Cargando…

Effect of Different Implant Designs on Strain and Stress Distribution under Non-Axial Loading: A Three-Dimensional Finite Element Analysis

Implant design evolved alongside the development of implant therapy. The purpose of this finite element analysis (FEA) study was to analyze the influence of different implant designs on the stress and strain distribution to the implants and surrounding bone. Three implant designs with the same lengt...

Descripción completa

Detalles Bibliográficos
Autores principales: Oliveira, Hélder, Brizuela Velasco, Aritza, Ríos-Santos, José-Vicente, Sánchez Lasheras, Fernando, Lemos, Bernardo Ferreira, Gil, Francisco Javier, Carvalho, Alexandrine, Herrero-Climent, Mariano
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7370002/
https://www.ncbi.nlm.nih.gov/pubmed/32630294
http://dx.doi.org/10.3390/ijerph17134738
Descripción
Sumario:Implant design evolved alongside the development of implant therapy. The purpose of this finite element analysis (FEA) study was to analyze the influence of different implant designs on the stress and strain distribution to the implants and surrounding bone. Three implant designs with the same length and diameter were used. The three-dimensional geometry of the bone was simulated with a cortical bone of three different thicknesses and two medullar bone densities: low density (150 Hounsfield units) and high density (850 Hounsfield units). A 30° oblique load of 150 N was applied to the implant restoration. Displacement and stress (von Mises) results were obtained for bone and dental implants. The strain and stress distributions to the bone were higher for the tissue-level implant for all types of bone. The maximum principal strain and stress decreased with an increase in cortical bone thickness for both cancellous bone densities. The distribution of the load was concentrated at the coronal portion of the bone and implants. All implants showed a good distribution of forces for non-axial loads, with higher forces concentrated at the crestal region of the bone–implant interface. Decrease in medullar bone density negatively affects the strain and stress produced by the implants.