Cargando…
Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study
BACKGROUND: 3D-time resolved flow (4DF) cardiovascular magnetic resonance (CMR) with retrospective analysis of atrioventricular valve regurgitation (AVVR) allows for internal validation by multiple direct and indirect methods. Limited data exist on direct measurement of AVVR by 4DF CMR in pediatric...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7222506/ https://www.ncbi.nlm.nih.gov/pubmed/32404159 http://dx.doi.org/10.1186/s12968-020-00612-4 |
_version_ | 1783533590293774336 |
---|---|
author | Jacobs, Kimberley Rigdon, Joseph Chan, Frandics Cheng, Joseph Y. Alley, Marcus T. Vasanawala, Shreyas Maskatia, Shiraz A. |
author_facet | Jacobs, Kimberley Rigdon, Joseph Chan, Frandics Cheng, Joseph Y. Alley, Marcus T. Vasanawala, Shreyas Maskatia, Shiraz A. |
author_sort | Jacobs, Kimberley |
collection | PubMed |
description | BACKGROUND: 3D-time resolved flow (4DF) cardiovascular magnetic resonance (CMR) with retrospective analysis of atrioventricular valve regurgitation (AVVR) allows for internal validation by multiple direct and indirect methods. Limited data exist on direct measurement of AVVR by 4DF CMR in pediatric congenital heart disease (CHD). We aimed to validate direct measurement of the AVVR jet as accurate and reliable compared to the volumetric method (clinical standard by 2D CMR) and as a superior method of internal validation than the annular inflow method. METHODS: We identified 44 consecutive patients with diverse CHD referred for evaluation of AVVR by CMR. 1.5 T or 3 T scanners, intravenous contrast, and a combination of parallel imaging and compressed sensing were used. Four methods of measuring AVVR volume (RVol) were used: volumetric method (VOL; the clinical standard) = stroke volume by 2D balanced steady-state free precession – semilunar valve forward flow (SLFF); annular inflow method (AIM) = atrioventricular valve forward flow [AVFF] – semilunar valve net flow (SLNF); and direct measurement (JET). AVFF was measured using static and retrospective valve tracking planes. SLFF, SLNF, AVFF, and JET were measured by 4DF phase contrast. Regurgitant fraction was calculated as [RVol/(RVol+SLNF)]× 100. Statistical methods included Spearman, Wilcoxon rank sum test/Student paired t-test, Bland Altman analysis, and intra-class coefficient (ICC), where appropriate. RESULTS: Regurgitant fraction by JET strongly correlated with the indirect methods (VOL and AIM) (ρ = 0.73–0.80, p < 0.001) and was similar to VOL with a median difference (interquartile range) of − 1.5% (− 8.3–7.2%; p = 0.624). VOL had weaker correlations with AIM and JET (ρ = 0.69–0.73, p < 0.001). AIM underestimated RF by 3.6–6.9% compared to VOL and JET, p < 0.03. Intra- and inter- observer reliability were excellent for all methods (ICC 0.94–0.99). The mean (±standard deviation) inter-observer difference for VOL was 2.4% (±5.1%), p < 0.05. CONCLUSIONS: In a diverse cohort of pediatric CHD, measurement of AVVR using JET is accurate and reliable to VOL and is a superior method of internal validation compared to AIM. This study supports use of 4DF CMR for measurement of AVVR, obviating need for expert prospective prescription during image acquisition by 2D CMR. |
format | Online Article Text |
id | pubmed-7222506 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-72225062020-05-20 Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study Jacobs, Kimberley Rigdon, Joseph Chan, Frandics Cheng, Joseph Y. Alley, Marcus T. Vasanawala, Shreyas Maskatia, Shiraz A. J Cardiovasc Magn Reson Research BACKGROUND: 3D-time resolved flow (4DF) cardiovascular magnetic resonance (CMR) with retrospective analysis of atrioventricular valve regurgitation (AVVR) allows for internal validation by multiple direct and indirect methods. Limited data exist on direct measurement of AVVR by 4DF CMR in pediatric congenital heart disease (CHD). We aimed to validate direct measurement of the AVVR jet as accurate and reliable compared to the volumetric method (clinical standard by 2D CMR) and as a superior method of internal validation than the annular inflow method. METHODS: We identified 44 consecutive patients with diverse CHD referred for evaluation of AVVR by CMR. 1.5 T or 3 T scanners, intravenous contrast, and a combination of parallel imaging and compressed sensing were used. Four methods of measuring AVVR volume (RVol) were used: volumetric method (VOL; the clinical standard) = stroke volume by 2D balanced steady-state free precession – semilunar valve forward flow (SLFF); annular inflow method (AIM) = atrioventricular valve forward flow [AVFF] – semilunar valve net flow (SLNF); and direct measurement (JET). AVFF was measured using static and retrospective valve tracking planes. SLFF, SLNF, AVFF, and JET were measured by 4DF phase contrast. Regurgitant fraction was calculated as [RVol/(RVol+SLNF)]× 100. Statistical methods included Spearman, Wilcoxon rank sum test/Student paired t-test, Bland Altman analysis, and intra-class coefficient (ICC), where appropriate. RESULTS: Regurgitant fraction by JET strongly correlated with the indirect methods (VOL and AIM) (ρ = 0.73–0.80, p < 0.001) and was similar to VOL with a median difference (interquartile range) of − 1.5% (− 8.3–7.2%; p = 0.624). VOL had weaker correlations with AIM and JET (ρ = 0.69–0.73, p < 0.001). AIM underestimated RF by 3.6–6.9% compared to VOL and JET, p < 0.03. Intra- and inter- observer reliability were excellent for all methods (ICC 0.94–0.99). The mean (±standard deviation) inter-observer difference for VOL was 2.4% (±5.1%), p < 0.05. CONCLUSIONS: In a diverse cohort of pediatric CHD, measurement of AVVR using JET is accurate and reliable to VOL and is a superior method of internal validation compared to AIM. This study supports use of 4DF CMR for measurement of AVVR, obviating need for expert prospective prescription during image acquisition by 2D CMR. BioMed Central 2020-05-14 /pmc/articles/PMC7222506/ /pubmed/32404159 http://dx.doi.org/10.1186/s12968-020-00612-4 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Jacobs, Kimberley Rigdon, Joseph Chan, Frandics Cheng, Joseph Y. Alley, Marcus T. Vasanawala, Shreyas Maskatia, Shiraz A. Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
title | Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
title_full | Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
title_fullStr | Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
title_full_unstemmed | Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
title_short | Direct measurement of atrioventricular valve regurgitant jets using 4D flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
title_sort | direct measurement of atrioventricular valve regurgitant jets using 4d flow cardiovascular magnetic resonance is accurate and reliable for children with congenital heart disease: a retrospective cohort study |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7222506/ https://www.ncbi.nlm.nih.gov/pubmed/32404159 http://dx.doi.org/10.1186/s12968-020-00612-4 |
work_keys_str_mv | AT jacobskimberley directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy AT rigdonjoseph directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy AT chanfrandics directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy AT chengjosephy directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy AT alleymarcust directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy AT vasanawalashreyas directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy AT maskatiashiraza directmeasurementofatrioventricularvalveregurgitantjetsusing4dflowcardiovascularmagneticresonanceisaccurateandreliableforchildrenwithcongenitalheartdiseasearetrospectivecohortstudy |