Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery

BACKGROUND/AIM: To compare the clinical use, image quality and viewing angle of a commonly used contact wide angle viewing (WAV) system (Advanced Visual Instruments (AVI) Panoramic Imaging Systems, NY, USA) with a commonly used noncontact WAV system (Leica RUV800, Leica Microsystems, Switzerland). M...

Descripción completa

Detalles Bibliográficos
Autores principales: BAYRAKÇEKEN, Kemal, HONDUR, Ahmet Murad, ATALAY, Tuba, ARIBAS, Yavuz Kemal
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Scientific and Technological Research Council of Turkey 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7379410/
https://www.ncbi.nlm.nih.gov/pubmed/32041381
http://dx.doi.org/10.3906/sag-1910-11
_version_ 1783562633236971520
author BAYRAKÇEKEN, Kemal
HONDUR, Ahmet Murad
ATALAY, Tuba
ARIBAS, Yavuz Kemal
author_facet BAYRAKÇEKEN, Kemal
HONDUR, Ahmet Murad
ATALAY, Tuba
ARIBAS, Yavuz Kemal
author_sort BAYRAKÇEKEN, Kemal
collection PubMed
description BACKGROUND/AIM: To compare the clinical use, image quality and viewing angle of a commonly used contact wide angle viewing (WAV) system (Advanced Visual Instruments (AVI) Panoramic Imaging Systems, NY, USA) with a commonly used noncontact WAV system (Leica RUV800, Leica Microsystems, Switzerland). MATERIALS AND METHODS: Images of 42 consecutively operated eyes were obtained with both systems at the same surgical stages and were compared for image quality using the Imatest Master 4.5.13 (Imatest LLC, Boulder, USA) software. The viewing angles of the images were calculated using the optic disc sizes measured from the OCT and infrared fundus images. The 68-degree AVI lens was compared with the 90-dioptre (D) Leica RUV800 lens, while the 130-degree AVI lens was compared with the 132-D Leica RUV800 lens. The surgical assistants were asked to grade the difficulty of holding the lens in place from 1 to 10, 1 being the easiest and 10 being the most difficult. RESULTS: The contact system provided wider viewing angles with higher quality compared to the noncontact system both under fluid and air media. The difference was clinically significant in eyes with impaired corneal clarity, very high myopia, or small pupil. The difficulty of holding the lens in place ranged from 4 to 7, and decreased gradually with practice. CONCLUSIONS: Both WAV systems provided high image quality and adequate viewing angles in most cases. However, the contact system appeared to provide a superior image quality and/or a wider viewing angle in more challenging situations. The difficulty of holding the contact lens in place was found to be moderate.
format Online
Article
Text
id pubmed-7379410
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher The Scientific and Technological Research Council of Turkey
record_format MEDLINE/PubMed
spelling pubmed-73794102020-07-27 Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery BAYRAKÇEKEN, Kemal HONDUR, Ahmet Murad ATALAY, Tuba ARIBAS, Yavuz Kemal Turk J Med Sci Article BACKGROUND/AIM: To compare the clinical use, image quality and viewing angle of a commonly used contact wide angle viewing (WAV) system (Advanced Visual Instruments (AVI) Panoramic Imaging Systems, NY, USA) with a commonly used noncontact WAV system (Leica RUV800, Leica Microsystems, Switzerland). MATERIALS AND METHODS: Images of 42 consecutively operated eyes were obtained with both systems at the same surgical stages and were compared for image quality using the Imatest Master 4.5.13 (Imatest LLC, Boulder, USA) software. The viewing angles of the images were calculated using the optic disc sizes measured from the OCT and infrared fundus images. The 68-degree AVI lens was compared with the 90-dioptre (D) Leica RUV800 lens, while the 130-degree AVI lens was compared with the 132-D Leica RUV800 lens. The surgical assistants were asked to grade the difficulty of holding the lens in place from 1 to 10, 1 being the easiest and 10 being the most difficult. RESULTS: The contact system provided wider viewing angles with higher quality compared to the noncontact system both under fluid and air media. The difference was clinically significant in eyes with impaired corneal clarity, very high myopia, or small pupil. The difficulty of holding the lens in place ranged from 4 to 7, and decreased gradually with practice. CONCLUSIONS: Both WAV systems provided high image quality and adequate viewing angles in most cases. However, the contact system appeared to provide a superior image quality and/or a wider viewing angle in more challenging situations. The difficulty of holding the contact lens in place was found to be moderate. The Scientific and Technological Research Council of Turkey 2020-06-23 /pmc/articles/PMC7379410/ /pubmed/32041381 http://dx.doi.org/10.3906/sag-1910-11 Text en Copyright © 2020 The Author(s) This article is distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use and redistribution provided that the original author and source are credited.
spellingShingle Article
BAYRAKÇEKEN, Kemal
HONDUR, Ahmet Murad
ATALAY, Tuba
ARIBAS, Yavuz Kemal
Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
title Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
title_full Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
title_fullStr Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
title_full_unstemmed Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
title_short Real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
title_sort real-life comparison of the viewing angle and the image quality of two commonly used viewing systems for vitreoretinal surgery
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7379410/
https://www.ncbi.nlm.nih.gov/pubmed/32041381
http://dx.doi.org/10.3906/sag-1910-11
work_keys_str_mv AT bayrakcekenkemal reallifecomparisonoftheviewingangleandtheimagequalityoftwocommonlyusedviewingsystemsforvitreoretinalsurgery
AT hondurahmetmurad reallifecomparisonoftheviewingangleandtheimagequalityoftwocommonlyusedviewingsystemsforvitreoretinalsurgery
AT atalaytuba reallifecomparisonoftheviewingangleandtheimagequalityoftwocommonlyusedviewingsystemsforvitreoretinalsurgery
AT aribasyavuzkemal reallifecomparisonoftheviewingangleandtheimagequalityoftwocommonlyusedviewingsystemsforvitreoretinalsurgery