Cargando…

Minimal Cardiac Perforation by Lead Pacemaker Complicated with Pericardial Effusion and Impending Tamponade: Optimal Management with No Pericardiocentesis Driven by Echocardiography

We report the case of a 51-year-old patient who underwent the implantation of a bi-ventricular implantable cardioverter defibrillator (ICD) complicated by a sub-acute right ventricular minimal perforation with pericardial effusion and echocardiographic signs of tamponade. A new echocardiographic pla...

Descripción completa

Detalles Bibliográficos
Autores principales: Caiati, Carlo, Pollice, Paolo, Truncellito, Luigi, Lepera, Mario Erminio, Favale, Stefano
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7235835/
https://www.ncbi.nlm.nih.gov/pubmed/32235447
http://dx.doi.org/10.3390/diagnostics10040191
Descripción
Sumario:We report the case of a 51-year-old patient who underwent the implantation of a bi-ventricular implantable cardioverter defibrillator (ICD) complicated by a sub-acute right ventricular minimal perforation with pericardial effusion and echocardiographic signs of tamponade. A new echocardiographic plane orientation allowed us to diagnose this condition in emergency and to make the right decision without delay, which consisting in unscrewing the active fixation screw under fluoroscopy guidance, while the pericardiocentesis was postponed. Thanks to the intervention focused on eliminating the cause of the postcardiac injury syndrome, the patient recovered rapidly and ultimately avoided the pericardiocentesis procedure.