CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality

PURPOSE: To compare CT pulmonary angiographies (CTPAs) as well as phantom scans obtained at 100 kVp with a conventional CT (C-CT) to virtual monochromatic images (VMI) obtained with a spectral detector CT (SD-CT) at equivalent dose levels as well as to compare the radiation exposure of both systems....

Descripción completa

Detalles Bibliográficos
Autores principales: Sauter, Andreas P., Shapira, Nadav, Kopp, Felix K., Aichele, Juliane, Bodden, Jannis, Knipfer, Andreas, Rummeny, Ernst J., Noël, Peter B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7215101/
https://www.ncbi.nlm.nih.gov/pubmed/32420413
http://dx.doi.org/10.1016/j.ejro.2020.100234
_version_ 1783532108376965120
author Sauter, Andreas P.
Shapira, Nadav
Kopp, Felix K.
Aichele, Juliane
Bodden, Jannis
Knipfer, Andreas
Rummeny, Ernst J.
Noël, Peter B.
author_facet Sauter, Andreas P.
Shapira, Nadav
Kopp, Felix K.
Aichele, Juliane
Bodden, Jannis
Knipfer, Andreas
Rummeny, Ernst J.
Noël, Peter B.
author_sort Sauter, Andreas P.
collection PubMed
description PURPOSE: To compare CT pulmonary angiographies (CTPAs) as well as phantom scans obtained at 100 kVp with a conventional CT (C-CT) to virtual monochromatic images (VMI) obtained with a spectral detector CT (SD-CT) at equivalent dose levels as well as to compare the radiation exposure of both systems. MATERIAL AND METHODS: In total, 2110 patients with suspected pulmonary embolism (PE) were examined with both systems. For each system (C-CT and SD-CT), imaging data of 30 patients with the same mean CT dose index (4.85 mGy) was used for the reader study. C-CT was performed with 100 kVp and SD-CT was performed with 120 kVp; for SD-CT, virtual monochromatic images (VMI) with 40, 60 and 70 keV were calculated. All datasets were evaluated by three blinded radiologists regarding image quality, diagnostic confidence and diagnostic performance (sensitivity, specificity). Contrast-to-noise ratio (CNR) for different iodine concentrations was evaluated in a phantom study. RESULTS: CNR was significantly higher with VMI at 40 keV compared to all other datasets. Subjective image quality as well as sensitivity and specificity showed the highest values with VMI at 60 keV and 70 keV. Hereby, a significant difference to 100 kVp (C-CT) was found for image quality. The highest sensitivity was found using VMI at 60 keV with a sensitivity of more than 97 % for all localizations of PE. For diagnostic confidence and subjective contrast, highest values were found with VMI at 40 keV. CONCLUSION: Higher levels of diagnostic performance and image quality were achieved for CPTAs with SD-CT compared to C-CT given similar dose levels. In the clinical setting SD-CT may be the modality of choice as additional spectral information can be obtained.
format Online
Article
Text
id pubmed-7215101
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-72151012020-05-15 CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality Sauter, Andreas P. Shapira, Nadav Kopp, Felix K. Aichele, Juliane Bodden, Jannis Knipfer, Andreas Rummeny, Ernst J. Noël, Peter B. Eur J Radiol Open Article PURPOSE: To compare CT pulmonary angiographies (CTPAs) as well as phantom scans obtained at 100 kVp with a conventional CT (C-CT) to virtual monochromatic images (VMI) obtained with a spectral detector CT (SD-CT) at equivalent dose levels as well as to compare the radiation exposure of both systems. MATERIAL AND METHODS: In total, 2110 patients with suspected pulmonary embolism (PE) were examined with both systems. For each system (C-CT and SD-CT), imaging data of 30 patients with the same mean CT dose index (4.85 mGy) was used for the reader study. C-CT was performed with 100 kVp and SD-CT was performed with 120 kVp; for SD-CT, virtual monochromatic images (VMI) with 40, 60 and 70 keV were calculated. All datasets were evaluated by three blinded radiologists regarding image quality, diagnostic confidence and diagnostic performance (sensitivity, specificity). Contrast-to-noise ratio (CNR) for different iodine concentrations was evaluated in a phantom study. RESULTS: CNR was significantly higher with VMI at 40 keV compared to all other datasets. Subjective image quality as well as sensitivity and specificity showed the highest values with VMI at 60 keV and 70 keV. Hereby, a significant difference to 100 kVp (C-CT) was found for image quality. The highest sensitivity was found using VMI at 60 keV with a sensitivity of more than 97 % for all localizations of PE. For diagnostic confidence and subjective contrast, highest values were found with VMI at 40 keV. CONCLUSION: Higher levels of diagnostic performance and image quality were achieved for CPTAs with SD-CT compared to C-CT given similar dose levels. In the clinical setting SD-CT may be the modality of choice as additional spectral information can be obtained. Elsevier 2020-05-07 /pmc/articles/PMC7215101/ /pubmed/32420413 http://dx.doi.org/10.1016/j.ejro.2020.100234 Text en © 2020 The Author(s) http://creativecommons.org/licenses/by/4.0/ This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Sauter, Andreas P.
Shapira, Nadav
Kopp, Felix K.
Aichele, Juliane
Bodden, Jannis
Knipfer, Andreas
Rummeny, Ernst J.
Noël, Peter B.
CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality
title CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality
title_full CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality
title_fullStr CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality
title_full_unstemmed CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality
title_short CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality
title_sort ctpa with a conventional ct at 100 kvp vs. a spectral-detector ct at 120 kvp: comparison of radiation exposure, diagnostic performance and image quality
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7215101/
https://www.ncbi.nlm.nih.gov/pubmed/32420413
http://dx.doi.org/10.1016/j.ejro.2020.100234
work_keys_str_mv AT sauterandreasp ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT shapiranadav ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT koppfelixk ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT aichelejuliane ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT boddenjannis ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT knipferandreas ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT rummenyernstj ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality
AT noelpeterb ctpawithaconventionalctat100kvpvsaspectraldetectorctat120kvpcomparisonofradiationexposurediagnosticperformanceandimagequality