Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging

BACKGROUND: Magnetic resonance (MR) black-blood thrombus imaging (BTI) is an accurate diagnostic technique for detecting deep vein thrombosis (DVT) but to date there have been no studies comparing the diagnostic performance and consistency of this technique at different field strengths. In this stud...

Descripción completa

Detalles Bibliográficos
Autores principales: Ye, Yufeng, He, Xueping, Huang, Chen, Shi, Caiyun, Deng, Wei, Luo, Wenfeng, Liang, Jianke, He, Zhuonan, Mao, Huan, Liang, Qiwei, Chen, Dongya, Chen, Hanwei, Xie, Guoxi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7288798/
https://www.ncbi.nlm.nih.gov/pubmed/32320276
http://dx.doi.org/10.1177/1076029620921235
_version_ 1783545339736752128
author Ye, Yufeng
He, Xueping
Huang, Chen
Shi, Caiyun
Deng, Wei
Luo, Wenfeng
Liang, Jianke
He, Zhuonan
Mao, Huan
Liang, Qiwei
Chen, Dongya
Chen, Hanwei
Xie, Guoxi
author_facet Ye, Yufeng
He, Xueping
Huang, Chen
Shi, Caiyun
Deng, Wei
Luo, Wenfeng
Liang, Jianke
He, Zhuonan
Mao, Huan
Liang, Qiwei
Chen, Dongya
Chen, Hanwei
Xie, Guoxi
author_sort Ye, Yufeng
collection PubMed
description BACKGROUND: Magnetic resonance (MR) black-blood thrombus imaging (BTI) is an accurate diagnostic technique for detecting deep vein thrombosis (DVT) but to date there have been no studies comparing the diagnostic performance and consistency of this technique at different field strengths. In this study, we evaluated and compared the diagnostic performance of BTI for detecting DVT at 1.5 T and 3.0 T field strengths. METHODS: A total of 40 patients with DVT were enrolled in this study from November 2015 up to October 2018. All patients underwent BTI, a contrast-free T1-weighted MR imaging technique for detecting DVT, and contrast-enhanced MR venography (CE-MRV) at 1.5 T or 3.0 T field strengths. The MR data analyses used 1160 segments from the venous lumen of the 40 patients. The signal-to-noise ratio and contrast-to-noise ratio between thrombus and muscle/lumen were calculated to compare BTI at 1.5 T or 3.0 T to determine the image performance for thrombus detection at 1.5 T or 3.0 T. Two physicians blinded to the study evaluated all BTI images and calculated the overall sensitivity (SE), specificity (SP), positive predictive value (PPV) and negative predictive value (NPV), accuracy, and diagnostic consistency at 1.5 T and 3.0 T. These images and values were compared to control CE-MRV images that had been obtained by 2 senior physicians and used as reference standards. In addition, the reliability and consistency of diagnoses between observers were also evaluated. RESULTS: Two study-blind physicians reviewed all BTI images to diagnose thrombus and to determine SE, SP, PPV, NPV, and accuracy. There were no statistical differences in SE, SP, PPV, NPV, or accuracy between the 1.5 T and 3.0 T groups. CONCLUSIONS: Black-blood thrombus imaging has high SE, SP, and accuracy for DVT diagnosis both at 1.5 T and 3.0 T field strengths. This noninvasive diagnostic technique, which does not require the use of contrast agents, can be widely used in the clinical screening of DVT and follow-up after treatment.
format Online
Article
Text
id pubmed-7288798
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-72887982020-06-19 Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging Ye, Yufeng He, Xueping Huang, Chen Shi, Caiyun Deng, Wei Luo, Wenfeng Liang, Jianke He, Zhuonan Mao, Huan Liang, Qiwei Chen, Dongya Chen, Hanwei Xie, Guoxi Clin Appl Thromb Hemost Original Article BACKGROUND: Magnetic resonance (MR) black-blood thrombus imaging (BTI) is an accurate diagnostic technique for detecting deep vein thrombosis (DVT) but to date there have been no studies comparing the diagnostic performance and consistency of this technique at different field strengths. In this study, we evaluated and compared the diagnostic performance of BTI for detecting DVT at 1.5 T and 3.0 T field strengths. METHODS: A total of 40 patients with DVT were enrolled in this study from November 2015 up to October 2018. All patients underwent BTI, a contrast-free T1-weighted MR imaging technique for detecting DVT, and contrast-enhanced MR venography (CE-MRV) at 1.5 T or 3.0 T field strengths. The MR data analyses used 1160 segments from the venous lumen of the 40 patients. The signal-to-noise ratio and contrast-to-noise ratio between thrombus and muscle/lumen were calculated to compare BTI at 1.5 T or 3.0 T to determine the image performance for thrombus detection at 1.5 T or 3.0 T. Two physicians blinded to the study evaluated all BTI images and calculated the overall sensitivity (SE), specificity (SP), positive predictive value (PPV) and negative predictive value (NPV), accuracy, and diagnostic consistency at 1.5 T and 3.0 T. These images and values were compared to control CE-MRV images that had been obtained by 2 senior physicians and used as reference standards. In addition, the reliability and consistency of diagnoses between observers were also evaluated. RESULTS: Two study-blind physicians reviewed all BTI images to diagnose thrombus and to determine SE, SP, PPV, NPV, and accuracy. There were no statistical differences in SE, SP, PPV, NPV, or accuracy between the 1.5 T and 3.0 T groups. CONCLUSIONS: Black-blood thrombus imaging has high SE, SP, and accuracy for DVT diagnosis both at 1.5 T and 3.0 T field strengths. This noninvasive diagnostic technique, which does not require the use of contrast agents, can be widely used in the clinical screening of DVT and follow-up after treatment. SAGE Publications 2020-04-22 /pmc/articles/PMC7288798/ /pubmed/32320276 http://dx.doi.org/10.1177/1076029620921235 Text en © The Author(s) 2020 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Article
Ye, Yufeng
He, Xueping
Huang, Chen
Shi, Caiyun
Deng, Wei
Luo, Wenfeng
Liang, Jianke
He, Zhuonan
Mao, Huan
Liang, Qiwei
Chen, Dongya
Chen, Hanwei
Xie, Guoxi
Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging
title Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging
title_full Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging
title_fullStr Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging
title_full_unstemmed Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging
title_short Comparison Between the Diagnostic Performance of 1.5 T and 3.0 T field Strengths for Detecting Deep Vein Thrombosis Using Magnetic Resonance Black-Blood Thrombus Imaging
title_sort comparison between the diagnostic performance of 1.5 t and 3.0 t field strengths for detecting deep vein thrombosis using magnetic resonance black-blood thrombus imaging
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7288798/
https://www.ncbi.nlm.nih.gov/pubmed/32320276
http://dx.doi.org/10.1177/1076029620921235
work_keys_str_mv AT yeyufeng comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT hexueping comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT huangchen comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT shicaiyun comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT dengwei comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT luowenfeng comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT liangjianke comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT hezhuonan comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT maohuan comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT liangqiwei comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT chendongya comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT chenhanwei comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging
AT xieguoxi comparisonbetweenthediagnosticperformanceof15tand30tfieldstrengthsfordetectingdeepveinthrombosisusingmagneticresonanceblackbloodthrombusimaging