Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. METHODS: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immers...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7288812/ https://www.ncbi.nlm.nih.gov/pubmed/32577607 http://dx.doi.org/10.1177/2515841420923190 |
_version_ | 1783545343025086464 |
---|---|
author | Kontadakis, George A. Pennos, Alexandros Pentari, Iro Kymionis, George D. Pallikaris, Ioannis G. Ginis, Harilaos |
author_facet | Kontadakis, George A. Pennos, Alexandros Pentari, Iro Kymionis, George D. Pallikaris, Ioannis G. Ginis, Harilaos |
author_sort | Kontadakis, George A. |
collection | PubMed |
description | PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. METHODS: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immersed in 35%, 40%, 50%, and 60% glycerin solutions for 3 hours. Subsequently, globes were mounted in a special holder, and their intraocular pressure was hydrostatically adjusted. Intraocular pressure was measured by means of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL while adjusting true intraocular pressure to 17, 33, and 50 mm Hg. Ultrasound pachymetry was performed. RESULTS: Mean corneal thickness was 914.5 ± 33.3 μm (730–1015 μm). In true intraocular pressure of 33 mm Hg, Goldmann applanation tonometry and dynamic contour tonometry significantly underestimated true intraocular pressure (mean Goldmann applanation tonometry: 14.7 ± 4.8 mm Hg, p < 0.001, mean dynamic contour tonometry: 21.6 ± 6.8, p < 0.001). Tono-Pen XL also underestimated, but difference was not statistically significant (Tono-Pen XL: 27.9 ± 9.7, p = 0.064). In true intraocular pressure of 50 mm Hg, all three methods significantly underestimated (Goldmann applanation tonometry: 17.6 ± 5.3 mm Hg, p < 0.001, dynamic contour tonometry: 26.8 ± 6.3 mm Hg, p < 0.001, Tono-Pen XL: 35.6 ± 8.4 mm Hg, p < 0.001). The error in measured intraocular pressure for each method (true minus measured intraocular pressure) was significantly correlated to true intraocular pressure (p < 0.001). The intraocular pressure measurements of each eye taken under true intraocular pressure of 17 and 33 mm Hg with the three methods were correlated to each other. Measurements taken under intraocular pressure of 50 mmHg were not correlated to each other. Corneal thickness was not correlated to intraocular pressure measurement. CONCLUSION: Goldmann applanation tonometry, dynamic contour tonometry, and Tono-Pen XL underestimate intraocular pressure when measured under edematous conditions. Tono-Pen XL showed better accuracy, especially in lower true intraocular pressure. The measurement error increases when true intraocular pressure increases in all three methods. |
format | Online Article Text |
id | pubmed-7288812 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-72888122020-06-22 Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas Kontadakis, George A. Pennos, Alexandros Pentari, Iro Kymionis, George D. Pallikaris, Ioannis G. Ginis, Harilaos Ther Adv Ophthalmol Original Research PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. METHODS: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immersed in 35%, 40%, 50%, and 60% glycerin solutions for 3 hours. Subsequently, globes were mounted in a special holder, and their intraocular pressure was hydrostatically adjusted. Intraocular pressure was measured by means of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL while adjusting true intraocular pressure to 17, 33, and 50 mm Hg. Ultrasound pachymetry was performed. RESULTS: Mean corneal thickness was 914.5 ± 33.3 μm (730–1015 μm). In true intraocular pressure of 33 mm Hg, Goldmann applanation tonometry and dynamic contour tonometry significantly underestimated true intraocular pressure (mean Goldmann applanation tonometry: 14.7 ± 4.8 mm Hg, p < 0.001, mean dynamic contour tonometry: 21.6 ± 6.8, p < 0.001). Tono-Pen XL also underestimated, but difference was not statistically significant (Tono-Pen XL: 27.9 ± 9.7, p = 0.064). In true intraocular pressure of 50 mm Hg, all three methods significantly underestimated (Goldmann applanation tonometry: 17.6 ± 5.3 mm Hg, p < 0.001, dynamic contour tonometry: 26.8 ± 6.3 mm Hg, p < 0.001, Tono-Pen XL: 35.6 ± 8.4 mm Hg, p < 0.001). The error in measured intraocular pressure for each method (true minus measured intraocular pressure) was significantly correlated to true intraocular pressure (p < 0.001). The intraocular pressure measurements of each eye taken under true intraocular pressure of 17 and 33 mm Hg with the three methods were correlated to each other. Measurements taken under intraocular pressure of 50 mmHg were not correlated to each other. Corneal thickness was not correlated to intraocular pressure measurement. CONCLUSION: Goldmann applanation tonometry, dynamic contour tonometry, and Tono-Pen XL underestimate intraocular pressure when measured under edematous conditions. Tono-Pen XL showed better accuracy, especially in lower true intraocular pressure. The measurement error increases when true intraocular pressure increases in all three methods. SAGE Publications 2020-06-10 /pmc/articles/PMC7288812/ /pubmed/32577607 http://dx.doi.org/10.1177/2515841420923190 Text en © The Author(s), 2020 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Research Kontadakis, George A. Pennos, Alexandros Pentari, Iro Kymionis, George D. Pallikaris, Ioannis G. Ginis, Harilaos Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas |
title | Accuracy of dynamic contour tonometry, Goldmann applanation
tonometry, and Tono-Pen XL in edematous corneas |
title_full | Accuracy of dynamic contour tonometry, Goldmann applanation
tonometry, and Tono-Pen XL in edematous corneas |
title_fullStr | Accuracy of dynamic contour tonometry, Goldmann applanation
tonometry, and Tono-Pen XL in edematous corneas |
title_full_unstemmed | Accuracy of dynamic contour tonometry, Goldmann applanation
tonometry, and Tono-Pen XL in edematous corneas |
title_short | Accuracy of dynamic contour tonometry, Goldmann applanation
tonometry, and Tono-Pen XL in edematous corneas |
title_sort | accuracy of dynamic contour tonometry, goldmann applanation
tonometry, and tono-pen xl in edematous corneas |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7288812/ https://www.ncbi.nlm.nih.gov/pubmed/32577607 http://dx.doi.org/10.1177/2515841420923190 |
work_keys_str_mv | AT kontadakisgeorgea accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas AT pennosalexandros accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas AT pentariiro accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas AT kymionisgeorged accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas AT pallikarisioannisg accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas AT ginisharilaos accuracyofdynamiccontourtonometrygoldmannapplanationtonometryandtonopenxlinedematouscorneas |