Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens

Spinal tumors and unstable vertebral body fractures usually require surgical treatment including vertebral body replacement. Regarding primary stability, however, the best possible treatment depends on the spinal region. The purpose of this in vitro study was to evaluate the effects of instrumentati...

Descripción completa

Detalles Bibliográficos
Autores principales: Liebsch, Christian, Kocak, Tugrul, Aleinikov, Viktor, Kerimbayev, Talgat, Akshulakov, Serik, Jansen, Jan Ulrich, Vogt, Morten, Wilke, Hans-Joachim
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7295896/
https://www.ncbi.nlm.nih.gov/pubmed/32582680
http://dx.doi.org/10.3389/fbioe.2020.00572
_version_ 1783546733211418624
author Liebsch, Christian
Kocak, Tugrul
Aleinikov, Viktor
Kerimbayev, Talgat
Akshulakov, Serik
Jansen, Jan Ulrich
Vogt, Morten
Wilke, Hans-Joachim
author_facet Liebsch, Christian
Kocak, Tugrul
Aleinikov, Viktor
Kerimbayev, Talgat
Akshulakov, Serik
Jansen, Jan Ulrich
Vogt, Morten
Wilke, Hans-Joachim
author_sort Liebsch, Christian
collection PubMed
description Spinal tumors and unstable vertebral body fractures usually require surgical treatment including vertebral body replacement. Regarding primary stability, however, the best possible treatment depends on the spinal region. The purpose of this in vitro study was to evaluate the effects of instrumentation length and approach size on thoracic spinal stability including the entire rib cage. Six fresh frozen human thoracic spine specimens with intact rib cages (C7-L1) were loaded with pure moments of 5 Nm in flexion/extension, lateral bending, and axial rotation, while monitoring the relative motions of all spinal segments using optical motion tracking. The specimens were tested (1) in the intact condition, followed by testing after vertebral body replacement at T6 level using a unilateral approach combined with (2) long instrumentation (T4–T8) and (3) short instrumentation (T5–T7) as well as a bilateral approach combined with (4) long and (5) short instrumentation. Significant increases of the range of motion (p < 0.05) were found in the entire thoracic spine (T1–T12) using the bilateral approach and short instrumentation in primary flexion/extension and in secondary axial rotation during primary lateral bending compared to both conditions with long instrumentation, as well as in secondary lateral bending during primary axial rotation compared to unilateral approach and long instrumentation. Compared to the intact condition, the range of motion was significantly decreased using unilateral approach and long instrumentation in flexion extension and secondary lateral bending during primary axial rotation, as well as using bilateral approach and long instrumentation in lateral bending. On the segmental level, the range of motion was significantly increased at T4–T5 level in lateral bending using unilateral approach and short instrumentation and significantly decreased using bilateral approach and long instrumentation compared to their respective previous conditions. Regardless of the approach type, which did not affect thoracic spinal stability in the present study, short instrumentation overall shows sufficient primary stability in the mid-thoracic spine with intact rib cage, while creating considerably more instability compared to long instrumentation, potentially being of importance regarding long-term implant failure. Moreover, short instrumentation could affect adjacent segment disease due to increased motion at the upper segmental level.
format Online
Article
Text
id pubmed-7295896
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-72958962020-06-23 Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens Liebsch, Christian Kocak, Tugrul Aleinikov, Viktor Kerimbayev, Talgat Akshulakov, Serik Jansen, Jan Ulrich Vogt, Morten Wilke, Hans-Joachim Front Bioeng Biotechnol Bioengineering and Biotechnology Spinal tumors and unstable vertebral body fractures usually require surgical treatment including vertebral body replacement. Regarding primary stability, however, the best possible treatment depends on the spinal region. The purpose of this in vitro study was to evaluate the effects of instrumentation length and approach size on thoracic spinal stability including the entire rib cage. Six fresh frozen human thoracic spine specimens with intact rib cages (C7-L1) were loaded with pure moments of 5 Nm in flexion/extension, lateral bending, and axial rotation, while monitoring the relative motions of all spinal segments using optical motion tracking. The specimens were tested (1) in the intact condition, followed by testing after vertebral body replacement at T6 level using a unilateral approach combined with (2) long instrumentation (T4–T8) and (3) short instrumentation (T5–T7) as well as a bilateral approach combined with (4) long and (5) short instrumentation. Significant increases of the range of motion (p < 0.05) were found in the entire thoracic spine (T1–T12) using the bilateral approach and short instrumentation in primary flexion/extension and in secondary axial rotation during primary lateral bending compared to both conditions with long instrumentation, as well as in secondary lateral bending during primary axial rotation compared to unilateral approach and long instrumentation. Compared to the intact condition, the range of motion was significantly decreased using unilateral approach and long instrumentation in flexion extension and secondary lateral bending during primary axial rotation, as well as using bilateral approach and long instrumentation in lateral bending. On the segmental level, the range of motion was significantly increased at T4–T5 level in lateral bending using unilateral approach and short instrumentation and significantly decreased using bilateral approach and long instrumentation compared to their respective previous conditions. Regardless of the approach type, which did not affect thoracic spinal stability in the present study, short instrumentation overall shows sufficient primary stability in the mid-thoracic spine with intact rib cage, while creating considerably more instability compared to long instrumentation, potentially being of importance regarding long-term implant failure. Moreover, short instrumentation could affect adjacent segment disease due to increased motion at the upper segmental level. Frontiers Media S.A. 2020-06-09 /pmc/articles/PMC7295896/ /pubmed/32582680 http://dx.doi.org/10.3389/fbioe.2020.00572 Text en Copyright © 2020 Liebsch, Kocak, Aleinikov, Kerimbayev, Akshulakov, Jansen, Vogt and Wilke. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Bioengineering and Biotechnology
Liebsch, Christian
Kocak, Tugrul
Aleinikov, Viktor
Kerimbayev, Talgat
Akshulakov, Serik
Jansen, Jan Ulrich
Vogt, Morten
Wilke, Hans-Joachim
Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens
title Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens
title_full Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens
title_fullStr Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens
title_full_unstemmed Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens
title_short Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens
title_sort thoracic spinal stability and motion behavior are affected by the length of posterior instrumentation after vertebral body replacement, but not by the surgical approach type: an in vitro study with entire rib cage specimens
topic Bioengineering and Biotechnology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7295896/
https://www.ncbi.nlm.nih.gov/pubmed/32582680
http://dx.doi.org/10.3389/fbioe.2020.00572
work_keys_str_mv AT liebschchristian thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT kocaktugrul thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT aleinikovviktor thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT kerimbayevtalgat thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT akshulakovserik thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT jansenjanulrich thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT vogtmorten thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens
AT wilkehansjoachim thoracicspinalstabilityandmotionbehaviorareaffectedbythelengthofposteriorinstrumentationaftervertebralbodyreplacementbutnotbythesurgicalapproachtypeaninvitrostudywithentireribcagespecimens