Cargando…

In vitro characterization of PrismaLung+: a novel ECCO(2)R device

BACKGROUND: Invasive mechanical ventilation is lifesaving in the setting of severe acute respiratory failure but can cause ventilation-induced lung injury. Advances in extracorporeal CO(2) removal (ECCO(2)R) technologies may facilitate more protective lung ventilation in acute respiratory distress s...

Descripción completa

Detalles Bibliográficos
Autores principales: Hospach, Ingeborg, Goldstein, Jacques, Harenski, Kai, Laffey, John G., Pouchoulin, Dominique, Raible, Manuela, Votteler, Stefanie, Storr, Markus
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7221037/
https://www.ncbi.nlm.nih.gov/pubmed/32405714
http://dx.doi.org/10.1186/s40635-020-00301-7
_version_ 1783533285231558656
author Hospach, Ingeborg
Goldstein, Jacques
Harenski, Kai
Laffey, John G.
Pouchoulin, Dominique
Raible, Manuela
Votteler, Stefanie
Storr, Markus
author_facet Hospach, Ingeborg
Goldstein, Jacques
Harenski, Kai
Laffey, John G.
Pouchoulin, Dominique
Raible, Manuela
Votteler, Stefanie
Storr, Markus
author_sort Hospach, Ingeborg
collection PubMed
description BACKGROUND: Invasive mechanical ventilation is lifesaving in the setting of severe acute respiratory failure but can cause ventilation-induced lung injury. Advances in extracorporeal CO(2) removal (ECCO(2)R) technologies may facilitate more protective lung ventilation in acute respiratory distress syndrome, and enable earlier weaning and/or avoid invasive mechanical ventilation entirely in chronic obstructive pulmonary disease exacerbations. We evaluated the in vitro CO(2) removal capacity of the novel PrismaLung+ ECCO(2)R device compared with two existing gas exchangers. METHODS: The in vitro CO(2) removal capacity of the PrismaLung+ (surface area 0.8 m(2), Baxter) was compared with the PrismaLung (surface area 0.35 m(2), Baxter) and A.L.ONE (surface area 1.35 m(2), Eurosets) devices, using a closed-loop bovine blood–perfused extracorporeal circuit. The efficacy of each device was measured at varying pCO(2) inlet (p(in)CO(2)) levels (45, 60, and 80 mmHg) and blood flow rates (Q(B)) of 200–450 mL/min; the PrismaLung+ and A.L.ONE devices were also tested at a Q(B) of 600 mL/min. The amount of CO(2) removed by each device was assessed by measurement of the CO(2) infused to maintain circuit equilibrium (CO(2) infusion method) and compared with measured CO(2) concentrations in the inlet and outlet of the CO(2) removal device (blood gas analysis method). RESULTS: The PrismaLung+ device performed similarly to the A.L.ONE device, with both devices demonstrating CO(2) removal rates ~ 50% greater than the PrismaLung device. CO(2) removal rates were 73 ± 4.0, 44 ± 2.5, and 72 ± 1.9 mL/min, for PrismaLung+, PrismaLung, and A.L.ONE, respectively, at Q(B) 300 mL/min and p(in)CO(2) 45 mmHg. A Bland–Altman plot demonstrated that the CO(2) infusion method was comparable to the blood gas analysis method for calculating CO(2) removal. The resistance to blood flow across the test device, as measured by pressure drop, varied as a function of blood flow rate, and was greatest for PrismaLung and lowest for the A.L.ONE device. CONCLUSIONS: The newly developed PrismaLung+ performed more effectively than PrismaLung, with performance of CO(2) removal comparable to A.L.ONE at the flow rates tested, despite the smaller membrane surface area of PrismaLung+ versus A.L.ONE. Clinical testing of PrismaLung+ is warranted to further characterize its performance.
format Online
Article
Text
id pubmed-7221037
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-72210372020-05-15 In vitro characterization of PrismaLung+: a novel ECCO(2)R device Hospach, Ingeborg Goldstein, Jacques Harenski, Kai Laffey, John G. Pouchoulin, Dominique Raible, Manuela Votteler, Stefanie Storr, Markus Intensive Care Med Exp Research BACKGROUND: Invasive mechanical ventilation is lifesaving in the setting of severe acute respiratory failure but can cause ventilation-induced lung injury. Advances in extracorporeal CO(2) removal (ECCO(2)R) technologies may facilitate more protective lung ventilation in acute respiratory distress syndrome, and enable earlier weaning and/or avoid invasive mechanical ventilation entirely in chronic obstructive pulmonary disease exacerbations. We evaluated the in vitro CO(2) removal capacity of the novel PrismaLung+ ECCO(2)R device compared with two existing gas exchangers. METHODS: The in vitro CO(2) removal capacity of the PrismaLung+ (surface area 0.8 m(2), Baxter) was compared with the PrismaLung (surface area 0.35 m(2), Baxter) and A.L.ONE (surface area 1.35 m(2), Eurosets) devices, using a closed-loop bovine blood–perfused extracorporeal circuit. The efficacy of each device was measured at varying pCO(2) inlet (p(in)CO(2)) levels (45, 60, and 80 mmHg) and blood flow rates (Q(B)) of 200–450 mL/min; the PrismaLung+ and A.L.ONE devices were also tested at a Q(B) of 600 mL/min. The amount of CO(2) removed by each device was assessed by measurement of the CO(2) infused to maintain circuit equilibrium (CO(2) infusion method) and compared with measured CO(2) concentrations in the inlet and outlet of the CO(2) removal device (blood gas analysis method). RESULTS: The PrismaLung+ device performed similarly to the A.L.ONE device, with both devices demonstrating CO(2) removal rates ~ 50% greater than the PrismaLung device. CO(2) removal rates were 73 ± 4.0, 44 ± 2.5, and 72 ± 1.9 mL/min, for PrismaLung+, PrismaLung, and A.L.ONE, respectively, at Q(B) 300 mL/min and p(in)CO(2) 45 mmHg. A Bland–Altman plot demonstrated that the CO(2) infusion method was comparable to the blood gas analysis method for calculating CO(2) removal. The resistance to blood flow across the test device, as measured by pressure drop, varied as a function of blood flow rate, and was greatest for PrismaLung and lowest for the A.L.ONE device. CONCLUSIONS: The newly developed PrismaLung+ performed more effectively than PrismaLung, with performance of CO(2) removal comparable to A.L.ONE at the flow rates tested, despite the smaller membrane surface area of PrismaLung+ versus A.L.ONE. Clinical testing of PrismaLung+ is warranted to further characterize its performance. Springer International Publishing 2020-05-13 /pmc/articles/PMC7221037/ /pubmed/32405714 http://dx.doi.org/10.1186/s40635-020-00301-7 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Research
Hospach, Ingeborg
Goldstein, Jacques
Harenski, Kai
Laffey, John G.
Pouchoulin, Dominique
Raible, Manuela
Votteler, Stefanie
Storr, Markus
In vitro characterization of PrismaLung+: a novel ECCO(2)R device
title In vitro characterization of PrismaLung+: a novel ECCO(2)R device
title_full In vitro characterization of PrismaLung+: a novel ECCO(2)R device
title_fullStr In vitro characterization of PrismaLung+: a novel ECCO(2)R device
title_full_unstemmed In vitro characterization of PrismaLung+: a novel ECCO(2)R device
title_short In vitro characterization of PrismaLung+: a novel ECCO(2)R device
title_sort in vitro characterization of prismalung+: a novel ecco(2)r device
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7221037/
https://www.ncbi.nlm.nih.gov/pubmed/32405714
http://dx.doi.org/10.1186/s40635-020-00301-7
work_keys_str_mv AT hospachingeborg invitrocharacterizationofprismalunganovelecco2rdevice
AT goldsteinjacques invitrocharacterizationofprismalunganovelecco2rdevice
AT harenskikai invitrocharacterizationofprismalunganovelecco2rdevice
AT laffeyjohng invitrocharacterizationofprismalunganovelecco2rdevice
AT pouchoulindominique invitrocharacterizationofprismalunganovelecco2rdevice
AT raiblemanuela invitrocharacterizationofprismalunganovelecco2rdevice
AT vottelerstefanie invitrocharacterizationofprismalunganovelecco2rdevice
AT storrmarkus invitrocharacterizationofprismalunganovelecco2rdevice