A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population

INTRODUCTION: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-fr...

Descripción completa

Detalles Bibliográficos
Autores principales: Singh, Shashi B., Kumar, Dewesh, Kashyap, Vivek, Singh, Surendra
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7266193/
https://www.ncbi.nlm.nih.gov/pubmed/32509634
http://dx.doi.org/10.4103/jfmpc.jfmpc_796_19
_version_ 1783541256361607168
author Singh, Shashi B.
Kumar, Dewesh
Kashyap, Vivek
Singh, Surendra
author_facet Singh, Shashi B.
Kumar, Dewesh
Kashyap, Vivek
Singh, Surendra
author_sort Singh, Shashi B.
collection PubMed
description INTRODUCTION: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-free LED BP devices. MATERIALS AND METHODS: This cross-sectional study was conducted in the tribal state of Jharkhand in India from January 2017 to June 2017. A total of 300 study participants aged more than 18 years were enrolled in this study. BP of the patients in the sitting position was measured three times each by automated device and mercury-free LED BP device. The different sets of readings were assessed by concordance correlation coefficient (CCC) and other validity measures. RESULTS: The CCC for systolic and diastolic BP measured by automated and mercury-free LED BP is 0.88 and 0.85, respectively. The mean difference between systolic and diastolic BP by both the instruments is statistically insignificant (P > 0.05). The sensitivity, specificity, PPV, NPV, and accuracy of automated BP devices to predict hypertension is 96.61%, 92.21%, 75%, 99%, and 93%. The area under ROC for systolic and diastolic BP is 0.984 and 0.97, respectively with P values < 0.0001 in both the cases. CONCLUSIONS: This study concluded that the overall automated BP machine has fair degree of agreement (CCC) with a manual BP device. The validity of this monitor to screen hypertension may also be considered in field settings.
format Online
Article
Text
id pubmed-7266193
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-72661932020-06-04 A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population Singh, Shashi B. Kumar, Dewesh Kashyap, Vivek Singh, Surendra J Family Med Prim Care Original Article INTRODUCTION: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-free LED BP devices. MATERIALS AND METHODS: This cross-sectional study was conducted in the tribal state of Jharkhand in India from January 2017 to June 2017. A total of 300 study participants aged more than 18 years were enrolled in this study. BP of the patients in the sitting position was measured three times each by automated device and mercury-free LED BP device. The different sets of readings were assessed by concordance correlation coefficient (CCC) and other validity measures. RESULTS: The CCC for systolic and diastolic BP measured by automated and mercury-free LED BP is 0.88 and 0.85, respectively. The mean difference between systolic and diastolic BP by both the instruments is statistically insignificant (P > 0.05). The sensitivity, specificity, PPV, NPV, and accuracy of automated BP devices to predict hypertension is 96.61%, 92.21%, 75%, 99%, and 93%. The area under ROC for systolic and diastolic BP is 0.984 and 0.97, respectively with P values < 0.0001 in both the cases. CONCLUSIONS: This study concluded that the overall automated BP machine has fair degree of agreement (CCC) with a manual BP device. The validity of this monitor to screen hypertension may also be considered in field settings. Wolters Kluwer - Medknow 2020-03-26 /pmc/articles/PMC7266193/ /pubmed/32509634 http://dx.doi.org/10.4103/jfmpc.jfmpc_796_19 Text en Copyright: © 2020 Journal of Family Medicine and Primary Care http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Singh, Shashi B.
Kumar, Dewesh
Kashyap, Vivek
Singh, Surendra
A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_full A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_fullStr A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_full_unstemmed A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_short A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_sort comparative study of automated blood pressure device and mercury-free led blood pressure device using lin's concordance correlation coefficient and other validity measures in indian population
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7266193/
https://www.ncbi.nlm.nih.gov/pubmed/32509634
http://dx.doi.org/10.4103/jfmpc.jfmpc_796_19
work_keys_str_mv AT singhshashib acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT kumardewesh acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT kashyapvivek acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT singhsurendra acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT singhshashib comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT kumardewesh comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT kashyapvivek comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT singhsurendra comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation