Cargando…

Magnetic Solid-Phase Extraction Based on Poly 4-Vinyl Pyridine for HPLC-FLD Analysis of Naproxen in Urine Samples

A magnetic solid phase extraction technique followed by liquid chromatography with a fluorescence detector for naproxen analysis in human urine samples was developed. The method includes the extraction of naproxen with a magnetic solid synthetized with magnetite and poly 4-vinylpriridine, followed b...

Descripción completa

Detalles Bibliográficos
Autores principales: Escamilla-Lara, Karen A., Heredia, Ana C., Peña-Alvarez, Araceli, Ibarra, Israel S., Barrado, Enrique, Rodriguez, Jose A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7355941/
https://www.ncbi.nlm.nih.gov/pubmed/32630475
http://dx.doi.org/10.3390/molecules25122924
Descripción
Sumario:A magnetic solid phase extraction technique followed by liquid chromatography with a fluorescence detector for naproxen analysis in human urine samples was developed. The method includes the extraction of naproxen with a magnetic solid synthetized with magnetite and poly 4-vinylpriridine, followed by the magnetic separation of the solid phase and desorption of the analyte with methanol. Under optimal conditions, the linear range of the calibration curve was 0.05–0.60 μg L(−1), with a limit of detection of 0.02 μg L(−1). In all cases values of repeatability were lower than 5.0% with recoveries of 99.4 ± 1.3%. Precision and accuracy values are adequate for naproxen (Npx) analysis in urine samples.