Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects

OBJECTIVE: This study is aimed at investigating if point-of-care testing for HbA1c (POCT-HbA1c) using the HemoCue® HbA1c 501 system could be an alternative method for diabetes screening and monitoring to replace the HbA1c measurement in a standard diagnostic laboratory. DESIGN: This was a cross-sect...

Descripción completa

Detalles Bibliográficos
Autores principales: Berbudi, Afiat, Rahmadika, Nofri, Tjahjadi, Adi Imam, Ruslami, Rovina
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7369652/
https://www.ncbi.nlm.nih.gov/pubmed/32733964
http://dx.doi.org/10.1155/2020/2037565
_version_ 1783560817505992704
author Berbudi, Afiat
Rahmadika, Nofri
Tjahjadi, Adi Imam
Ruslami, Rovina
author_facet Berbudi, Afiat
Rahmadika, Nofri
Tjahjadi, Adi Imam
Ruslami, Rovina
author_sort Berbudi, Afiat
collection PubMed
description OBJECTIVE: This study is aimed at investigating if point-of-care testing for HbA1c (POCT-HbA1c) using the HemoCue® HbA1c 501 system could be an alternative method for diabetes screening and monitoring to replace the HbA1c measurement in a standard diagnostic laboratory. DESIGN: This was a cross-sectional study to assess the agreement between POCT and a standard laboratory measurement method for determining the level of HbA1c. Setting and Participants. In total, 108 participants were recruited to participate in this study, consisting of 61 diabetics and 47 nondiabetics. The diabetic group comprised 37 females and 24 males, diagnosed with type 2 diabetes mellitus (DM) and undergoing diabetes treatment at several community health care centres in Bandung, West Java. The nondiabetic group consisted of 15 female and 32 male patients of several community health care centres and healthy volunteers. Sample Collection and Analysis. A venous blood sample was taken for routine HbA1c analysis by the diagnostic laboratory method. For the POCT-HbA1c, a blood sample was taken from the fingertip at the same time and analysed with the HemoCue® HbA1c 501 system. Outcome Measures. The HbA1c results of both methods were compared and analysed with a Bland-Altman agreement plot. The sensitivity and specificity of the POCT-HbA1c data were also compared with those of the standard diagnostic results. RESULTS: Based on the Bland-Altman plot, the HbA1c level for 100 out of 108 (92.59%) subjects analysed by the POCT-HbA1c was within the range of the 95% limit of agreement. Compared with the standard diagnostic assay, the sensitivity of the POCT-HbA1c was 97.83% and its specificity was 77.42%. CONCLUSIONS: The high sensitivity and accuracy of POCT-HbA1c indicate that it is a potential method for diabetes screening and monitoring to replace the routine diagnostic laboratory HbA1c measurement, especially when a rapid result is required.
format Online
Article
Text
id pubmed-7369652
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-73696522020-07-29 Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects Berbudi, Afiat Rahmadika, Nofri Tjahjadi, Adi Imam Ruslami, Rovina J Diabetes Res Research Article OBJECTIVE: This study is aimed at investigating if point-of-care testing for HbA1c (POCT-HbA1c) using the HemoCue® HbA1c 501 system could be an alternative method for diabetes screening and monitoring to replace the HbA1c measurement in a standard diagnostic laboratory. DESIGN: This was a cross-sectional study to assess the agreement between POCT and a standard laboratory measurement method for determining the level of HbA1c. Setting and Participants. In total, 108 participants were recruited to participate in this study, consisting of 61 diabetics and 47 nondiabetics. The diabetic group comprised 37 females and 24 males, diagnosed with type 2 diabetes mellitus (DM) and undergoing diabetes treatment at several community health care centres in Bandung, West Java. The nondiabetic group consisted of 15 female and 32 male patients of several community health care centres and healthy volunteers. Sample Collection and Analysis. A venous blood sample was taken for routine HbA1c analysis by the diagnostic laboratory method. For the POCT-HbA1c, a blood sample was taken from the fingertip at the same time and analysed with the HemoCue® HbA1c 501 system. Outcome Measures. The HbA1c results of both methods were compared and analysed with a Bland-Altman agreement plot. The sensitivity and specificity of the POCT-HbA1c data were also compared with those of the standard diagnostic results. RESULTS: Based on the Bland-Altman plot, the HbA1c level for 100 out of 108 (92.59%) subjects analysed by the POCT-HbA1c was within the range of the 95% limit of agreement. Compared with the standard diagnostic assay, the sensitivity of the POCT-HbA1c was 97.83% and its specificity was 77.42%. CONCLUSIONS: The high sensitivity and accuracy of POCT-HbA1c indicate that it is a potential method for diabetes screening and monitoring to replace the routine diagnostic laboratory HbA1c measurement, especially when a rapid result is required. Hindawi 2020-07-09 /pmc/articles/PMC7369652/ /pubmed/32733964 http://dx.doi.org/10.1155/2020/2037565 Text en Copyright © 2020 Afiat Berbudi et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Berbudi, Afiat
Rahmadika, Nofri
Tjahjadi, Adi Imam
Ruslami, Rovina
Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects
title Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects
title_full Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects
title_fullStr Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects
title_full_unstemmed Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects
title_short Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects
title_sort performance of point-of-care testing compared with the standard laboratory diagnostic test in the measurement of hba1c in indonesian diabetic and nondiabetic subjects
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7369652/
https://www.ncbi.nlm.nih.gov/pubmed/32733964
http://dx.doi.org/10.1155/2020/2037565
work_keys_str_mv AT berbudiafiat performanceofpointofcaretestingcomparedwiththestandardlaboratorydiagnostictestinthemeasurementofhba1cinindonesiandiabeticandnondiabeticsubjects
AT rahmadikanofri performanceofpointofcaretestingcomparedwiththestandardlaboratorydiagnostictestinthemeasurementofhba1cinindonesiandiabeticandnondiabeticsubjects
AT tjahjadiadiimam performanceofpointofcaretestingcomparedwiththestandardlaboratorydiagnostictestinthemeasurementofhba1cinindonesiandiabeticandnondiabeticsubjects
AT ruslamirovina performanceofpointofcaretestingcomparedwiththestandardlaboratorydiagnostictestinthemeasurementofhba1cinindonesiandiabeticandnondiabeticsubjects