Cargando…
Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
BACKGROUND: Cross-linked polyethylene (XLPE) liners have shown lower wear rates than conventional polyethylene liners in total hip arthroplasty. The primary aim of our study was to report our most recent analysis of wear rates and clinical outcomes of a third-generation XLPE liner. Secondary aims we...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7267711/ https://www.ncbi.nlm.nih.gov/pubmed/32514421 http://dx.doi.org/10.1016/j.artd.2020.04.006 |
Sumario: | BACKGROUND: Cross-linked polyethylene (XLPE) liners have shown lower wear rates than conventional polyethylene liners in total hip arthroplasty. The primary aim of our study was to report our most recent analysis of wear rates and clinical outcomes of a third-generation XLPE liner. Secondary aims were to investigate the rate of adverse events related to mechanical failure or oxidation of this liner. METHODS: A series of 266 total hip arthroplasties using a specific XLPE liner were retrospectively reviewed. Radiographs were examined to determine linear and volumetric wear rates and presence of osteolysis. Clinical outcomes, revision rates, mechanical failures, and risk factors for accelerated polyethylene wear were additionally investigated. RESULTS: The mean age at the time of surgery was 65.8 years and the mean follow-up was 5.5 years. The mean linear wear rate was 0.003 mm/year and the mean volumetric wear rate was 0.42 mm(3)/year, and there was no evidence of osteolysis. Harris hip scores increased from 50.9 preoperatively to 96.0 at the latest follow-up. The revision rate was 0.4%, with no liner rim fractures and no liner dissociations/loosenings. Femoral head material, head size, age, body mass index, and time since implantation had no effect on wear rates. CONCLUSION: Wear rates for this third-generation XLPE liner were low at mid-term follow-up, and no adverse sequelae of oxidation or deleterious mechanical properties were observed. This remained true regardless of femoral head size and material or patient age and body mass index. Further analysis will be necessary to ensure continued wear resistance, oxidative stability, and mechanical strength at long-term follow-up. |
---|