Dosimetric Limitations in Treating Breast Cancer with Accelerated Partial Breast Irradiation Using Strut Adjusted Volume Implant (SAVI)
We present one case of accelerated partial breast irradiation (APBI) using strut adjusted volume implant (SAVI) where there were limitations in delivering the dose as per the standard guidelines. The device was placed close to both the chest wall and the skin with little tissue surrounding the tip....
Autores principales: | Youssef, Ashraf, Mahadevan, Anand, Philippides, Adele, Thieme, Heather, Soto Hamlin, Angela |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cureus
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7198095/ https://www.ncbi.nlm.nih.gov/pubmed/32377476 http://dx.doi.org/10.7759/cureus.7528 |
Ejemplares similares
-
Comparison of planning techniques when air/fluid is present using the strut‐adjusted volume implant (SAVI) for HDR‐based accelerated partial breast irradiation
por: Harmon, Joseph F., et al.
Publicado: (2013) -
Inter-fractional variations in the dosimetric parameters of accelerated partial breast irradiation using a strut-adjusted volume implant
por: Iijima, Kotaro, et al.
Publicado: (2019) -
Dosimetry evaluation of SAVI-based HDR brachytherapy for partial breast irradiation
por: Manoharan, Sivasubramanian R., et al.
Publicado: (2010) -
Dosimetric effects of air pocket sizes in MammoSite treatment as accelerated partial breast irradiation for early breast cancer
por: Huang, Y. Jessica, et al.
Publicado: (2009) -
Dosimetric characteristics of accelerated partial breast irradiation by interstitial multicatheter brachytherapy with intraoperative free‐hand implantation in the treatment of early breast cancer
por: Li, Chuan, et al.
Publicado: (2021)