Cargando…

Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty

BACKGROUND: Cross-linked polyethylene (XLPE) liners have shown lower wear rates than conventional polyethylene liners in total hip arthroplasty. The primary aim of our study was to report our most recent analysis of wear rates and clinical outcomes of a third-generation XLPE liner. Secondary aims we...

Descripción completa

Detalles Bibliográficos
Autores principales: Jalali, Omid, Scudday, Travis, Fickenscher, Marie-Claire, Barnett, Steven, Gorab, Robert
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7267711/
https://www.ncbi.nlm.nih.gov/pubmed/32514421
http://dx.doi.org/10.1016/j.artd.2020.04.006
_version_ 1783541460606386176
author Jalali, Omid
Scudday, Travis
Fickenscher, Marie-Claire
Barnett, Steven
Gorab, Robert
author_facet Jalali, Omid
Scudday, Travis
Fickenscher, Marie-Claire
Barnett, Steven
Gorab, Robert
author_sort Jalali, Omid
collection PubMed
description BACKGROUND: Cross-linked polyethylene (XLPE) liners have shown lower wear rates than conventional polyethylene liners in total hip arthroplasty. The primary aim of our study was to report our most recent analysis of wear rates and clinical outcomes of a third-generation XLPE liner. Secondary aims were to investigate the rate of adverse events related to mechanical failure or oxidation of this liner. METHODS: A series of 266 total hip arthroplasties using a specific XLPE liner were retrospectively reviewed. Radiographs were examined to determine linear and volumetric wear rates and presence of osteolysis. Clinical outcomes, revision rates, mechanical failures, and risk factors for accelerated polyethylene wear were additionally investigated. RESULTS: The mean age at the time of surgery was 65.8 years and the mean follow-up was 5.5 years. The mean linear wear rate was 0.003 mm/year and the mean volumetric wear rate was 0.42 mm(3)/year, and there was no evidence of osteolysis. Harris hip scores increased from 50.9 preoperatively to 96.0 at the latest follow-up. The revision rate was 0.4%, with no liner rim fractures and no liner dissociations/loosenings. Femoral head material, head size, age, body mass index, and time since implantation had no effect on wear rates. CONCLUSION: Wear rates for this third-generation XLPE liner were low at mid-term follow-up, and no adverse sequelae of oxidation or deleterious mechanical properties were observed. This remained true regardless of femoral head size and material or patient age and body mass index. Further analysis will be necessary to ensure continued wear resistance, oxidative stability, and mechanical strength at long-term follow-up.
format Online
Article
Text
id pubmed-7267711
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-72677112020-06-07 Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty Jalali, Omid Scudday, Travis Fickenscher, Marie-Claire Barnett, Steven Gorab, Robert Arthroplast Today Original Research BACKGROUND: Cross-linked polyethylene (XLPE) liners have shown lower wear rates than conventional polyethylene liners in total hip arthroplasty. The primary aim of our study was to report our most recent analysis of wear rates and clinical outcomes of a third-generation XLPE liner. Secondary aims were to investigate the rate of adverse events related to mechanical failure or oxidation of this liner. METHODS: A series of 266 total hip arthroplasties using a specific XLPE liner were retrospectively reviewed. Radiographs were examined to determine linear and volumetric wear rates and presence of osteolysis. Clinical outcomes, revision rates, mechanical failures, and risk factors for accelerated polyethylene wear were additionally investigated. RESULTS: The mean age at the time of surgery was 65.8 years and the mean follow-up was 5.5 years. The mean linear wear rate was 0.003 mm/year and the mean volumetric wear rate was 0.42 mm(3)/year, and there was no evidence of osteolysis. Harris hip scores increased from 50.9 preoperatively to 96.0 at the latest follow-up. The revision rate was 0.4%, with no liner rim fractures and no liner dissociations/loosenings. Femoral head material, head size, age, body mass index, and time since implantation had no effect on wear rates. CONCLUSION: Wear rates for this third-generation XLPE liner were low at mid-term follow-up, and no adverse sequelae of oxidation or deleterious mechanical properties were observed. This remained true regardless of femoral head size and material or patient age and body mass index. Further analysis will be necessary to ensure continued wear resistance, oxidative stability, and mechanical strength at long-term follow-up. Elsevier 2020-06-02 /pmc/articles/PMC7267711/ /pubmed/32514421 http://dx.doi.org/10.1016/j.artd.2020.04.006 Text en © 2020 Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research
Jalali, Omid
Scudday, Travis
Fickenscher, Marie-Claire
Barnett, Steven
Gorab, Robert
Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
title Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
title_full Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
title_fullStr Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
title_full_unstemmed Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
title_short Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty
title_sort third-generation medium cross-linked polyethylene demonstrates very low wear in total hip arthroplasty
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7267711/
https://www.ncbi.nlm.nih.gov/pubmed/32514421
http://dx.doi.org/10.1016/j.artd.2020.04.006
work_keys_str_mv AT jalaliomid thirdgenerationmediumcrosslinkedpolyethylenedemonstratesverylowwearintotalhiparthroplasty
AT scuddaytravis thirdgenerationmediumcrosslinkedpolyethylenedemonstratesverylowwearintotalhiparthroplasty
AT fickenschermarieclaire thirdgenerationmediumcrosslinkedpolyethylenedemonstratesverylowwearintotalhiparthroplasty
AT barnettsteven thirdgenerationmediumcrosslinkedpolyethylenedemonstratesverylowwearintotalhiparthroplasty
AT gorabrobert thirdgenerationmediumcrosslinkedpolyethylenedemonstratesverylowwearintotalhiparthroplasty