Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas

PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. METHODS: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immers...

Descripción completa

Detalles Bibliográficos
Autores principales: Kontadakis, George A., Pennos, Alexandros, Pentari, Iro, Kymionis, George D., Pallikaris, Ioannis G., Ginis, Harilaos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7288812/
https://www.ncbi.nlm.nih.gov/pubmed/32577607
http://dx.doi.org/10.1177/2515841420923190
_version_ 1783545343025086464
author Kontadakis, George A.
Pennos, Alexandros
Pentari, Iro
Kymionis, George D.
Pallikaris, Ioannis G.
Ginis, Harilaos
author_facet Kontadakis, George A.
Pennos, Alexandros
Pentari, Iro
Kymionis, George D.
Pallikaris, Ioannis G.
Ginis, Harilaos
author_sort Kontadakis, George A.
collection PubMed
description PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. METHODS: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immersed in 35%, 40%, 50%, and 60% glycerin solutions for 3 hours. Subsequently, globes were mounted in a special holder, and their intraocular pressure was hydrostatically adjusted. Intraocular pressure was measured by means of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL while adjusting true intraocular pressure to 17, 33, and 50 mm Hg. Ultrasound pachymetry was performed. RESULTS: Mean corneal thickness was 914.5 ± 33.3 μm (730–1015 μm). In true intraocular pressure of 33 mm Hg, Goldmann applanation tonometry and dynamic contour tonometry significantly underestimated true intraocular pressure (mean Goldmann applanation tonometry: 14.7 ± 4.8 mm Hg, p < 0.001, mean dynamic contour tonometry: 21.6 ± 6.8, p < 0.001). Tono-Pen XL also underestimated, but difference was not statistically significant (Tono-Pen XL: 27.9 ± 9.7, p = 0.064). In true intraocular pressure of 50 mm Hg, all three methods significantly underestimated (Goldmann applanation tonometry: 17.6 ± 5.3 mm Hg, p < 0.001, dynamic contour tonometry: 26.8 ± 6.3 mm Hg, p < 0.001, Tono-Pen XL: 35.6 ± 8.4 mm Hg, p < 0.001). The error in measured intraocular pressure for each method (true minus measured intraocular pressure) was significantly correlated to true intraocular pressure (p < 0.001). The intraocular pressure measurements of each eye taken under true intraocular pressure of 17 and 33 mm Hg with the three methods were correlated to each other. Measurements taken under intraocular pressure of 50 mmHg were not correlated to each other. Corneal thickness was not correlated to intraocular pressure measurement. CONCLUSION: Goldmann applanation tonometry, dynamic contour tonometry, and Tono-Pen XL underestimate intraocular pressure when measured under edematous conditions. Tono-Pen XL showed better accuracy, especially in lower true intraocular pressure. The measurement error increases when true intraocular pressure increases in all three methods.
format Online
Article
Text
id pubmed-7288812
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-72888122020-06-22 Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas Kontadakis, George A. Pennos, Alexandros Pentari, Iro Kymionis, George D. Pallikaris, Ioannis G. Ginis, Harilaos Ther Adv Ophthalmol Original Research PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. METHODS: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immersed in 35%, 40%, 50%, and 60% glycerin solutions for 3 hours. Subsequently, globes were mounted in a special holder, and their intraocular pressure was hydrostatically adjusted. Intraocular pressure was measured by means of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL while adjusting true intraocular pressure to 17, 33, and 50 mm Hg. Ultrasound pachymetry was performed. RESULTS: Mean corneal thickness was 914.5 ± 33.3 μm (730–1015 μm). In true intraocular pressure of 33 mm Hg, Goldmann applanation tonometry and dynamic contour tonometry significantly underestimated true intraocular pressure (mean Goldmann applanation tonometry: 14.7 ± 4.8 mm Hg, p < 0.001, mean dynamic contour tonometry: 21.6 ± 6.8, p < 0.001). Tono-Pen XL also underestimated, but difference was not statistically significant (Tono-Pen XL: 27.9 ± 9.7, p = 0.064). In true intraocular pressure of 50 mm Hg, all three methods significantly underestimated (Goldmann applanation tonometry: 17.6 ± 5.3 mm Hg, p < 0.001, dynamic contour tonometry: 26.8 ± 6.3 mm Hg, p < 0.001, Tono-Pen XL: 35.6 ± 8.4 mm Hg, p < 0.001). The error in measured intraocular pressure for each method (true minus measured intraocular pressure) was significantly correlated to true intraocular pressure (p < 0.001). The intraocular pressure measurements of each eye taken under true intraocular pressure of 17 and 33 mm Hg with the three methods were correlated to each other. Measurements taken under intraocular pressure of 50 mmHg were not correlated to each other. Corneal thickness was not correlated to intraocular pressure measurement. CONCLUSION: Goldmann applanation tonometry, dynamic contour tonometry, and Tono-Pen XL underestimate intraocular pressure when measured under edematous conditions. Tono-Pen XL showed better accuracy, especially in lower true intraocular pressure. The measurement error increases when true intraocular pressure increases in all three methods. SAGE Publications 2020-06-10 /pmc/articles/PMC7288812/ /pubmed/32577607 http://dx.doi.org/10.1177/2515841420923190 Text en © The Author(s), 2020 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Research
Kontadakis, George A.
Pennos, Alexandros
Pentari, Iro
Kymionis, George D.
Pallikaris, Ioannis G.
Ginis, Harilaos
Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
title Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
title_full Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
title_fullStr Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
title_full_unstemmed Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
title_short Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
title_sort accuracy of dynamic contour tonometry, goldmann applanation tonometry, and tono-pen xl in edematous corneas
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7288812/
https://www.ncbi.nlm.nih.gov/pubmed/32577607
http://dx.doi.org/10.1177/2515841420923190
work_keys_str_mv AT kontadakisgeorgea accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas
AT pennosalexandros accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas
AT pentariiro accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas
AT kymionisgeorged accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas
AT pallikarisioannisg accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas
AT ginisharilaos accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas