Patient-based quality control for glucometers: using the moving sum of positive patient results and moving average
INTRODUCTION: The capability of glucometer internal quality control (QC) in detecting varying magnitude of systematic error (bias), and the potential use of moving sum of positive results (MovSum) and moving average (MA) techniques as potential alternatives were evaluated. MATERIALS AND METHODS: The...
Autores principales: | Lim, Chun Yee, Badrick, Tony, Loh, Tze Ping |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7271757/ https://www.ncbi.nlm.nih.gov/pubmed/32550817 http://dx.doi.org/10.11613/BM.2020.020709 |
Ejemplares similares
-
Impact of combining data from multiple instruments on performance of patient-based real-time quality control
por: Zhou, Qianqian, et al.
Publicado: (2021) -
Use of Binary Cumulative Sums and Moving Averages in Nosocomial Infection Cluster Detection
por: Brown, Samuel M., et al.
Publicado: (2002) -
Exponentially weighted moving average—Moving average charts for monitoring the process mean
por: Sukparungsee, Saowanit, et al.
Publicado: (2020) -
Performance of new nonparametric Tukey modified exponentially weighted moving average—Moving average control chart
por: Talordphop, Khanittha, et al.
Publicado: (2022) -
Majorization involving the cyclic moving average
por: Zhang, Tao, et al.
Publicado: (2018)