Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T
BACKGROUND: Cardiovascular magnetic resonance (CMR) studies in patients with implanted cardioverter/defibrillators (ICD) are increasingly required in daily clinical practice. However, the clinical experience regarding the feasibility as well as clinical value of CMR studies in patients with subcutan...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7232845/ https://www.ncbi.nlm.nih.gov/pubmed/32418537 http://dx.doi.org/10.1186/s12968-020-00626-y |
_version_ | 1783535463922925568 |
---|---|
author | Holtstiege, Viktoria Meier, Claudia Bietenbeck, Michael Chatzantonis, Grigorios Florian, Anca Köbe, Julia Reinke, Florian Eckardt, Lars Yilmaz, Ali |
author_facet | Holtstiege, Viktoria Meier, Claudia Bietenbeck, Michael Chatzantonis, Grigorios Florian, Anca Köbe, Julia Reinke, Florian Eckardt, Lars Yilmaz, Ali |
author_sort | Holtstiege, Viktoria |
collection | PubMed |
description | BACKGROUND: Cardiovascular magnetic resonance (CMR) studies in patients with implanted cardioverter/defibrillators (ICD) are increasingly required in daily clinical practice. However, the clinical experience regarding the feasibility as well as clinical value of CMR studies in patients with subcutaneous ICD (S-ICD) is still limited. Besides safety issues, image quality and analysis can be impaired primarily due the presence of image artefacts associated with the generator. METHODS: Twenty-three patients with an implanted S-ICD (EMBLEM, Boston Scientific, Marlborough, Massachusetts, USA; MR-conditional) with suspected cardiomyopathy and/or myocarditis underwent multi-parametric CMR imaging. Studies were performed on a 1.5 T CMR scanner after device interrogation and comprised standard a) balanced steady state free precession cine, b) T2 weighted-edema, c) velocity-encoded cine flow, d) myocardial perfusion, e) late-gadolinium-enhancement (LGE)-imaging and f) 3D-CMR angiography of the aorta. In case of substantial artefacts, alternative CMR techniques such as spoiled gradient-echo cine-sequences and wide-band inversion-recovery LGE (wb-LGE) sequences were applied. RESULTS: Successful CMR studies could be performed in all patients without any case of unexpected early termination or relevant technical complication other than permanent loss of the S-ICD system beeper volume in 52% of our patients. Assessment of cine-CMR images was predominantly impaired in the left ventricular (LV) anterior, lateral and inferior wall segments and a switch to spoiled gradient echo-based cine-CMR allowed an accurate assessment of cine-images in N = 17 (74%) patients with only limited artefacts. Hyperintensity artefacts in conventional LGE-images were predominantly observed in the LV anterior, lateral and inferior wall segments and image optimisation by use of the wb-LGE was helpful in 15 (65%) cases. Aortic flow measurements and 3D-CMR angiography were assessable in all patients Perfusion imaging artefacts precluded a meaningful assessment in at least one half of the patients. A benefit in clinical-decision making was documented in 17 (74%) patients in the present study. CONCLUSION: Safe 1.5 T CMR imaging was possible in all patients with an S-ICD, though the majority had permanent loss of the S-ICD beeper volume. Achieving good image quality may be challenging in some patients - particularly for perfusion imaging. Using spoiled gradient echo-based cine-sequences and wb-LGE sequences may help to reduce the extent of artefacts, thereby allowing accurate cardiac assessment. Thus, 1.5 T CMR studies should not be withhold in patients with S-ICD for safety concerns and/or fear of extensive imaging artefacts precluding successful image analysis. |
format | Online Article Text |
id | pubmed-7232845 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-72328452020-05-27 Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T Holtstiege, Viktoria Meier, Claudia Bietenbeck, Michael Chatzantonis, Grigorios Florian, Anca Köbe, Julia Reinke, Florian Eckardt, Lars Yilmaz, Ali J Cardiovasc Magn Reson Research BACKGROUND: Cardiovascular magnetic resonance (CMR) studies in patients with implanted cardioverter/defibrillators (ICD) are increasingly required in daily clinical practice. However, the clinical experience regarding the feasibility as well as clinical value of CMR studies in patients with subcutaneous ICD (S-ICD) is still limited. Besides safety issues, image quality and analysis can be impaired primarily due the presence of image artefacts associated with the generator. METHODS: Twenty-three patients with an implanted S-ICD (EMBLEM, Boston Scientific, Marlborough, Massachusetts, USA; MR-conditional) with suspected cardiomyopathy and/or myocarditis underwent multi-parametric CMR imaging. Studies were performed on a 1.5 T CMR scanner after device interrogation and comprised standard a) balanced steady state free precession cine, b) T2 weighted-edema, c) velocity-encoded cine flow, d) myocardial perfusion, e) late-gadolinium-enhancement (LGE)-imaging and f) 3D-CMR angiography of the aorta. In case of substantial artefacts, alternative CMR techniques such as spoiled gradient-echo cine-sequences and wide-band inversion-recovery LGE (wb-LGE) sequences were applied. RESULTS: Successful CMR studies could be performed in all patients without any case of unexpected early termination or relevant technical complication other than permanent loss of the S-ICD system beeper volume in 52% of our patients. Assessment of cine-CMR images was predominantly impaired in the left ventricular (LV) anterior, lateral and inferior wall segments and a switch to spoiled gradient echo-based cine-CMR allowed an accurate assessment of cine-images in N = 17 (74%) patients with only limited artefacts. Hyperintensity artefacts in conventional LGE-images were predominantly observed in the LV anterior, lateral and inferior wall segments and image optimisation by use of the wb-LGE was helpful in 15 (65%) cases. Aortic flow measurements and 3D-CMR angiography were assessable in all patients Perfusion imaging artefacts precluded a meaningful assessment in at least one half of the patients. A benefit in clinical-decision making was documented in 17 (74%) patients in the present study. CONCLUSION: Safe 1.5 T CMR imaging was possible in all patients with an S-ICD, though the majority had permanent loss of the S-ICD beeper volume. Achieving good image quality may be challenging in some patients - particularly for perfusion imaging. Using spoiled gradient echo-based cine-sequences and wb-LGE sequences may help to reduce the extent of artefacts, thereby allowing accurate cardiac assessment. Thus, 1.5 T CMR studies should not be withhold in patients with S-ICD for safety concerns and/or fear of extensive imaging artefacts precluding successful image analysis. BioMed Central 2020-05-18 /pmc/articles/PMC7232845/ /pubmed/32418537 http://dx.doi.org/10.1186/s12968-020-00626-y Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Holtstiege, Viktoria Meier, Claudia Bietenbeck, Michael Chatzantonis, Grigorios Florian, Anca Köbe, Julia Reinke, Florian Eckardt, Lars Yilmaz, Ali Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T |
title | Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T |
title_full | Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T |
title_fullStr | Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T |
title_full_unstemmed | Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T |
title_short | Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T |
title_sort | clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (s-icd) at 1.5 t |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7232845/ https://www.ncbi.nlm.nih.gov/pubmed/32418537 http://dx.doi.org/10.1186/s12968-020-00626-y |
work_keys_str_mv | AT holtstiegeviktoria clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT meierclaudia clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT bietenbeckmichael clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT chatzantonisgrigorios clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT floriananca clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT kobejulia clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT reinkeflorian clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT eckardtlars clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t AT yilmazali clinicalexperienceregardingsafetyanddiagnosticvalueofcardiovascularmagneticresonanceinpatientswithasubcutaneousimplantedcardioverterdefibrillatorsicdat15t |