Cargando…
Balloon-expandable versus self-expanding transcatheter aortic valve replacement for bioprosthetic dysfunction: A systematic review and meta-analysis
BACKGROUND: Transcatheter aortic valve-in-valve (VIV) procedure is a safe alternative to conventional reoperation for bioprosthetic dysfunction. Balloon-expandable valve (BEV) and self-expanding valve (SEV) are the 2 major types of devices used. Evidence regarding the comparison of the 2 valves rema...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7263630/ https://www.ncbi.nlm.nih.gov/pubmed/32479546 http://dx.doi.org/10.1371/journal.pone.0233894 |
Sumario: | BACKGROUND: Transcatheter aortic valve-in-valve (VIV) procedure is a safe alternative to conventional reoperation for bioprosthetic dysfunction. Balloon-expandable valve (BEV) and self-expanding valve (SEV) are the 2 major types of devices used. Evidence regarding the comparison of the 2 valves remains scarce. METHODS: A systematic review and meta-analysis was conducted to compare the outcomes of BEV and SEV in transcatheter VIV for aortic bioprostheses dysfunction. A computerized search of Medline, PubMed, Embase, and Cochrane databases was performed. English-language journal articles reporting SEV or BEV outcomes of at least 10 patients were included. RESULTS: In total, 27 studies were included, with 2,269 and 1,671 patients in the BEV and SEV groups, respectively. Rates of 30-day mortality and stroke did not differ significantly between the 2 groups. However, BEV was associated with significantly lower rates of postprocedural permanent pacemaker implantation (3.8% vs. 12%; P < 0.001). Regarding echocardiographic parameters, SEV was associated with larger postprocedural effective orifice area at 30 days (1.53 cm(2) vs. 1.23 cm(2); P < 0.001) and 1 year (1.55 cm(2) vs. 1.22 cm(2); P < 0.001). CONCLUSIONS: For patients who underwent transcatheter aortic VIV, SEV was associated with larger postprocedural effective orifice area but higher rates of permanent pacemaker implantation. These findings provide valuable information for optimizing device selection for transcatheter aortic VIV. |
---|