A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease

INTRODUCTION: Hyperkalaemia is a common electrolyte disorder that may cause life-threatening cardiac arrythmias. We aimed to determine the agreement of potassium concentrations between GEM premier 3500 point-of-care blood gas analyser (POC-BGA) and Roche Cobas 6000 c501 auto-analyser in patients wit...

Descripción completa

Detalles Bibliográficos
Autores principales: Chothia, Mogamat-Yazied, Kassum, Patricia, Zemlin, Annalise
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7394258/
https://www.ncbi.nlm.nih.gov/pubmed/32774124
http://dx.doi.org/10.11613/BM.2020.030702
_version_ 1783565200718299136
author Chothia, Mogamat-Yazied
Kassum, Patricia
Zemlin, Annalise
author_facet Chothia, Mogamat-Yazied
Kassum, Patricia
Zemlin, Annalise
author_sort Chothia, Mogamat-Yazied
collection PubMed
description INTRODUCTION: Hyperkalaemia is a common electrolyte disorder that may cause life-threatening cardiac arrythmias. We aimed to determine the agreement of potassium concentrations between GEM premier 3500 point-of-care blood gas analyser (POC-BGA) and Roche Cobas 6000 c501 auto-analyser in patients with hyperkalaemia. METHODS: A prospective, cross-sectional study of all consecutive adult patients referred to the Renal Unit with a serum potassium concentration ≥ 5.5 mmol/L was performed. A total of 59 paired venous blood samples were included in the final statistical analysis. Passing-Bablok regression and Bland Altman analysis were used to compare the two methods. RESULTS: The median laboratory auto-analyser potassium concentration was 6.1 (5.9-7.1) mmol/L as compared to the POC-BGA potassium concentration of 5.7 (5.5-6.8) mmol/L with a mean difference of - 0.43 mmol/L and 95% upper and lower limits of agreement of 0.35 mmol/L and - 1.21 mmol/L, respectively. Regression analysis revealed proportional systematic error. Test for linearity did not indicate significant deviation (P = 0.297). CONCLUSION: Although regression analysis indicated proportional systematic error, on Bland Altman analysis, the mean difference appeared to remain relatively constant across the potassium range that was evaluated. Therefore, in patients presenting to the emergency department with a clinical suspicion of hyperkalaemia, POC-BGA potassium concentrations may be considered a surrogate for laboratory auto-analyser measurements once clinicians have been cautioned about this difference.
format Online
Article
Text
id pubmed-7394258
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-73942582020-08-08 A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease Chothia, Mogamat-Yazied Kassum, Patricia Zemlin, Annalise Biochem Med (Zagreb) Original Articles INTRODUCTION: Hyperkalaemia is a common electrolyte disorder that may cause life-threatening cardiac arrythmias. We aimed to determine the agreement of potassium concentrations between GEM premier 3500 point-of-care blood gas analyser (POC-BGA) and Roche Cobas 6000 c501 auto-analyser in patients with hyperkalaemia. METHODS: A prospective, cross-sectional study of all consecutive adult patients referred to the Renal Unit with a serum potassium concentration ≥ 5.5 mmol/L was performed. A total of 59 paired venous blood samples were included in the final statistical analysis. Passing-Bablok regression and Bland Altman analysis were used to compare the two methods. RESULTS: The median laboratory auto-analyser potassium concentration was 6.1 (5.9-7.1) mmol/L as compared to the POC-BGA potassium concentration of 5.7 (5.5-6.8) mmol/L with a mean difference of - 0.43 mmol/L and 95% upper and lower limits of agreement of 0.35 mmol/L and - 1.21 mmol/L, respectively. Regression analysis revealed proportional systematic error. Test for linearity did not indicate significant deviation (P = 0.297). CONCLUSION: Although regression analysis indicated proportional systematic error, on Bland Altman analysis, the mean difference appeared to remain relatively constant across the potassium range that was evaluated. Therefore, in patients presenting to the emergency department with a clinical suspicion of hyperkalaemia, POC-BGA potassium concentrations may be considered a surrogate for laboratory auto-analyser measurements once clinicians have been cautioned about this difference. Croatian Society of Medical Biochemistry and Laboratory Medicine 2020-08-05 2020-10-15 /pmc/articles/PMC7394258/ /pubmed/32774124 http://dx.doi.org/10.11613/BM.2020.030702 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Chothia, Mogamat-Yazied
Kassum, Patricia
Zemlin, Annalise
A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
title A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
title_full A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
title_fullStr A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
title_full_unstemmed A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
title_short A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
title_sort method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7394258/
https://www.ncbi.nlm.nih.gov/pubmed/32774124
http://dx.doi.org/10.11613/BM.2020.030702
work_keys_str_mv AT chothiamogamatyazied amethodcomparisonstudyofapointofcarebloodgasanalyserwithalaboratoryautoanalyserforthedeterminationofpotassiumconcentrationsduringhyperkalaemiainpatientswithkidneydisease
AT kassumpatricia amethodcomparisonstudyofapointofcarebloodgasanalyserwithalaboratoryautoanalyserforthedeterminationofpotassiumconcentrationsduringhyperkalaemiainpatientswithkidneydisease
AT zemlinannalise amethodcomparisonstudyofapointofcarebloodgasanalyserwithalaboratoryautoanalyserforthedeterminationofpotassiumconcentrationsduringhyperkalaemiainpatientswithkidneydisease
AT chothiamogamatyazied methodcomparisonstudyofapointofcarebloodgasanalyserwithalaboratoryautoanalyserforthedeterminationofpotassiumconcentrationsduringhyperkalaemiainpatientswithkidneydisease
AT kassumpatricia methodcomparisonstudyofapointofcarebloodgasanalyserwithalaboratoryautoanalyserforthedeterminationofpotassiumconcentrationsduringhyperkalaemiainpatientswithkidneydisease
AT zemlinannalise methodcomparisonstudyofapointofcarebloodgasanalyserwithalaboratoryautoanalyserforthedeterminationofpotassiumconcentrationsduringhyperkalaemiainpatientswithkidneydisease