Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density

PURPOSE: This study was conducted evaluate the influence of reconstruction parameters of micro-computed tomography (micro-CT) images on bone mineral density (BMD) analyses. MATERIALS AND METHODS: The sample consisted of micro-CT images of the maxillae of 5 Wistar rats, acquired using a SkyScan 1174...

Descripción completa

Detalles Bibliográficos
Autores principales: Gaêta-Araujo, Hugo, Nascimento, Eduarda Helena Leandro, Brasil, Danieli Moura, Madlum, Daniela Verardi, Haiter-Neto, Francisco, Oliveira-Santos, Christiano
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Academy of Oral and Maxillofacial Radiology 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7314611/
https://www.ncbi.nlm.nih.gov/pubmed/32601590
http://dx.doi.org/10.5624/isd.2020.50.2.153
_version_ 1783550096570318848
author Gaêta-Araujo, Hugo
Nascimento, Eduarda Helena Leandro
Brasil, Danieli Moura
Madlum, Daniela Verardi
Haiter-Neto, Francisco
Oliveira-Santos, Christiano
author_facet Gaêta-Araujo, Hugo
Nascimento, Eduarda Helena Leandro
Brasil, Danieli Moura
Madlum, Daniela Verardi
Haiter-Neto, Francisco
Oliveira-Santos, Christiano
author_sort Gaêta-Araujo, Hugo
collection PubMed
description PURPOSE: This study was conducted evaluate the influence of reconstruction parameters of micro-computed tomography (micro-CT) images on bone mineral density (BMD) analyses. MATERIALS AND METHODS: The sample consisted of micro-CT images of the maxillae of 5 Wistar rats, acquired using a SkyScan 1174 unit (Bruker, Kontich, Belgium). Each acquisition was reconstructed following the manufacturer's recommendations (standard protocol; SP) for the application of artifact correction tools (beam hardening correction [BHC], 45%; smoothing filter, degree 2; and ring artifact correction [RAC], level 5). Additionally, images were reconstructed with 36 protocols combining different settings of artifact correction tools (P0 to P35). BMD analysis was performed for each reconstructed image. The BMD values obtained for each protocol were compared to those obtained using the SP through repeated-measures analysis of variance with the Dunnett post hoc test (α=0.05). RESULTS: The BMD values obtained from all protocols that used a BHC of 45% did not significantly differ from those obtained using the SP (P>0.05). The other protocols all yielded significantly different BMD values from the SP (P<0.05). The smoothing and RAC tools did not affect BMD values. CONCLUSION: BMD values measured on micro-CT images were influenced by the BHC level. Higher levels of BHC induced higher values of BMD.
format Online
Article
Text
id pubmed-7314611
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Korean Academy of Oral and Maxillofacial Radiology
record_format MEDLINE/PubMed
spelling pubmed-73146112020-06-28 Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density Gaêta-Araujo, Hugo Nascimento, Eduarda Helena Leandro Brasil, Danieli Moura Madlum, Daniela Verardi Haiter-Neto, Francisco Oliveira-Santos, Christiano Imaging Sci Dent Original Article PURPOSE: This study was conducted evaluate the influence of reconstruction parameters of micro-computed tomography (micro-CT) images on bone mineral density (BMD) analyses. MATERIALS AND METHODS: The sample consisted of micro-CT images of the maxillae of 5 Wistar rats, acquired using a SkyScan 1174 unit (Bruker, Kontich, Belgium). Each acquisition was reconstructed following the manufacturer's recommendations (standard protocol; SP) for the application of artifact correction tools (beam hardening correction [BHC], 45%; smoothing filter, degree 2; and ring artifact correction [RAC], level 5). Additionally, images were reconstructed with 36 protocols combining different settings of artifact correction tools (P0 to P35). BMD analysis was performed for each reconstructed image. The BMD values obtained for each protocol were compared to those obtained using the SP through repeated-measures analysis of variance with the Dunnett post hoc test (α=0.05). RESULTS: The BMD values obtained from all protocols that used a BHC of 45% did not significantly differ from those obtained using the SP (P>0.05). The other protocols all yielded significantly different BMD values from the SP (P<0.05). The smoothing and RAC tools did not affect BMD values. CONCLUSION: BMD values measured on micro-CT images were influenced by the BHC level. Higher levels of BHC induced higher values of BMD. Korean Academy of Oral and Maxillofacial Radiology 2020-06 2020-06-18 /pmc/articles/PMC7314611/ /pubmed/32601590 http://dx.doi.org/10.5624/isd.2020.50.2.153 Text en Copyright © 2020 by Korean Academy of Oral and Maxillofacial Radiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Gaêta-Araujo, Hugo
Nascimento, Eduarda Helena Leandro
Brasil, Danieli Moura
Madlum, Daniela Verardi
Haiter-Neto, Francisco
Oliveira-Santos, Christiano
Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
title Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
title_full Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
title_fullStr Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
title_full_unstemmed Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
title_short Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
title_sort influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7314611/
https://www.ncbi.nlm.nih.gov/pubmed/32601590
http://dx.doi.org/10.5624/isd.2020.50.2.153
work_keys_str_mv AT gaetaaraujohugo influenceofreconstructionparametersofmicrocomputedtomographyontheanalysisofbonemineraldensity
AT nascimentoeduardahelenaleandro influenceofreconstructionparametersofmicrocomputedtomographyontheanalysisofbonemineraldensity
AT brasildanielimoura influenceofreconstructionparametersofmicrocomputedtomographyontheanalysisofbonemineraldensity
AT madlumdanielaverardi influenceofreconstructionparametersofmicrocomputedtomographyontheanalysisofbonemineraldensity
AT haiternetofrancisco influenceofreconstructionparametersofmicrocomputedtomographyontheanalysisofbonemineraldensity
AT oliveirasantoschristiano influenceofreconstructionparametersofmicrocomputedtomographyontheanalysisofbonemineraldensity