The potential of an optical surface tracking system in non‐coplanar single isocenter treatments of multiple brain metastases

To evaluate the accuracy of a commercial optical surface tracking (OST) system and to demonstrate how it can be implemented to monitor patient positioning during non‐coplanar single isocenter stereotactic treatments of brain metastases. A 3‐camera OST system was used (Catalyst HD™, C‐RAD) on a Trueb...

Descripción completa

Detalles Bibliográficos
Autores principales: Swinnen, Ans C. C., Öllers, Michel C., Loon Ong, Chin, Verhaegen, Frank
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7324699/
https://www.ncbi.nlm.nih.gov/pubmed/32237274
http://dx.doi.org/10.1002/acm2.12866
Descripción
Sumario:To evaluate the accuracy of a commercial optical surface tracking (OST) system and to demonstrate how it can be implemented to monitor patient positioning during non‐coplanar single isocenter stereotactic treatments of brain metastases. A 3‐camera OST system was used (Catalyst HD™, C‐RAD) on a TruebeamSTx with a 6DoF couch. The setup accuracy and agreement between the OST system, and CBCT and kV‐MV imaging at couch angles 0° and 270°, respectively, were examined. Film measurements at 3 depths in the Rando‐Alderson phantom were performed using a single isocenter non‐coplanar VMAT plan containing 4 brain lesions. Setup of the phantom was performed with CBCT at couch 0° and subsequently monitored by OST at other couch angles. Setup data for 7 volunteers were collected to evaluate the accuracy and reproducibility of the OST system at couch angles 0°, 45°, 90°, 315°, and 270°. These results were also correlated to the couch rotation offsets obtained by a Winston‐Lutz (WL) test. The Rando‐Alderson phantom, as well as volunteers, were fixated using open face masks (Orfit). For repeated tests with the Rando‐Alderson phantom, deviations between rotational and translational isocenter corrections for CBCT and OST systems are always within 0.2° (pitch, roll, yaw), and 0.1mm and 0.5mm (longitudinal, lateral, vertical) for couch positions 0° and 270°, respectively. Dose deviations between the film and TPS doses in the center of the 4 lesions were −1.2%, −0.1%, −0.0%, and −1.9%. Local gamma evaluation criteria of 2%/2 mm and 3%/1 mm yielded pass rates of 99.2%, 99.2%, 98.6%, 89.9% and 98.8%, 97.5%, 81.7%, 78.1% for the 4 lesions. Regarding the volunteers, the mean translational and rotational isocenter shift values were (0.24 ± 0.09) mm and (0.15 ± 0.07) degrees. Largest isocenter shifts were found for couch angles 45˚ and 90˚, confirmed by WL couch rotation offsets. Patient monitoring during non‐coplanar VMAT treatments of brain metastases is feasible with submillimeter accuracy.