Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector

PURPOSE: This study investigated the use of high spatial resolution solid‐state detectors (DUO and Octa) combined with an inclinometer for machine‐based quality assurance (QA) of Volumetric Modulated Arc Therapy (VMAT) with flattened and flattening filter‐free beams. METHOD: The proposed system was...

Descripción completa

Detalles Bibliográficos
Autores principales: Matar, F. S., Wilkinson, D., Davis, J., Biasi, G., Causer, T., Fuduli, I., Brace, O., Stansook, N., Carolan, M., Rosenfeld, A. B., Petasecca, Marco
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7324694/
https://www.ncbi.nlm.nih.gov/pubmed/32277745
http://dx.doi.org/10.1002/acm2.12864
_version_ 1783551993238781952
author Matar, F. S.
Wilkinson, D.
Davis, J.
Biasi, G.
Causer, T.
Fuduli, I.
Brace, O.
Stansook, N.
Carolan, M.
Rosenfeld, A. B.
Petasecca, Marco
author_facet Matar, F. S.
Wilkinson, D.
Davis, J.
Biasi, G.
Causer, T.
Fuduli, I.
Brace, O.
Stansook, N.
Carolan, M.
Rosenfeld, A. B.
Petasecca, Marco
author_sort Matar, F. S.
collection PubMed
description PURPOSE: This study investigated the use of high spatial resolution solid‐state detectors (DUO and Octa) combined with an inclinometer for machine‐based quality assurance (QA) of Volumetric Modulated Arc Therapy (VMAT) with flattened and flattening filter‐free beams. METHOD: The proposed system was inserted in the accessory tray of the gantry head of a Varian 21iX Clinac and a Truebeam linear accelerator. Mutual dependence of the dose rate (DR) and gantry speed (GS) was assessed using the standard Varian customer acceptance plan (CAP). The multi‐leaf collimator (MLC) leaf speed was evaluated under static gantry conditions in directions parallel and orthogonal to gravity as well as under dynamic gantry conditions. Measurements were compared to machine log files. RESULTS: DR and GS as a function of gantry angle were reconstructed using the DUO/inclinometer and in agreement to within 1% with the machine log files in the sectors of constant DR and GS. The MLC leaf speeds agreed with the nominal speeds and those extracted from the machine log files to within 0.03 cm s(−1). The effect of gravity on the leaf motion was only observed when the leaves traveled faster than the nominal maximum velocity stated by the vendor. Under dynamic gantry conditions, MLC leaf speeds ranging between 0.33 and 1.42 cm s(−1) were evaluated. Comparing the average MLC leaf speeds with the machine log files found differences between 0.9% and 5.7%, with the largest discrepancy occurring under conditions of fastest leaf velocity, lowest DR and lowest detector signal. CONCLUSIONS: The investigation on the use of solid‐state detectors in combination with an inclinometer has demonstrated the capability to provide efficient and independent verification of DR, GS, and MLC leaf speed during dynamic VMAT delivery. Good agreement with machine log files suggests the detector/inclinometer system is a useful tool for machine‐specific VMAT QA.
format Online
Article
Text
id pubmed-7324694
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-73246942020-07-01 Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector Matar, F. S. Wilkinson, D. Davis, J. Biasi, G. Causer, T. Fuduli, I. Brace, O. Stansook, N. Carolan, M. Rosenfeld, A. B. Petasecca, Marco J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: This study investigated the use of high spatial resolution solid‐state detectors (DUO and Octa) combined with an inclinometer for machine‐based quality assurance (QA) of Volumetric Modulated Arc Therapy (VMAT) with flattened and flattening filter‐free beams. METHOD: The proposed system was inserted in the accessory tray of the gantry head of a Varian 21iX Clinac and a Truebeam linear accelerator. Mutual dependence of the dose rate (DR) and gantry speed (GS) was assessed using the standard Varian customer acceptance plan (CAP). The multi‐leaf collimator (MLC) leaf speed was evaluated under static gantry conditions in directions parallel and orthogonal to gravity as well as under dynamic gantry conditions. Measurements were compared to machine log files. RESULTS: DR and GS as a function of gantry angle were reconstructed using the DUO/inclinometer and in agreement to within 1% with the machine log files in the sectors of constant DR and GS. The MLC leaf speeds agreed with the nominal speeds and those extracted from the machine log files to within 0.03 cm s(−1). The effect of gravity on the leaf motion was only observed when the leaves traveled faster than the nominal maximum velocity stated by the vendor. Under dynamic gantry conditions, MLC leaf speeds ranging between 0.33 and 1.42 cm s(−1) were evaluated. Comparing the average MLC leaf speeds with the machine log files found differences between 0.9% and 5.7%, with the largest discrepancy occurring under conditions of fastest leaf velocity, lowest DR and lowest detector signal. CONCLUSIONS: The investigation on the use of solid‐state detectors in combination with an inclinometer has demonstrated the capability to provide efficient and independent verification of DR, GS, and MLC leaf speed during dynamic VMAT delivery. Good agreement with machine log files suggests the detector/inclinometer system is a useful tool for machine‐specific VMAT QA. John Wiley and Sons Inc. 2020-04-11 /pmc/articles/PMC7324694/ /pubmed/32277745 http://dx.doi.org/10.1002/acm2.12864 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Matar, F. S.
Wilkinson, D.
Davis, J.
Biasi, G.
Causer, T.
Fuduli, I.
Brace, O.
Stansook, N.
Carolan, M.
Rosenfeld, A. B.
Petasecca, Marco
Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector
title Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector
title_full Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector
title_fullStr Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector
title_full_unstemmed Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector
title_short Quality assurance of VMAT on flattened and flattening filter‐free accelerators using a high spatial resolution detector
title_sort quality assurance of vmat on flattened and flattening filter‐free accelerators using a high spatial resolution detector
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7324694/
https://www.ncbi.nlm.nih.gov/pubmed/32277745
http://dx.doi.org/10.1002/acm2.12864
work_keys_str_mv AT matarfs qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT wilkinsond qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT davisj qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT biasig qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT causert qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT fudulii qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT braceo qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT stansookn qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT carolanm qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT rosenfeldab qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector
AT petaseccamarco qualityassuranceofvmatonflattenedandflatteningfilterfreeacceleratorsusingahighspatialresolutiondetector