Assessment and correction of macroscopic field variations in 2D spoiled gradient‐echo sequences
PURPOSE: To model and correct the dephasing effects in the gradient‐echo signal for arbitrary RF excitation pulses with large flip angles in the presence of macroscopic field variations. METHODS: The dephasing of the spoiled 2D gradient‐echo signal was modeled using a numerical solution of the Bloch...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7216950/ https://www.ncbi.nlm.nih.gov/pubmed/31868260 http://dx.doi.org/10.1002/mrm.28139 |
Sumario: | PURPOSE: To model and correct the dephasing effects in the gradient‐echo signal for arbitrary RF excitation pulses with large flip angles in the presence of macroscopic field variations. METHODS: The dephasing of the spoiled 2D gradient‐echo signal was modeled using a numerical solution of the Bloch equations to calculate the magnitude and phase of the transverse magnetization across the slice profile. Additionally, regional variations of the transmit RF field and slice profile scaling due to macroscopic field gradients were included. Simulations, phantom, and in vivo measurements at 3 T were conducted for [Formula: see text] and myelin water fraction (MWF) mapping. RESULTS: The influence of macroscopic field gradients on [Formula: see text] and myelin water fraction estimation can be substantially reduced by applying the proposed model. Moreover, it was shown that the dephasing over time for flip angles of 60° or greater also depends on the polarity of the slice‐selection gradient because of phase variation along the slice profile. CONCLUSION: Substantial improvements in [Formula: see text] accuracy and myelin water fraction mapping coverage can be achieved using the proposed model if higher flip angles are required. In this context, we demonstrated that the phase along the slice profile and the polarity of the slice‐selection gradient are essential for proper modeling of the gradient‐echo signal in the presence of macroscopic field variations. |
---|