The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations

PURPOSE: This study focused on determining risks from stereotactic radiotherapy using flattening filter‐free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radios...

Descripción completa

Detalles Bibliográficos
Autores principales: Aslian, Hossein, Kron, Tomas, Watts, Troy, Akalanli, Cagla, Hardcastle, Nicholas, Lonski, Peta, Montaseri, Atousa, Hay, Barry, Korte, James, Berk, Kemal, Longo, Francesco, Severgnini, Mara
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7324704/
https://www.ncbi.nlm.nih.gov/pubmed/32277741
http://dx.doi.org/10.1002/acm2.12873
_version_ 1783551996190523392
author Aslian, Hossein
Kron, Tomas
Watts, Troy
Akalanli, Cagla
Hardcastle, Nicholas
Lonski, Peta
Montaseri, Atousa
Hay, Barry
Korte, James
Berk, Kemal
Longo, Francesco
Severgnini, Mara
author_facet Aslian, Hossein
Kron, Tomas
Watts, Troy
Akalanli, Cagla
Hardcastle, Nicholas
Lonski, Peta
Montaseri, Atousa
Hay, Barry
Korte, James
Berk, Kemal
Longo, Francesco
Severgnini, Mara
author_sort Aslian, Hossein
collection PubMed
description PURPOSE: This study focused on determining risks from stereotactic radiotherapy using flattening filter‐free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs. METHODS: A retrospective review was performed. Subsequently, a phantom study was performed using 30 fully functional explanted CIEDs from two different manufacturers. Irradiation was carried out in a slab phantom with 6‐MV and 10‐MV FFF beams. First, a repetition‐rate test (RRT) with a range of beam pulse frequencies was conducted. Then, multifraction SBRT (48 Gy/4 Fx) and single‐fraction SBRT (28 Gy/1 Fx) treatment plans were used for lung tumors delivered to the phantom. RESULTS: Between 2014 and 2018, 13 cases were treated with an FFF beam (6 MV, 1400 MU/min or 10 MV, 2400 MU/min), and 15 cases were treated with a flattening filter (FF) beam (6 MV, 600 MU/min). All the devices were positioned outside the treatment field at a distance of more than 5 cm, except for one case, and no failures were reported due to SBRT/SRS. In the phantom rep‐rate tests, inappropriate sensing occurred, starting at a rep‐rate of 1200 MU/min. Cardiac implantable electronic device anomalies during and after delivering VMAT‐SBRT with a 10‐MV FFF beam were observed. CONCLUSIONS: The study showed that caution should be paid to managing CIED patients when they undergo SBRT using FFF beams, as it is recommended by AAPM TG‐203. Correspondingly, it was found that for FFF beams although there is small risk from dose‐rate effects, delivering high dose of radiation with beam energy greater than 6 MV and high‐dose rate to CIEDs positioned in close vicinity of the PTV may present issues.
format Online
Article
Text
id pubmed-7324704
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-73247042020-07-01 The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations Aslian, Hossein Kron, Tomas Watts, Troy Akalanli, Cagla Hardcastle, Nicholas Lonski, Peta Montaseri, Atousa Hay, Barry Korte, James Berk, Kemal Longo, Francesco Severgnini, Mara J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: This study focused on determining risks from stereotactic radiotherapy using flattening filter‐free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs. METHODS: A retrospective review was performed. Subsequently, a phantom study was performed using 30 fully functional explanted CIEDs from two different manufacturers. Irradiation was carried out in a slab phantom with 6‐MV and 10‐MV FFF beams. First, a repetition‐rate test (RRT) with a range of beam pulse frequencies was conducted. Then, multifraction SBRT (48 Gy/4 Fx) and single‐fraction SBRT (28 Gy/1 Fx) treatment plans were used for lung tumors delivered to the phantom. RESULTS: Between 2014 and 2018, 13 cases were treated with an FFF beam (6 MV, 1400 MU/min or 10 MV, 2400 MU/min), and 15 cases were treated with a flattening filter (FF) beam (6 MV, 600 MU/min). All the devices were positioned outside the treatment field at a distance of more than 5 cm, except for one case, and no failures were reported due to SBRT/SRS. In the phantom rep‐rate tests, inappropriate sensing occurred, starting at a rep‐rate of 1200 MU/min. Cardiac implantable electronic device anomalies during and after delivering VMAT‐SBRT with a 10‐MV FFF beam were observed. CONCLUSIONS: The study showed that caution should be paid to managing CIED patients when they undergo SBRT using FFF beams, as it is recommended by AAPM TG‐203. Correspondingly, it was found that for FFF beams although there is small risk from dose‐rate effects, delivering high dose of radiation with beam energy greater than 6 MV and high‐dose rate to CIEDs positioned in close vicinity of the PTV may present issues. John Wiley and Sons Inc. 2020-04-11 /pmc/articles/PMC7324704/ /pubmed/32277741 http://dx.doi.org/10.1002/acm2.12873 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Aslian, Hossein
Kron, Tomas
Watts, Troy
Akalanli, Cagla
Hardcastle, Nicholas
Lonski, Peta
Montaseri, Atousa
Hay, Barry
Korte, James
Berk, Kemal
Longo, Francesco
Severgnini, Mara
The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
title The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
title_full The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
title_fullStr The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
title_full_unstemmed The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
title_short The effect of stereotactic body radiotherapy (SBRT) using flattening filter‐free beams on cardiac implantable electronic devices (CIEDs) in clinical situations
title_sort effect of stereotactic body radiotherapy (sbrt) using flattening filter‐free beams on cardiac implantable electronic devices (cieds) in clinical situations
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7324704/
https://www.ncbi.nlm.nih.gov/pubmed/32277741
http://dx.doi.org/10.1002/acm2.12873
work_keys_str_mv AT aslianhossein theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT krontomas theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT wattstroy theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT akalanlicagla theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT hardcastlenicholas theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT lonskipeta theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT montaseriatousa theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT haybarry theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT kortejames theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT berkkemal theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT longofrancesco theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT severgninimara theeffectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT aslianhossein effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT krontomas effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT wattstroy effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT akalanlicagla effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT hardcastlenicholas effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT lonskipeta effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT montaseriatousa effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT haybarry effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT kortejames effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT berkkemal effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT longofrancesco effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations
AT severgninimara effectofstereotacticbodyradiotherapysbrtusingflatteningfilterfreebeamsoncardiacimplantableelectronicdevicesciedsinclinicalsituations