Cargando…

Falsely Elevated Thyroid-Stimulating Hormone Results due to Interference by M-Component of IgG-Lambda Type

Heterophilic antibodies but also M-components can interfere with laboratory tests causing erroneous results. We report the case of a 75-year-old man with myeloma and a monoclonal immunoglobulin component (M-component) that caused elevated thyroid-stimulating hormone (TSH) results. The M-component wa...

Descripción completa

Detalles Bibliográficos
Autores principales: Larsson, Anders, Gåfvels, Mats, Karlsson, Torbjörn
Formato: Online Artículo Texto
Lenguaje:English
Publicado: S. Karger AG 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7383168/
https://www.ncbi.nlm.nih.gov/pubmed/32774256
http://dx.doi.org/10.1159/000507754
Descripción
Sumario:Heterophilic antibodies but also M-components can interfere with laboratory tests causing erroneous results. We report the case of a 75-year-old man with myeloma and a monoclonal immunoglobulin component (M-component) that caused elevated thyroid-stimulating hormone (TSH) results. The M-component was of the IgG-lambda type. Thyroid markers were analyzed repeatedly, and there was a clear association between IgG concentrations and TSH values (R(2) = 0.724). The highest TSH value was 75 mIU/L. Polyethylene glycol (PEG) precipitation of intact immunoglobulins was used to investigate if there was an antibody-related interference problem. The PEG treatment normalized the TSH value, showing that the cause of the elevated TSH result was due to interference caused by the M-component. In conclusion, it is important to remember that both heterophilic antibodies and M-components may cause erroneous results.