Cargando…
Impact of CT Injector Technology and Contrast Media Viscosity on Vascular Enhancement: Evaluation in a Circulation Phantom
OBJECTIVE: To assess the impact of piston-based vs peristaltic injection system technology and contrast media viscosity on achievable iodine delivery rates (IDRs) and vascular enhancement in a pre-clinical study. METHODS: Four injectors were tested: MEDRAD(®) Centargo, MEDRAD(®) Stellant, CT Exprès(...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The British Institute of Radiology.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7217576/ https://www.ncbi.nlm.nih.gov/pubmed/32017607 http://dx.doi.org/10.1259/bjr.20190868 |
Sumario: | OBJECTIVE: To assess the impact of piston-based vs peristaltic injection system technology and contrast media viscosity on achievable iodine delivery rates (IDRs) and vascular enhancement in a pre-clinical study. METHODS: Four injectors were tested: MEDRAD(®) Centargo, MEDRAD(®) Stellant, CT Exprès(®), and CT motion(™) using five contrast media [iopromide (300 and 370 mgI ml(−1)), iodixanol 320 mgI ml(−1), iohexol 350 mgI ml(−1), iomeprol 400 mgI ml(−1)]. Three experiments were performed evaluating achievable IDR and corresponding enhancement in a circulation phantom. RESULTS: Experiment I: Centargo provided the highest achievable IDRs with all tested contrast media (p < 0.05). Iopromide 370 yielded the highest IDR with an 18G catheter (3.15 gI/s); iopromide 300 yielded the highest IDR with 20G (2.70 gI/s) and 22G (1.65 gI/s) catheters (p < 0.05). Experiment II: with higher achievable IDRs, piston-based injectors provided significantly higher peak vascular enhancement (up to 48% increase) than the peristaltic injectors with programmed IDRs from 1.8 to 2.4 gI/s (p < 0.05). Experiment III: with programmed IDRs (e.g. 1.5 gI/s) achievable by all injection systems, Centargo, with sharper measured bolus shape, provided significant increases in enhancement of 34–73 HU in the pulmonary artery with iopromide 370 (p < 0.05). CONCLUSION: The tested piston-based injection systems combined with low viscosity contrast media provide higher achievable IDRs and higher peak vascular enhancement than the tested peristaltic-based injectors. With equivalent IDRs, Centargo provides higher peak vascular enhancement due to improved bolus shape. ADVANCES IN KNOWLEDGE: This paper introduces a new parameter to compare expected performance among contrast media: the concentration/viscosity ratio. Additionally, it demonstrates previously unexplored impacts of bolus shape on vascular enhancement. |
---|