Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients

The measurement of the optic nerve sheath diameter (ONSD) by ultrasonography (USG) is particularly important for intracranial pressure (ICP) monitoring when invasive measurements are not possible or are contraindicated. Standard USG (SUDs) devices can be bulky and may break down the workflow. The va...

Descripción completa

Detalles Bibliográficos
Autores principales: Özdemir, Uğur, Çimen, Meltem, Güney, Tuba, Gürsel, Gül
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7223003/
https://www.ncbi.nlm.nih.gov/pubmed/31278545
http://dx.doi.org/10.1007/s10877-019-00351-7
_version_ 1783533686285664256
author Özdemir, Uğur
Çimen, Meltem
Güney, Tuba
Gürsel, Gül
author_facet Özdemir, Uğur
Çimen, Meltem
Güney, Tuba
Gürsel, Gül
author_sort Özdemir, Uğur
collection PubMed
description The measurement of the optic nerve sheath diameter (ONSD) by ultrasonography (USG) is particularly important for intracranial pressure (ICP) monitoring when invasive measurements are not possible or are contraindicated. Standard USG (SUDs) devices can be bulky and may break down the workflow. The validity and reliability of pocket-sized USG devices (PSUDs) compared to SUDs for ONSD measurement has not been investigated, yet. We compared the reliability and validity of PSUDs with SUDs for ONSD measurement. 35 patients were included in the study. ONSD measurements with PSUDs and SUDs were performed by two clinicians as three different measurements in both sagittal and transverse axis of the eye. There was agreement between mean transverse ONSD (mtONSD), mean sagittal ONSD (msONSD), and mean ONSD(mONSD) values measured with each device according to Bland–Altman test (p = 0.166, p = 0.135, p = 0.066, respectively) with no proportional bias (p = 0.544, p = 0.868, p = 0.929, respectively). Intraclass correlation coefficients (ICC) were found to be greater than 0.7 for mtONSD, msONSD, and mONSD values measured by SUD (ICC = 0.790, ICC = 0.817, and ICC = 0.844, respectively). Similarly, for mtONSD, msONSD, and mONSD values measured by PSUD, ICC were found greater than 0.7 (ICC = 0.763, ICC = 0.814, and ICC = 0.843, respectively). There was no statistically significant difference between mtONSD and msONSD (p = 0.441 for SUD and p = 0.893 for PSUD). There was a good correlation between mtONSD and msONSD (r = 0.767 for SUD and r = 0.816 for PSUD). The total variance between three different measurements in the transverse and sagittal axis was ± 0.6 mm. PSUDs can be used with similar validity and reliability as SUDs for ONSD measurement.
format Online
Article
Text
id pubmed-7223003
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Springer Netherlands
record_format MEDLINE/PubMed
spelling pubmed-72230032020-05-15 Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients Özdemir, Uğur Çimen, Meltem Güney, Tuba Gürsel, Gül J Clin Monit Comput Original Research The measurement of the optic nerve sheath diameter (ONSD) by ultrasonography (USG) is particularly important for intracranial pressure (ICP) monitoring when invasive measurements are not possible or are contraindicated. Standard USG (SUDs) devices can be bulky and may break down the workflow. The validity and reliability of pocket-sized USG devices (PSUDs) compared to SUDs for ONSD measurement has not been investigated, yet. We compared the reliability and validity of PSUDs with SUDs for ONSD measurement. 35 patients were included in the study. ONSD measurements with PSUDs and SUDs were performed by two clinicians as three different measurements in both sagittal and transverse axis of the eye. There was agreement between mean transverse ONSD (mtONSD), mean sagittal ONSD (msONSD), and mean ONSD(mONSD) values measured with each device according to Bland–Altman test (p = 0.166, p = 0.135, p = 0.066, respectively) with no proportional bias (p = 0.544, p = 0.868, p = 0.929, respectively). Intraclass correlation coefficients (ICC) were found to be greater than 0.7 for mtONSD, msONSD, and mONSD values measured by SUD (ICC = 0.790, ICC = 0.817, and ICC = 0.844, respectively). Similarly, for mtONSD, msONSD, and mONSD values measured by PSUD, ICC were found greater than 0.7 (ICC = 0.763, ICC = 0.814, and ICC = 0.843, respectively). There was no statistically significant difference between mtONSD and msONSD (p = 0.441 for SUD and p = 0.893 for PSUD). There was a good correlation between mtONSD and msONSD (r = 0.767 for SUD and r = 0.816 for PSUD). The total variance between three different measurements in the transverse and sagittal axis was ± 0.6 mm. PSUDs can be used with similar validity and reliability as SUDs for ONSD measurement. Springer Netherlands 2019-07-05 2020 /pmc/articles/PMC7223003/ /pubmed/31278545 http://dx.doi.org/10.1007/s10877-019-00351-7 Text en © Springer Nature B.V. 2019 This article is made available via the PMC Open Access Subset for unrestricted research re-use and secondary analysis in any form or by any means with acknowledgement of the original source. These permissions are granted for the duration of the World Health Organization (WHO) declaration of COVID-19 as a global pandemic.
spellingShingle Original Research
Özdemir, Uğur
Çimen, Meltem
Güney, Tuba
Gürsel, Gül
Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients
title Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients
title_full Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients
title_fullStr Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients
title_full_unstemmed Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients
title_short Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients
title_sort validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in icu patients
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7223003/
https://www.ncbi.nlm.nih.gov/pubmed/31278545
http://dx.doi.org/10.1007/s10877-019-00351-7
work_keys_str_mv AT ozdemirugur validityandreliabilityofpocketsizedultrasounddevicesinmeasurementofopticnervesheathdiameterinicupatients
AT cimenmeltem validityandreliabilityofpocketsizedultrasounddevicesinmeasurementofopticnervesheathdiameterinicupatients
AT guneytuba validityandreliabilityofpocketsizedultrasounddevicesinmeasurementofopticnervesheathdiameterinicupatients
AT gurselgul validityandreliabilityofpocketsizedultrasounddevicesinmeasurementofopticnervesheathdiameterinicupatients