Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard

BACKGROUND: The aim of this study is to conduct physical performance evaluation on the uMI550 whole-body PET/CT system according to the NEMA NU2-2018 standard. METHODS: According to the NEMA NU2-2018, spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Shuguang, Hu, Pengcheng, Gu, Yushen, Yu, Haojun, Shi, Hongcheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7316913/
https://www.ncbi.nlm.nih.gov/pubmed/32588139
http://dx.doi.org/10.1186/s40658-020-00315-w
Descripción
Sumario:BACKGROUND: The aim of this study is to conduct physical performance evaluation on the uMI550 whole-body PET/CT system according to the NEMA NU2-2018 standard. METHODS: According to the NEMA NU2-2018, spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and timing resolution were evaluated. Spatial resolution was measured by using a (22)Na point source. System sensitivity was measured by inserting an (18)F line-source in six concentric aluminum sleeves with varying diameters. Scatter fraction, count-rate performance, accuracy of count loss, and timing resolution were all calculated by analyzing dynamically acquired data of an (18)F line-source inside a polyethylene cylinder in 20 cm diameter and 70 cm length. Image quality was assessed using a NEMA IEC body phantom with a 4:1 ratio of activity concentration of spheres to the warm background. Additionally, three patient studies were performed, with one brain scan and two whole-body scans, separately. The patient images were evaluated to get a visual first impression of uMI550. RESULTS: The tangential, radial, and axial spatial resolutions were measured as 2.91 mm, 2.98 mm, and 2.97 mm FWHM, respectively, at 1 cm radial offset. The total system sensitivity to line source at center was 10.24 cps/kBq. A NECR peak was measured as 124.4 kcps at 18.85 kBq/mL. The scatter fraction at NECR peak was 36.65%, and the maximum count-rate error at and below NEC peak was 1.55%. Contrast recovery coefficients were from 46.5 (10 mm) to 83.9% (37 mm). The timing resolution was measured as 372 ps at low count rate. CONCLUSION: NEMA NU-2 2018 testing was performed on the new SiPM-based uMI550 PET/CT system. The uMI550 shows a high-spatial resolution of less than 3 mm and a good timing resolution of 372 ps. It shows clinical significances on improving potentially diagnostic ability on small lesions.