PROCEDURES TO MEASURE MEAN AMBIENT DOSE EQUIVALENT RATES USING ELECTRET ION CHAMBERS

The capabilities of electret ion chambers (EICs) to measure mean ambient dose equivalent rates were investigated by performing both laboratory and field studies of their properties. First, EICs were ‘calibrated’ to measure ambient gamma dose equivalent in the Ionizing Calibration Laboratory of the G...

Descripción completa

Detalles Bibliográficos
Autores principales: Leontaris, F, Boziari, A, Clouvas, A, Kolovou, M, Guilhot, J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7397567/
https://www.ncbi.nlm.nih.gov/pubmed/32488254
http://dx.doi.org/10.1093/rpd/ncaa061
Descripción
Sumario:The capabilities of electret ion chambers (EICs) to measure mean ambient dose equivalent rates were investigated by performing both laboratory and field studies of their properties. First, EICs were ‘calibrated’ to measure ambient gamma dose equivalent in the Ionizing Calibration Laboratory of the Greek Atomic Energy Commission. The EICs were irradiated with different gamma photon energies and from different angles. Calibration factors were deduced (electret’s voltage drop due to irradiation in terms of ambient dose equivalent). In the field studies, EICs were installed at eight locations belonging to the Greek Early Warning System Network (which is based on Reuter-Stokes ionization chambers) for three periods, averaging 5 months each. In the same locations, in situ gamma spectrometry measurements were performed with portable germanium detectors. Gamma ambient dose equivalent rates were deduced by the in situ gamma spectrometry measurements and by soil sample analysis. The mean daily electret potential drop (in Volts) was compared with the mean daily ambient dose equivalent, measured with a portable HPGe detector and Reuter-Stokes high-pressure ionization chambers. From these measurements, ‘field’ calibration factors (electret’s voltage drop due to gamma radiation in terms of ambient dose equivalent) were deduced and found in very good agreement with the values deduced in Laboratory. The influence of cosmic radiation and the intrinsic voltage loss when performing long-term environmental gamma measurements with EICs, was estimated.