Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head

OBJECTIVE: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. METHODS: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator,...

Descripción completa

Detalles Bibliográficos
Autores principales: Guimarães, Daniel Peres, Caneo, Luiz Fernando, Matte, Gregory, Carletto, Luciana P., Policarpo, Valéria Camargo, Castro, Ana Vitória C. X., Miranda, Matheus H. C., Costa, Priscila S., Jatene, Marcelo B., Cestari, Idagene, Jatene, Fabio B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Cirurgia Cardiovascular 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7199987/
https://www.ncbi.nlm.nih.gov/pubmed/32369291
http://dx.doi.org/10.21470/1678-9741-2019-0311
_version_ 1783529251255877632
author Guimarães, Daniel Peres
Caneo, Luiz Fernando
Matte, Gregory
Carletto, Luciana P.
Policarpo, Valéria Camargo
Castro, Ana Vitória C. X.
Miranda, Matheus H. C.
Costa, Priscila S.
Jatene, Marcelo B.
Cestari, Idagene
Jatene, Fabio B.
author_facet Guimarães, Daniel Peres
Caneo, Luiz Fernando
Matte, Gregory
Carletto, Luciana P.
Policarpo, Valéria Camargo
Castro, Ana Vitória C. X.
Miranda, Matheus H. C.
Costa, Priscila S.
Jatene, Marcelo B.
Cestari, Idagene
Jatene, Fabio B.
author_sort Guimarães, Daniel Peres
collection PubMed
description OBJECTIVE: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. METHODS: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min. RESULTS: The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD. CONCLUSION: The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position. An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass.
format Online
Article
Text
id pubmed-7199987
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Sociedade Brasileira de Cirurgia Cardiovascular
record_format MEDLINE/PubMed
spelling pubmed-71999872020-05-08 Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head Guimarães, Daniel Peres Caneo, Luiz Fernando Matte, Gregory Carletto, Luciana P. Policarpo, Valéria Camargo Castro, Ana Vitória C. X. Miranda, Matheus H. C. Costa, Priscila S. Jatene, Marcelo B. Cestari, Idagene Jatene, Fabio B. Braz J Cardiovasc Surg Original Article OBJECTIVE: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. METHODS: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min. RESULTS: The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD. CONCLUSION: The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position. An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass. Sociedade Brasileira de Cirurgia Cardiovascular 2020 /pmc/articles/PMC7199987/ /pubmed/32369291 http://dx.doi.org/10.21470/1678-9741-2019-0311 Text en http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Guimarães, Daniel Peres
Caneo, Luiz Fernando
Matte, Gregory
Carletto, Luciana P.
Policarpo, Valéria Camargo
Castro, Ana Vitória C. X.
Miranda, Matheus H. C.
Costa, Priscila S.
Jatene, Marcelo B.
Cestari, Idagene
Jatene, Fabio B.
Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_full Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_fullStr Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_full_unstemmed Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_short Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_sort impact of vacuum-assisted venous drainage on forward flow in simulated pediatric cardiopulmonary bypass circuits utilizing a centrifugal arterial pump head
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7199987/
https://www.ncbi.nlm.nih.gov/pubmed/32369291
http://dx.doi.org/10.21470/1678-9741-2019-0311
work_keys_str_mv AT guimaraesdanielperes impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT caneoluizfernando impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT mattegregory impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT carlettolucianap impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT policarpovaleriacamargo impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT castroanavitoriacx impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT mirandamatheushc impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT costapriscilas impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT jatenemarcelob impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT cestariidagene impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead
AT jatenefabiob impactofvacuumassistedvenousdrainageonforwardflowinsimulatedpediatriccardiopulmonarybypasscircuitsutilizingacentrifugalarterialpumphead