实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用
OBJECTIVE: To set internal quality control system of BCR-ABL (P210) transcript levels for real-time quantitative PCR (RQ-PCR). METHODS: Using K562 cells and HL-60 cells, we prepared high- and low-level BCR-ABL internal quality control substance. The BCR-ABL (P210) transcript levels of internal quali...
Formato: | Online Artículo Texto |
---|---|
Lenguaje: | English |
Publicado: |
Editorial office of Chinese Journal of Hematology
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7342103/ https://www.ncbi.nlm.nih.gov/pubmed/27719725 http://dx.doi.org/10.3760/cma.j.issn.0253-2727.2016.09.014 |
_version_ | 1783555365611241472 |
---|---|
collection | PubMed |
description | OBJECTIVE: To set internal quality control system of BCR-ABL (P210) transcript levels for real-time quantitative PCR (RQ-PCR). METHODS: Using K562 cells and HL-60 cells, we prepared high- and low-level BCR-ABL internal quality control substance. The BCR-ABL (P210) transcript levels of internal quality control substance have been determined for 184 times together with clinical samples from August 2013 to October 2015. The slope rate, intercept and correlation coefficient of standard curve were calculated according to different reagent lots (lots number 20130303, 20131212, 20140411 and 20150327 are called R1、R2、R3 and R4 for short respectively), and the detection results of quality control substance were calculated according to different reagent lots and quality control substance lots (lots number 20130725, 20140611 are called Q1、Q2 for short respectively). Then the results were analyzed by Levey-Jennings quality control chart combined with Westgard multi-rules theory. RESULTS: ①We analyzed the slope rate and intercept of standard curve. Fifty-three times of the R1 reagent detection, 80 times of the R3 reagent detection and 14 times of the R4 reagent detection were all under control. For 37 times detection of R2 reagent, the slope rate was out of control for 6 times. It was lower than x−s for the 2–8 tests and upper the average for the 12–37 tests. The intercept was out of control for 9 times, upper the x+s for the 1–8 tests and lower the average for the 12–37 tests. ②According to the detection results of quality control substance, for Q1 quality control substance, 49 tests by R1 reagent were under control, and 1 out of 23 tests by R2 reagent was out of control. For Q2 quality control substance, 14 tests by R2 reagent detection, 72 tests by R3 reagent detection and 14 tests by R4 reagent were all under control. CONCLUSION: The preparation of high- and low-level quality control substance using K562 and HL-60 cells was convenient and the detection results were reliable and stable. The application of quality control substance combined with slope rate and intercept in the internal quality control may contribute to quality assurance for quantitative detection of BCR-ABL (P210) transcript levels. |
format | Online Article Text |
id | pubmed-7342103 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Editorial office of Chinese Journal of Hematology |
record_format | MEDLINE/PubMed |
spelling | pubmed-73421032020-07-16 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 Zhonghua Xue Ye Xue Za Zhi 论著 OBJECTIVE: To set internal quality control system of BCR-ABL (P210) transcript levels for real-time quantitative PCR (RQ-PCR). METHODS: Using K562 cells and HL-60 cells, we prepared high- and low-level BCR-ABL internal quality control substance. The BCR-ABL (P210) transcript levels of internal quality control substance have been determined for 184 times together with clinical samples from August 2013 to October 2015. The slope rate, intercept and correlation coefficient of standard curve were calculated according to different reagent lots (lots number 20130303, 20131212, 20140411 and 20150327 are called R1、R2、R3 and R4 for short respectively), and the detection results of quality control substance were calculated according to different reagent lots and quality control substance lots (lots number 20130725, 20140611 are called Q1、Q2 for short respectively). Then the results were analyzed by Levey-Jennings quality control chart combined with Westgard multi-rules theory. RESULTS: ①We analyzed the slope rate and intercept of standard curve. Fifty-three times of the R1 reagent detection, 80 times of the R3 reagent detection and 14 times of the R4 reagent detection were all under control. For 37 times detection of R2 reagent, the slope rate was out of control for 6 times. It was lower than x−s for the 2–8 tests and upper the average for the 12–37 tests. The intercept was out of control for 9 times, upper the x+s for the 1–8 tests and lower the average for the 12–37 tests. ②According to the detection results of quality control substance, for Q1 quality control substance, 49 tests by R1 reagent were under control, and 1 out of 23 tests by R2 reagent was out of control. For Q2 quality control substance, 14 tests by R2 reagent detection, 72 tests by R3 reagent detection and 14 tests by R4 reagent were all under control. CONCLUSION: The preparation of high- and low-level quality control substance using K562 and HL-60 cells was convenient and the detection results were reliable and stable. The application of quality control substance combined with slope rate and intercept in the internal quality control may contribute to quality assurance for quantitative detection of BCR-ABL (P210) transcript levels. Editorial office of Chinese Journal of Hematology 2016-09 /pmc/articles/PMC7342103/ /pubmed/27719725 http://dx.doi.org/10.3760/cma.j.issn.0253-2727.2016.09.014 Text en 2016年版权归中华医学会所有 http://creativecommons.org/licenses/by-nc-sa/3.0/ This work is licensed under a Creative Commons Attribution 3.0 License (CC-BY-NC). The Copyright own by Publisher. Without authorization, shall not reprint, except this publication article, shall not use this publication format design. Unless otherwise stated, all articles published in this journal do not represent the views of the Chinese Medical Association or the editorial board of this journal. |
spellingShingle | 论著 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 |
title | 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 |
title_full | 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 |
title_fullStr | 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 |
title_full_unstemmed | 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 |
title_short | 实时定量PCR检测BCR-ABL (P210)转录本水平室内长期质控体系的建立及应用 |
title_sort | 实时定量pcr检测bcr-abl (p210)转录本水平室内长期质控体系的建立及应用 |
topic | 论著 |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7342103/ https://www.ncbi.nlm.nih.gov/pubmed/27719725 http://dx.doi.org/10.3760/cma.j.issn.0253-2727.2016.09.014 |
work_keys_str_mv | AT shíshídìngliàngpcrjiǎncèbcrablp210zhuǎnlùběnshuǐpíngshìnèizhǎngqīzhìkòngtǐxìdejiànlìjíyīngyòng AT shíshídìngliàngpcrjiǎncèbcrablp210zhuǎnlùběnshuǐpíngshìnèizhǎngqīzhìkòngtǐxìdejiànlìjíyīngyòng AT shíshídìngliàngpcrjiǎncèbcrablp210zhuǎnlùběnshuǐpíngshìnèizhǎngqīzhìkòngtǐxìdejiànlìjíyīngyòng AT shíshídìngliàngpcrjiǎncèbcrablp210zhuǎnlùběnshuǐpíngshìnèizhǎngqīzhìkòngtǐxìdejiànlìjíyīngyòng AT shíshídìngliàngpcrjiǎncèbcrablp210zhuǎnlùběnshuǐpíngshìnèizhǎngqīzhìkòngtǐxìdejiànlìjíyīngyòng AT shíshídìngliàngpcrjiǎncèbcrablp210zhuǎnlùběnshuǐpíngshìnèizhǎngqīzhìkòngtǐxìdejiànlìjíyīngyòng |