Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T
This study aimed to validate the accuracy of the Omron HEM‐9601T, an automatic wrist‐type device for self‐blood pressure (BP) measurement with a timer function for automatic measurement of nocturnal BP, in the sitting position according to the American National Standards Institute/Association for th...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7383585/ https://www.ncbi.nlm.nih.gov/pubmed/32447831 http://dx.doi.org/10.1111/jch.13864 |
_version_ | 1783563448838258688 |
---|---|
author | Kuwabara, Mitsuo Harada, Kanako Hishiki, Yukiko Ohkubo, Takayoshi Kario, Kazuomi Imai, Yutaka |
author_facet | Kuwabara, Mitsuo Harada, Kanako Hishiki, Yukiko Ohkubo, Takayoshi Kario, Kazuomi Imai, Yutaka |
author_sort | Kuwabara, Mitsuo |
collection | PubMed |
description | This study aimed to validate the accuracy of the Omron HEM‐9601T, an automatic wrist‐type device for self‐blood pressure (BP) measurement with a timer function for automatic measurement of nocturnal BP, in the sitting position according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 guidelines, and to assess its performance in the supine position by applying the same protocol as conducted in the sitting position. The mean differences between the reference BPs and HEM‐9601T readings were 1.2 ± 6.9/1.1 ± 5.5 mmHg, 2.2 ± 6.5/1.8 ± 5.7 mmHg, 0.1 ± 6.6/1.5 ± 6.2 mmHg, and −0.8 ± 7.2/0.5 ± 6.4 mmHg for systolic BP/diastolic BP for criterion 1 in the sitting position, supine with sideways palm position, supine with upward palm position, and supine with downward palm position, respectively. In addition, the mean differences and their standard deviations for systolic BP and diastolic BP calculated according to criterion 2 in the ANSI/AAMI/ISO 81060‐2:2013 guidelines were acceptable in all four positions. In conclusion, the Omron HEM‐9601T fulfilled the validation criteria of the ANSI/AAMI/ISO81060‐2:2013 guidelines when used in the sitting position with the wrist at heart level, and its accuracy in the supine position was acceptable and roughly equivalent to that in the sitting position. The wrist‐type home BP monitor could be a more suitable tool for repeated nocturnal BP measurements at home than upper‐arm devices, and could improve the reliability of diagnosis and management of nocturnal hypertension. |
format | Online Article Text |
id | pubmed-7383585 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-73835852020-07-27 Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T Kuwabara, Mitsuo Harada, Kanako Hishiki, Yukiko Ohkubo, Takayoshi Kario, Kazuomi Imai, Yutaka J Clin Hypertens (Greenwich) Blood Pressure Measurement This study aimed to validate the accuracy of the Omron HEM‐9601T, an automatic wrist‐type device for self‐blood pressure (BP) measurement with a timer function for automatic measurement of nocturnal BP, in the sitting position according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 guidelines, and to assess its performance in the supine position by applying the same protocol as conducted in the sitting position. The mean differences between the reference BPs and HEM‐9601T readings were 1.2 ± 6.9/1.1 ± 5.5 mmHg, 2.2 ± 6.5/1.8 ± 5.7 mmHg, 0.1 ± 6.6/1.5 ± 6.2 mmHg, and −0.8 ± 7.2/0.5 ± 6.4 mmHg for systolic BP/diastolic BP for criterion 1 in the sitting position, supine with sideways palm position, supine with upward palm position, and supine with downward palm position, respectively. In addition, the mean differences and their standard deviations for systolic BP and diastolic BP calculated according to criterion 2 in the ANSI/AAMI/ISO 81060‐2:2013 guidelines were acceptable in all four positions. In conclusion, the Omron HEM‐9601T fulfilled the validation criteria of the ANSI/AAMI/ISO81060‐2:2013 guidelines when used in the sitting position with the wrist at heart level, and its accuracy in the supine position was acceptable and roughly equivalent to that in the sitting position. The wrist‐type home BP monitor could be a more suitable tool for repeated nocturnal BP measurements at home than upper‐arm devices, and could improve the reliability of diagnosis and management of nocturnal hypertension. John Wiley and Sons Inc. 2020-05-24 /pmc/articles/PMC7383585/ /pubmed/32447831 http://dx.doi.org/10.1111/jch.13864 Text en © 2020 The Authors. The Journal of Clinical Hypertension published by Wiley Periodicals LLC https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes. |
spellingShingle | Blood Pressure Measurement Kuwabara, Mitsuo Harada, Kanako Hishiki, Yukiko Ohkubo, Takayoshi Kario, Kazuomi Imai, Yutaka Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T |
title | Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T |
title_full | Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T |
title_fullStr | Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T |
title_full_unstemmed | Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T |
title_short | Validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060‐2:2013 guidelines: Omron HEM‐9601T |
title_sort | validation of a wrist‐type home nocturnal blood pressure monitor in the sitting and supine position according to the ansi/aami/iso81060‐2:2013 guidelines: omron hem‐9601t |
topic | Blood Pressure Measurement |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7383585/ https://www.ncbi.nlm.nih.gov/pubmed/32447831 http://dx.doi.org/10.1111/jch.13864 |
work_keys_str_mv | AT kuwabaramitsuo validationofawristtypehomenocturnalbloodpressuremonitorinthesittingandsupinepositionaccordingtotheansiaamiiso8106022013guidelinesomronhem9601t AT haradakanako validationofawristtypehomenocturnalbloodpressuremonitorinthesittingandsupinepositionaccordingtotheansiaamiiso8106022013guidelinesomronhem9601t AT hishikiyukiko validationofawristtypehomenocturnalbloodpressuremonitorinthesittingandsupinepositionaccordingtotheansiaamiiso8106022013guidelinesomronhem9601t AT ohkubotakayoshi validationofawristtypehomenocturnalbloodpressuremonitorinthesittingandsupinepositionaccordingtotheansiaamiiso8106022013guidelinesomronhem9601t AT kariokazuomi validationofawristtypehomenocturnalbloodpressuremonitorinthesittingandsupinepositionaccordingtotheansiaamiiso8106022013guidelinesomronhem9601t AT imaiyutaka validationofawristtypehomenocturnalbloodpressuremonitorinthesittingandsupinepositionaccordingtotheansiaamiiso8106022013guidelinesomronhem9601t |