Cargando…

Fracture of cobalt-crome modular neck in total hip arthroplasty

Despite the advantages of modular total hip arthroplasty in terms of neck version, offset and length precise reproduction, titanium necks breakage became a concern. Consequently, titanium has been replaced by cobalt-chrome (Co-Cr). However, four cases of Co-Cr modular neck breakage have been reporte...

Descripción completa

Detalles Bibliográficos
Autores principales: Murena, Luigi, Maritan, Guido, Concina, Chiara, Scamacca, Veronica, Ratti, Chiara, Canton, Gianluca
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mattioli 1885 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7233721/
https://www.ncbi.nlm.nih.gov/pubmed/31821307
http://dx.doi.org/10.23750/abm.v90i12-S.8941
Descripción
Sumario:Despite the advantages of modular total hip arthroplasty in terms of neck version, offset and length precise reproduction, titanium necks breakage became a concern. Consequently, titanium has been replaced by cobalt-chrome (Co-Cr). However, four cases of Co-Cr modular neck breakage have been reported in the literature. In the present paper, two cases of Co-Cr modular neck fractures are described together with a literature review. The aim of this work is to discuss the risk factors and characteristics of this rare complication. We described two cases of fracture of long varus Co-Cr modular femoral neck connected with cementless press-fit stem. Some risk factors, such as long varus type of modular neck, overweight and/or high demanding physical activity, might have contributed to implant failure. (www.actabiomedica.it)