Surveillance Accuracy of Smartphone-Dependent Glucose Meters in the Measurement of Plasma Glucose

BACKGROUND: Along with stand-alone glucose monitors, smartphone-dependent glucose meters (SDGM) are available for self-monitoring of blood glucose (SMBG). These display-less devices, connected to a smartphone, measure glucose and display report on a smartphone screen, which can be easily shared with...

Descripción completa

Detalles Bibliográficos
Autores principales: Mondal, Shaikat, Mondal, Himel, Biri, Sairavi Kiran
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7333746/
https://www.ncbi.nlm.nih.gov/pubmed/32699787
http://dx.doi.org/10.4103/ijem.IJEM_580_19
Descripción
Sumario:BACKGROUND: Along with stand-alone glucose monitors, smartphone-dependent glucose meters (SDGM) are available for self-monitoring of blood glucose (SMBG). These display-less devices, connected to a smartphone, measure glucose and display report on a smartphone screen, which can be easily shared with doctors. AIM: The aim of this study was to check the accuracy of the available SDGMs in relation to the laboratory reference method. MATERIALS AND METHODS: Two available SDGMs: BeatO™ and GlucoMe™ were tested. A total of 104 venous blood samples were tested for glucose in SDGMs and by the reference method. The accuracy of the report was checked by the International Standards Organization (ISO), American Diabetes Association (ADA), and Clinical Laboratory Improvement Amendments (CLIA) criteria. Clarke error grid analysis (CEGA) was carried out to find the suitability of SDGM in clinical usage. RESULTS: Mean plasma glucose (n = 104) level from laboratory report was 106.48 ± 44.58 mg/dL, that from BeatO™ was 105.78 ± 52.6, and GlucoMe™ was 99.72 ± 51.22. The accuracy was lower than that recommended by ISO 2013 (29.8% and 30.77% error in BeatO™ and GlucoMe™, respectively), ADA (63.46% and 74% error in BeatO™ and GlucoMe™, respectively), and CLIA (42.31% and 46.15% error in BeatO™ and GlucoMe™, respectively) criteria. According to CEGA, BeatO™ and GlucoMe™ can guide correct treatment in diabetes mellitus patients for 74% and 71% of measurement, respectively. CONCLUSION: The accuracy of both the SDGMs was lower than that suggested by ISO, ADA, and CLIA criteria. The result of this study would help patients and doctors in informed choice for the procurement of glucose monitors.