Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length
BACKGROUND: A major focus of toric intraocular lens (IOL) implantation is the rotational stability, especially in the patients with long axial length (AL). In this study, we aimed to evaluate the clinical outcomes after implantation of TECNIS toric IOL in eyes with long AL and identify factors influ...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7249429/ https://www.ncbi.nlm.nih.gov/pubmed/32450828 http://dx.doi.org/10.1186/s12886-020-01465-2 |
_version_ | 1783538591882805248 |
---|---|
author | He, Suhong Chen, Xiang Wu, Xingdi Ma, Yajuan Yu, Xuewen Xu, Wen |
author_facet | He, Suhong Chen, Xiang Wu, Xingdi Ma, Yajuan Yu, Xuewen Xu, Wen |
author_sort | He, Suhong |
collection | PubMed |
description | BACKGROUND: A major focus of toric intraocular lens (IOL) implantation is the rotational stability, especially in the patients with long axial length (AL). In this study, we aimed to evaluate the clinical outcomes after implantation of TECNIS toric IOL in eyes with long AL and identify factors influencing their early-stage stability with preoperative corneal astigmatism. METHODS: The study population consisted of 64 eyes from 52 cataract patients, and these patients had preoperative corneal astigmatism between 1.0 and 3.7 diopters (D) and underwent phacoemulsification and TECNIS toric IOL implantation. Ophthalmic biological measurements were carried out preoperatively, including AL, anterior chamber depth (ACD), lens thickness (LT), vitreous length (VL), anterior chamber volume (ACV), sulcus-to-sulcus (STS) and keratometric value (K). Clinical examinations, including visual acuity, manifest refraction, keratometry, digital anterior segment photographs with pupillary dilation, were performed at 1 and 3 months after surgery. RESULTS: The mean best corrected distance visual acuity (BCDVA) was improved from 0.93 ± 0.35 logarithms of the minimal angle of resolution (logMAR) preoperatively to 0.07 ± 0.10 logMAR postoperatively at 3 months after surgery. The mean residual astigmatism (RAS) was 0.91 ± 0.74D at 3 months, which was significantly decreased compared with the preoperative corneal astigmatism of 1.71 ± 0.55 D. The mean absolute rotation of TECNIS toric IOL at 1 and 3 months was 7.42 ± 11.32 degree (°) (0–79°) and 7.48 ± 11.19°(0–79°), respectively. The mean area of capsulorhexis and the overlapped area between capsulorhexis and IOL optic intraoperatively was 21.04 ± 3.30 mm(2) and 7.40 ± 2.87 mm(2).A positive correlation was found between IOL rotation and the area of capsulorhexis (p = 0.017) at 3 months after surgery. No correlation was found between IOL rotation and AL (p = 0.876), ACD (p = 0.387), LT (p = 0.523), VL (p = 0.546), ACV (p = 0.480), STS (p = 0.884), K1 (p = 0.429), K2 (p = 0.644), average of K1 and K2 (p = 0.520), intraoperative IOL axial direction (p = 0.396), preoperative corneal astigmatism (p = 0.269) or the overlapped area between capsulorhexis and IOL optic intraoperatively (p = 0.131) . CONCLUSIONS: The large CCC was a risk factor for toric IOL rotation. An appropriately smaller sized CCC was conducive to increase the rotational stability of TECNIS toric IOL implantation in cataract cases with long AL. |
format | Online Article Text |
id | pubmed-7249429 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-72494292020-06-04 Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length He, Suhong Chen, Xiang Wu, Xingdi Ma, Yajuan Yu, Xuewen Xu, Wen BMC Ophthalmol Research Article BACKGROUND: A major focus of toric intraocular lens (IOL) implantation is the rotational stability, especially in the patients with long axial length (AL). In this study, we aimed to evaluate the clinical outcomes after implantation of TECNIS toric IOL in eyes with long AL and identify factors influencing their early-stage stability with preoperative corneal astigmatism. METHODS: The study population consisted of 64 eyes from 52 cataract patients, and these patients had preoperative corneal astigmatism between 1.0 and 3.7 diopters (D) and underwent phacoemulsification and TECNIS toric IOL implantation. Ophthalmic biological measurements were carried out preoperatively, including AL, anterior chamber depth (ACD), lens thickness (LT), vitreous length (VL), anterior chamber volume (ACV), sulcus-to-sulcus (STS) and keratometric value (K). Clinical examinations, including visual acuity, manifest refraction, keratometry, digital anterior segment photographs with pupillary dilation, were performed at 1 and 3 months after surgery. RESULTS: The mean best corrected distance visual acuity (BCDVA) was improved from 0.93 ± 0.35 logarithms of the minimal angle of resolution (logMAR) preoperatively to 0.07 ± 0.10 logMAR postoperatively at 3 months after surgery. The mean residual astigmatism (RAS) was 0.91 ± 0.74D at 3 months, which was significantly decreased compared with the preoperative corneal astigmatism of 1.71 ± 0.55 D. The mean absolute rotation of TECNIS toric IOL at 1 and 3 months was 7.42 ± 11.32 degree (°) (0–79°) and 7.48 ± 11.19°(0–79°), respectively. The mean area of capsulorhexis and the overlapped area between capsulorhexis and IOL optic intraoperatively was 21.04 ± 3.30 mm(2) and 7.40 ± 2.87 mm(2).A positive correlation was found between IOL rotation and the area of capsulorhexis (p = 0.017) at 3 months after surgery. No correlation was found between IOL rotation and AL (p = 0.876), ACD (p = 0.387), LT (p = 0.523), VL (p = 0.546), ACV (p = 0.480), STS (p = 0.884), K1 (p = 0.429), K2 (p = 0.644), average of K1 and K2 (p = 0.520), intraoperative IOL axial direction (p = 0.396), preoperative corneal astigmatism (p = 0.269) or the overlapped area between capsulorhexis and IOL optic intraoperatively (p = 0.131) . CONCLUSIONS: The large CCC was a risk factor for toric IOL rotation. An appropriately smaller sized CCC was conducive to increase the rotational stability of TECNIS toric IOL implantation in cataract cases with long AL. BioMed Central 2020-05-25 /pmc/articles/PMC7249429/ /pubmed/32450828 http://dx.doi.org/10.1186/s12886-020-01465-2 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Article He, Suhong Chen, Xiang Wu, Xingdi Ma, Yajuan Yu, Xuewen Xu, Wen Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length |
title | Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length |
title_full | Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length |
title_fullStr | Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length |
title_full_unstemmed | Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length |
title_short | Early-stage clinical outcomes and rotational stability of TECNIS toric intraocular lens implantation in cataract cases with long axial length |
title_sort | early-stage clinical outcomes and rotational stability of tecnis toric intraocular lens implantation in cataract cases with long axial length |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7249429/ https://www.ncbi.nlm.nih.gov/pubmed/32450828 http://dx.doi.org/10.1186/s12886-020-01465-2 |
work_keys_str_mv | AT hesuhong earlystageclinicaloutcomesandrotationalstabilityoftecnistoricintraocularlensimplantationincataractcaseswithlongaxiallength AT chenxiang earlystageclinicaloutcomesandrotationalstabilityoftecnistoricintraocularlensimplantationincataractcaseswithlongaxiallength AT wuxingdi earlystageclinicaloutcomesandrotationalstabilityoftecnistoricintraocularlensimplantationincataractcaseswithlongaxiallength AT mayajuan earlystageclinicaloutcomesandrotationalstabilityoftecnistoricintraocularlensimplantationincataractcaseswithlongaxiallength AT yuxuewen earlystageclinicaloutcomesandrotationalstabilityoftecnistoricintraocularlensimplantationincataractcaseswithlongaxiallength AT xuwen earlystageclinicaloutcomesandrotationalstabilityoftecnistoricintraocularlensimplantationincataractcaseswithlongaxiallength |