Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T

PURPOSE: Inhomogeneous excitation at ultrahigh field strengths (7T and above) compromises the reliability of quantified dynamic contrast‐enhanced breast MRI. This can hamper the introduction of ultrahigh field MRI into the clinic. Compensation for this non‐uniformity effect can consist of both hardw...

Descripción completa

Detalles Bibliográficos
Autores principales: van Rijssel, Michael J., Pluim, Josien P. W., Chan, Hui‐Shan M., van den Wildenberg, Lieke, Schmitz, Alexander M. Th., Luijten, Peter R., Gilhuijs, Kenneth G. A., Klomp, Dennis W. J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7217168/
https://www.ncbi.nlm.nih.gov/pubmed/31880346
http://dx.doi.org/10.1002/mrm.28147
_version_ 1783532560525033472
author van Rijssel, Michael J.
Pluim, Josien P. W.
Chan, Hui‐Shan M.
van den Wildenberg, Lieke
Schmitz, Alexander M. Th.
Luijten, Peter R.
Gilhuijs, Kenneth G. A.
Klomp, Dennis W. J.
author_facet van Rijssel, Michael J.
Pluim, Josien P. W.
Chan, Hui‐Shan M.
van den Wildenberg, Lieke
Schmitz, Alexander M. Th.
Luijten, Peter R.
Gilhuijs, Kenneth G. A.
Klomp, Dennis W. J.
author_sort van Rijssel, Michael J.
collection PubMed
description PURPOSE: Inhomogeneous excitation at ultrahigh field strengths (7T and above) compromises the reliability of quantified dynamic contrast‐enhanced breast MRI. This can hamper the introduction of ultrahigh field MRI into the clinic. Compensation for this non‐uniformity effect can consist of both hardware improvements and post‐acquisition corrections. This paper investigated the correctable radiofrequency transmit ([Formula: see text]) range post‐acquisition in both simulations and patient data for 7T MRI. METHODS: Simulations were conducted to determine the minimum [Formula: see text] level at which corrections were still beneficial because of noise amplification. Two correction strategies leading to differences in noise amplification were tested. The effect of the corrections on a 7T patient data set (N = 38) with a wide range of [Formula: see text] levels was investigated in terms of time‐intensity curve types as well as washin, washout and peak enhancement values. RESULTS: In simulations assuming a common amount of T(1) saturation, the lowest [Formula: see text] level at which the SNR of the corrected images was at least that of the original precontrast image was 43% of the nominal angle. After correction, time‐intensity curve types changed in 24% of included patients, and the distribution of curve types corresponded better to the distribution found in literature. Additionally, the overlap between the distributions of washin, washout, and peak enhancement values for grade 1 and grade 2 tumors was slightly reduced. CONCLUSION: Although the correctable range varies with the amount of T(1) saturation, post‐acquisition correction for inhomogeneous excitation was feasible down to [Formula: see text] levels of 43% of the nominal angle in vivo.
format Online
Article
Text
id pubmed-7217168
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-72171682020-05-13 Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T van Rijssel, Michael J. Pluim, Josien P. W. Chan, Hui‐Shan M. van den Wildenberg, Lieke Schmitz, Alexander M. Th. Luijten, Peter R. Gilhuijs, Kenneth G. A. Klomp, Dennis W. J. Magn Reson Med Full Papers—Computer Processing and Modeling PURPOSE: Inhomogeneous excitation at ultrahigh field strengths (7T and above) compromises the reliability of quantified dynamic contrast‐enhanced breast MRI. This can hamper the introduction of ultrahigh field MRI into the clinic. Compensation for this non‐uniformity effect can consist of both hardware improvements and post‐acquisition corrections. This paper investigated the correctable radiofrequency transmit ([Formula: see text]) range post‐acquisition in both simulations and patient data for 7T MRI. METHODS: Simulations were conducted to determine the minimum [Formula: see text] level at which corrections were still beneficial because of noise amplification. Two correction strategies leading to differences in noise amplification were tested. The effect of the corrections on a 7T patient data set (N = 38) with a wide range of [Formula: see text] levels was investigated in terms of time‐intensity curve types as well as washin, washout and peak enhancement values. RESULTS: In simulations assuming a common amount of T(1) saturation, the lowest [Formula: see text] level at which the SNR of the corrected images was at least that of the original precontrast image was 43% of the nominal angle. After correction, time‐intensity curve types changed in 24% of included patients, and the distribution of curve types corresponded better to the distribution found in literature. Additionally, the overlap between the distributions of washin, washout, and peak enhancement values for grade 1 and grade 2 tumors was slightly reduced. CONCLUSION: Although the correctable range varies with the amount of T(1) saturation, post‐acquisition correction for inhomogeneous excitation was feasible down to [Formula: see text] levels of 43% of the nominal angle in vivo. John Wiley and Sons Inc. 2019-12-27 2020-08 /pmc/articles/PMC7217168/ /pubmed/31880346 http://dx.doi.org/10.1002/mrm.28147 Text en © 2019 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Full Papers—Computer Processing and Modeling
van Rijssel, Michael J.
Pluim, Josien P. W.
Chan, Hui‐Shan M.
van den Wildenberg, Lieke
Schmitz, Alexander M. Th.
Luijten, Peter R.
Gilhuijs, Kenneth G. A.
Klomp, Dennis W. J.
Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T
title Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T
title_full Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T
title_fullStr Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T
title_full_unstemmed Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T
title_short Correcting time‐intensity curves in dynamic contrast‐enhanced breast MRI for inhomogeneous excitation fields at 7T
title_sort correcting time‐intensity curves in dynamic contrast‐enhanced breast mri for inhomogeneous excitation fields at 7t
topic Full Papers—Computer Processing and Modeling
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7217168/
https://www.ncbi.nlm.nih.gov/pubmed/31880346
http://dx.doi.org/10.1002/mrm.28147
work_keys_str_mv AT vanrijsselmichaelj correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT pluimjosienpw correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT chanhuishanm correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT vandenwildenberglieke correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT schmitzalexandermth correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT luijtenpeterr correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT gilhuijskennethga correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t
AT klompdenniswj correctingtimeintensitycurvesindynamiccontrastenhancedbreastmriforinhomogeneousexcitationfieldsat7t