Falsely Elevated Thyroid-Stimulating Hormone Results due to Interference by M-Component of IgG-Lambda Type
Heterophilic antibodies but also M-components can interfere with laboratory tests causing erroneous results. We report the case of a 75-year-old man with myeloma and a monoclonal immunoglobulin component (M-component) that caused elevated thyroid-stimulating hormone (TSH) results. The M-component wa...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
S. Karger AG
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7383168/ https://www.ncbi.nlm.nih.gov/pubmed/32774256 http://dx.doi.org/10.1159/000507754 |
Sumario: | Heterophilic antibodies but also M-components can interfere with laboratory tests causing erroneous results. We report the case of a 75-year-old man with myeloma and a monoclonal immunoglobulin component (M-component) that caused elevated thyroid-stimulating hormone (TSH) results. The M-component was of the IgG-lambda type. Thyroid markers were analyzed repeatedly, and there was a clear association between IgG concentrations and TSH values (R(2) = 0.724). The highest TSH value was 75 mIU/L. Polyethylene glycol (PEG) precipitation of intact immunoglobulins was used to investigate if there was an antibody-related interference problem. The PEG treatment normalized the TSH value, showing that the cause of the elevated TSH result was due to interference caused by the M-component. In conclusion, it is important to remember that both heterophilic antibodies and M-components may cause erroneous results. |
---|