The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions

INTRODUCTION: The use of activity trackers has increased both among private consumers and in healthcare. It is therefore relevant to consider whether a consumer-graded activity tracker is comparable to or may substitute a research-graded activity tracker, which could further increase the use of acti...

Descripción completa

Detalles Bibliográficos
Autores principales: Svarre, Frederik Rose, Jensen, Mads Møller, Nielsen, Josephine, Villumsen, Morten
Formato: Online Artículo Texto
Lenguaje:English
Publicado: PeerJ Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7367048/
https://www.ncbi.nlm.nih.gov/pubmed/32742766
http://dx.doi.org/10.7717/peerj.9381
_version_ 1783560347105361920
author Svarre, Frederik Rose
Jensen, Mads Møller
Nielsen, Josephine
Villumsen, Morten
author_facet Svarre, Frederik Rose
Jensen, Mads Møller
Nielsen, Josephine
Villumsen, Morten
author_sort Svarre, Frederik Rose
collection PubMed
description INTRODUCTION: The use of activity trackers has increased both among private consumers and in healthcare. It is therefore relevant to consider whether a consumer-graded activity tracker is comparable to or may substitute a research-graded activity tracker, which could further increase the use of activity trackers in healthcare and rehabilitation. Such use will require knowledge of their accuracy as the clinical implications may be significant. Studies have indicated that activity trackers are not sufficiently accurate, especially at lower walking speeds. The present study seeks to inform decision makers and healthcare personnel considering implementing physical activity trackers in clinical practice. This study investigates the criterion validity of the consumer-graded Garmin Vivosmart(®) HR and the research-graded StepWatch(™) 3 compared with manual step count (gold standard) at different walking speeds under controlled conditions. METHODS: Thirty participants, wearing Garmin Vivosmart(®) HR at the wrist and StepWatch(™) 3 at the ankle, completed six trials on a treadmill at different walking speeds: 1.6 km/h, 2.4 km/h, 3.2 km/h, 4.0 km/h, 4.8 km/h, and 5.6 km/h. The participants were video recorded, and steps were registered by manual step count. Medians and inter-quartile ranges (IQR) were calculated for steps and differences in steps between manually counted steps and the two devices. In order to assess the clinical relevance of the tested devices, the mean absolute percentage error (MAPE) was determined at each speed. A MAPE ≤3% was considered to be clinically irrelevant. Furthermore, differences between manually counted steps and steps recorded by the two devices were presented in Bland–Altman style plots. RESULTS: The median of differences in steps between Garmin Vivosmart(®) HR and manual step count ranged from −49.5 (IQR = 101) at 1.6 km/h to −1 (IQR = 4) at 4.0 km/h. The median of differences in steps between StepWatch(™) 3 and manual step count were 4 (IQR = 14) at 1.6 km/h and 0 (IQR = 1) at all other walking speeds. The results of the MAPE showed that differences in steps counted by Garmin Vivosmart(®) HR were clinically irrelevant at walking speeds 3.2–4.8 km/h (MAPE: 0.61–1.27%) as the values were below 3%. Differences in steps counted by StepWatch(™) 3 were clinically irrelevant at walking speeds 2.4–5.6 km/h (MAPE: 0.08–0.35%). CONCLUSION: Garmin Vivosmart(®) HR tended to undercount steps compared with the manual step count, and StepWatch(™) 3 slightly overcounted steps compared with the manual step count. Both the consumer-graded activity tracker (Garmin Vivosmart(®) HR) and the research-graded (StepWatch(™) 3) are valid in detecting steps at selected walking speeds in healthy adults under controlled conditions. However, both activity trackers miscount steps at slow walking speeds, and the consumer graded activity tracker also miscounts steps at fast walking speeds.
format Online
Article
Text
id pubmed-7367048
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher PeerJ Inc.
record_format MEDLINE/PubMed
spelling pubmed-73670482020-07-31 The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions Svarre, Frederik Rose Jensen, Mads Møller Nielsen, Josephine Villumsen, Morten PeerJ Kinesiology INTRODUCTION: The use of activity trackers has increased both among private consumers and in healthcare. It is therefore relevant to consider whether a consumer-graded activity tracker is comparable to or may substitute a research-graded activity tracker, which could further increase the use of activity trackers in healthcare and rehabilitation. Such use will require knowledge of their accuracy as the clinical implications may be significant. Studies have indicated that activity trackers are not sufficiently accurate, especially at lower walking speeds. The present study seeks to inform decision makers and healthcare personnel considering implementing physical activity trackers in clinical practice. This study investigates the criterion validity of the consumer-graded Garmin Vivosmart(®) HR and the research-graded StepWatch(™) 3 compared with manual step count (gold standard) at different walking speeds under controlled conditions. METHODS: Thirty participants, wearing Garmin Vivosmart(®) HR at the wrist and StepWatch(™) 3 at the ankle, completed six trials on a treadmill at different walking speeds: 1.6 km/h, 2.4 km/h, 3.2 km/h, 4.0 km/h, 4.8 km/h, and 5.6 km/h. The participants were video recorded, and steps were registered by manual step count. Medians and inter-quartile ranges (IQR) were calculated for steps and differences in steps between manually counted steps and the two devices. In order to assess the clinical relevance of the tested devices, the mean absolute percentage error (MAPE) was determined at each speed. A MAPE ≤3% was considered to be clinically irrelevant. Furthermore, differences between manually counted steps and steps recorded by the two devices were presented in Bland–Altman style plots. RESULTS: The median of differences in steps between Garmin Vivosmart(®) HR and manual step count ranged from −49.5 (IQR = 101) at 1.6 km/h to −1 (IQR = 4) at 4.0 km/h. The median of differences in steps between StepWatch(™) 3 and manual step count were 4 (IQR = 14) at 1.6 km/h and 0 (IQR = 1) at all other walking speeds. The results of the MAPE showed that differences in steps counted by Garmin Vivosmart(®) HR were clinically irrelevant at walking speeds 3.2–4.8 km/h (MAPE: 0.61–1.27%) as the values were below 3%. Differences in steps counted by StepWatch(™) 3 were clinically irrelevant at walking speeds 2.4–5.6 km/h (MAPE: 0.08–0.35%). CONCLUSION: Garmin Vivosmart(®) HR tended to undercount steps compared with the manual step count, and StepWatch(™) 3 slightly overcounted steps compared with the manual step count. Both the consumer-graded activity tracker (Garmin Vivosmart(®) HR) and the research-graded (StepWatch(™) 3) are valid in detecting steps at selected walking speeds in healthy adults under controlled conditions. However, both activity trackers miscount steps at slow walking speeds, and the consumer graded activity tracker also miscounts steps at fast walking speeds. PeerJ Inc. 2020-07-14 /pmc/articles/PMC7367048/ /pubmed/32742766 http://dx.doi.org/10.7717/peerj.9381 Text en © 2020 Svarre et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, reproduction and adaptation in any medium and for any purpose provided that it is properly attributed. For attribution, the original author(s), title, publication source (PeerJ) and either DOI or URL of the article must be cited.
spellingShingle Kinesiology
Svarre, Frederik Rose
Jensen, Mads Møller
Nielsen, Josephine
Villumsen, Morten
The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions
title The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions
title_full The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions
title_fullStr The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions
title_full_unstemmed The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions
title_short The validity of activity trackers is affected by walking speed: the criterion validity of Garmin Vivosmart(®) HR and StepWatch(™) 3 for measuring steps at various walking speeds under controlled conditions
title_sort validity of activity trackers is affected by walking speed: the criterion validity of garmin vivosmart(®) hr and stepwatch(™) 3 for measuring steps at various walking speeds under controlled conditions
topic Kinesiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7367048/
https://www.ncbi.nlm.nih.gov/pubmed/32742766
http://dx.doi.org/10.7717/peerj.9381
work_keys_str_mv AT svarrefrederikrose thevalidityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT jensenmadsmøller thevalidityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT nielsenjosephine thevalidityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT villumsenmorten thevalidityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT svarrefrederikrose validityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT jensenmadsmøller validityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT nielsenjosephine validityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions
AT villumsenmorten validityofactivitytrackersisaffectedbywalkingspeedthecriterionvalidityofgarminvivosmarthrandstepwatch3formeasuringstepsatvariouswalkingspeedsundercontrolledconditions