Mechanically Assisted Crevice Corrosion in a Metal-on-Polyethylene Total Hip Presenting With Lower Extremity Vascular Compromise

Mechanically assisted crevice corrosion in modular total hip replacements may lead to an adverse local tissue reaction (ALTR) with a variety of sequelae. Although an ALTR is most commonly recognized with metal-on-metal modular hip constructs, tribocorrosion at the head-neck junction of metal-on-poly...

Descripción completa

Detalles Bibliográficos
Autores principales: Zink, Thomas M., McGrory, Brian J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7327377/
https://www.ncbi.nlm.nih.gov/pubmed/32637514
http://dx.doi.org/10.1016/j.artd.2020.04.015
Descripción
Sumario:Mechanically assisted crevice corrosion in modular total hip replacements may lead to an adverse local tissue reaction (ALTR) with a variety of sequelae. Although an ALTR is most commonly recognized with metal-on-metal modular hip constructs, tribocorrosion at the head-neck junction of metal-on-polyethylene (MoP) total hip arthroplasties may also lead to an ALTR. We present a case of a 79-year-old woman with a history of MoP total hip arthroplasty who presented with unilateral leg swelling, joint pain, and stiffness and subsequently underwent revision for an ALTR secondary to mechanically assisted crevice corrosion. This unique case of lower extremity vascular compromise resulting from an ALTR is important because clinicians should consider corrosion-related ALTRs when treating patients with an MoP hip prosthesis presenting with new-onset lower extremity swelling.