Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report

BACKGROUND: Percutaneous left atrial appendage (LAA) closure may reduce the risk of cardioembolic stroke in patients with non-valvular atrial fibrillation. Given the prophylactic nature of the procedure, identifying and managing complications are paramount. CASE SUMMARY: A 73-year-old man presented...

Descripción completa

Detalles Bibliográficos
Autores principales: Wilkins, Ben, Carranza, Christian L, Søndergaard, Lars, De Backer, Ole
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7319829/
https://www.ncbi.nlm.nih.gov/pubmed/32617484
http://dx.doi.org/10.1093/ehjcr/ytaa079
_version_ 1783551126631612416
author Wilkins, Ben
Carranza, Christian L
Søndergaard, Lars
De Backer, Ole
author_facet Wilkins, Ben
Carranza, Christian L
Søndergaard, Lars
De Backer, Ole
author_sort Wilkins, Ben
collection PubMed
description BACKGROUND: Percutaneous left atrial appendage (LAA) closure may reduce the risk of cardioembolic stroke in patients with non-valvular atrial fibrillation. Given the prophylactic nature of the procedure, identifying and managing complications are paramount. CASE SUMMARY: A 73-year-old man presented 14 months after percutaneous LAA closure with syncope and acute pericardial tamponade which required surgical exploration and haemostasis; the most temporally remote account of this complication albeit amongst very few case reports. Tissue erosion by the Amplatzer™ Amulet™ LAA closure device (Abbott, Plymouth, MN, USA) was noted at two separate anatomical locations, corresponding to the device disc and lobe, which has not been described previously. DISCUSSION: This case report highlights the anatomical relationship between the LAA and its surrounding structures, and the importance of recognizing the risk of late device erosion.
format Online
Article
Text
id pubmed-7319829
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-73198292020-07-01 Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report Wilkins, Ben Carranza, Christian L Søndergaard, Lars De Backer, Ole Eur Heart J Case Rep Case Reports BACKGROUND: Percutaneous left atrial appendage (LAA) closure may reduce the risk of cardioembolic stroke in patients with non-valvular atrial fibrillation. Given the prophylactic nature of the procedure, identifying and managing complications are paramount. CASE SUMMARY: A 73-year-old man presented 14 months after percutaneous LAA closure with syncope and acute pericardial tamponade which required surgical exploration and haemostasis; the most temporally remote account of this complication albeit amongst very few case reports. Tissue erosion by the Amplatzer™ Amulet™ LAA closure device (Abbott, Plymouth, MN, USA) was noted at two separate anatomical locations, corresponding to the device disc and lobe, which has not been described previously. DISCUSSION: This case report highlights the anatomical relationship between the LAA and its surrounding structures, and the importance of recognizing the risk of late device erosion. Oxford University Press 2020-05-03 /pmc/articles/PMC7319829/ /pubmed/32617484 http://dx.doi.org/10.1093/ehjcr/ytaa079 Text en © The Author(s) 2020. Published by Oxford University Press on behalf of the European Society of Cardiology. http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Case Reports
Wilkins, Ben
Carranza, Christian L
Søndergaard, Lars
De Backer, Ole
Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
title Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
title_full Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
title_fullStr Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
title_full_unstemmed Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
title_short Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
title_sort late presentation of left atrial appendage erosion and perforation by an amplatzer™ amulet™ closure device: a case report
topic Case Reports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7319829/
https://www.ncbi.nlm.nih.gov/pubmed/32617484
http://dx.doi.org/10.1093/ehjcr/ytaa079
work_keys_str_mv AT wilkinsben latepresentationofleftatrialappendageerosionandperforationbyanamplatzeramuletclosuredeviceacasereport
AT carranzachristianl latepresentationofleftatrialappendageerosionandperforationbyanamplatzeramuletclosuredeviceacasereport
AT søndergaardlars latepresentationofleftatrialappendageerosionandperforationbyanamplatzeramuletclosuredeviceacasereport
AT debackerole latepresentationofleftatrialappendageerosionandperforationbyanamplatzeramuletclosuredeviceacasereport