Determination of paramagnetic ferrous gel sensitivity in low energy x-ray beam produced by a miniature accelerator
The INTRABEAM Carl Zeiss Surgical system (Oberkochen, Germany) is a miniature accelerator producing low energy photons (50 keV maximum). The published dosimetric characterization of the INTRABEAM was based on detectors (radiochromic films or ionization chambers) not allowing measuring the absorbed d...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7197784/ https://www.ncbi.nlm.nih.gov/pubmed/32365071 http://dx.doi.org/10.1371/journal.pone.0232315 |
Sumario: | The INTRABEAM Carl Zeiss Surgical system (Oberkochen, Germany) is a miniature accelerator producing low energy photons (50 keV maximum). The published dosimetric characterization of the INTRABEAM was based on detectors (radiochromic films or ionization chambers) not allowing measuring the absorbed dose in the first millimeters of the irradiated medium, where the dose is actually prescribed. This study aims at determining with Magnetic Resonance Imaging (MRI) the sensitivity of a paramagnetic gel in order to measure the dose deposit produced with the INTRABEAM from 0 to 20 mm. Although spherical applicators are mostly used with the INTRABEAM system for breast applications, this study focuses on surface applicators that are of interest for cutaneous carcinomas. The irradiations at 12 different dose levels (between 2 Gy and 50 Gy at the gel surface) were performed with the INTRABEAM and a 4 cm surface applicator. The gel used in this study is a new « sensitive » material. In order to compare gel sensitivity at low energy with high energy, gels were irradiated by an 18 MV photon beam produced by a Varian Clinac 2100 CD. T(2) weighted multi echo MRI sequences were performed with 16 echo times. The T(2) signal versus echo times was fitted with a mono-exponential function with 95% confidence interval. The calibration curve determined at low energy is a linear function (r(2) = 0.9893) with a sensitivity of 0.0381 s(-1).Gy(-1), a similar linear function was obtained at high energy (0.0372 s(-1).Gy(-1) with r(2) = 0.9662). The calibration curve at low energy was used to draw isodose maps from the MR images. The PDD (Percent Depth Dose) determined in the gel is within 5%-1mm of the ionization chamber PDD except for one point. The dosimetric sensitivity of this new paramagnetic ferrous gel was determined with MRI measurements. It allowed measuring the dose distribution specifically in the first millimeters for an irradiation with the INTRABEAM miniature accelerator equipped with a surface applicator. |
---|