Cargando…

Guidewire Method for Measuring Local Left Ventricular Electrical Activation Time During Cardiac Resynchronization Implantation

The timing of local activation at left ventricular (LV) pacing leads is measured from the onset of the QRS complex to the peak of the LV electrogram (QLV). Pacing from the sites of late activation is associated with higher response rates to cardiac resynchronization therapy (CRT). Prior studies have...

Descripción completa

Detalles Bibliográficos
Autores principales: Rials, Seth J., Pershing, Michele, Collins, Christy
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MediaSphere Medical 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7252739/
https://www.ncbi.nlm.nih.gov/pubmed/32477783
http://dx.doi.org/10.19102/icrm.2018.090102
Descripción
Sumario:The timing of local activation at left ventricular (LV) pacing leads is measured from the onset of the QRS complex to the peak of the LV electrogram (QLV). Pacing from the sites of late activation is associated with higher response rates to cardiac resynchronization therapy (CRT). Prior studies have measured QLV from permanent pacing leads, or have used electroanatomic mapping systems. The current study compares QLV measurements made with a guidewire to those collected from permanent LV pacing leads positioned at the same venous site without the use of electroanatomic mapping systems. In this study, 20 patients undergoing CRT implantation (14 males, mean QRS: 164.0 ms) had QLV measurements taken using a guidewire. QLV and LV electrogram duration measurements were made at LV pacing sites, and were repeated after positioning the permanent LV pacing lead at the same site. There was no difference in QLV measurements obtained using a guidewire and those obtained using the permanent pacing lead placed at the same site (p = 0.569). QLV measurements obtained with a guidewire and the permanent LV pacing lead at the same site, respectively, were strongly correlated (r = 0.965; p < 0.001). The median absolute difference in electrogram duration was 7.0 ms (p = 0.55). The average time required to make QLV measurements using the guidewire was 11.7 minutes [standard deviation (SD): 6.8]. The average total fluoroscopy time for the entire CRT implant procedure was 10.9 minutes (SD: 5.1). In light of these results, it can be suggested that a guidewire can be used to prospectively measure LV prior to selection or placement of a permanent pacing lead without the use of an electroanatomic mapping system.