The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets—Quantification and possible corrections
PURPOSE: To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D(98), target volume coverage (TVC), conformity (CI), and gradient index (GI). METHODS: Winston‐Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereot...
Autores principales: | Wack, Linda J., Exner, Florian, Wegener, Sonja, Sauer, Otto A. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7286018/ https://www.ncbi.nlm.nih.gov/pubmed/32196950 http://dx.doi.org/10.1002/acm2.12854 |
Ejemplares similares
-
Implementing corrections of isocentric shifts for the stereotactic irradiation of cerebral targets: Clinical validation
por: Wegener, Sonja, et al.
Publicado: (2022) -
Verification procedure for isocentric alignment of proton beams
por: Ciangaru, George, et al.
Publicado: (2007) -
Validating robotic couch isocentricity with 3D surface imaging
por: El‐Sherif, Omar, et al.
Publicado: (2020) -
Electrometer offset current due to scattered radiation
por: Wegener, Sonja, et al.
Publicado: (2018) -
Comparing the monoisocentric and dual isocentric techniques in chest wall radiotherapy of mastectomy patients
por: Banaei, Amin, et al.
Publicado: (2015)