kardie Corbisiero, Raffaele Fish, Frank Kazemian, Pedram Muller, David An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? |
title | An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? |
title_full | An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? |
title_fullStr | An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? |
title_full_unstemmed | An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? |
title_short | An Alarming Change: What Caused an Instance of Inappropriate Detection in a Wearable Cardioverter-defibrillator? |
title_sort | alarming change: what caused an instance of inappropriate detection in a wearable cardioverter-defibrillator? |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7192131/ https://www.ncbi.nlm.nih.gov/pubmed/32368370 http://dx.doi.org/10.19102/icrm.2020.110201 |
work_keys_str_mv | AT tobbkardie analarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT corbisieroraffaele analarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT fishfrank analarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT kazemianpedram analarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT mullerdavid analarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT tobbkardie alarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT corbisieroraffaele alarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT fishfrank alarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT kazemianpedram alarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator AT mullerdavid alarmingchangewhatcausedaninstanceofinappropriatedetectioninawearablecardioverterdefibrillator |