Cargando…
Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory
BACKGROUND: The ExcelsiusGPS® (Globus Medical, Inc., Audubon, PA) is a next-generation spine surgery robotic system recently approved for use in the United States. The objective of the current study is to assess pedicle screw accuracy and clinical outcomes among two of the first operative cases util...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7398380/ https://www.ncbi.nlm.nih.gov/pubmed/32922884 http://dx.doi.org/10.1186/s41016-018-0131-x |
_version_ | 1783565954606694400 |
---|---|
author | Jiang, Bowen Karim Ahmed, A. Zygourakis, Corinna C. Kalb, Samuel Zhu, Alex M. Godzik, Jakub Molina, Camilo A. Blitz, Ari M. Bydon, Ali Crawford, Neil Theodore, Nicholas |
author_facet | Jiang, Bowen Karim Ahmed, A. Zygourakis, Corinna C. Kalb, Samuel Zhu, Alex M. Godzik, Jakub Molina, Camilo A. Blitz, Ari M. Bydon, Ali Crawford, Neil Theodore, Nicholas |
author_sort | Jiang, Bowen |
collection | PubMed |
description | BACKGROUND: The ExcelsiusGPS® (Globus Medical, Inc., Audubon, PA) is a next-generation spine surgery robotic system recently approved for use in the United States. The objective of the current study is to assess pedicle screw accuracy and clinical outcomes among two of the first operative cases utilizing the ExcelsiusGPS® robotic system and describe a novel metric to quantify screw deviation. METHODS: Two patients who underwent lumbar fusion at a single institution with the ExcelsiusGPS® surgical robot were included. Pre-operative trajectory planning was performed from an intra-operative CT scan using the O-arm (Medtronic, Inc., Minneapolis, MN). After robotic-assisted screw implantation, a post-operative CT scan was obtained to confirm ideal screw placement and accuracy with the planned trajectory. A novel pedicle screw accuracy algorithm was devised to measure screw tip/tail deviation distance and angular offset on axial and sagittal planes. Screw accuracy was concurrently determined by a blinded neuroradiologist using the traditional Gertzbein-Robbins method. Clinical variables such as symptomatology, operative data, and post-operative follow-up were also collected. RESULTS: Eight pedicle screws were placed in two L4-L5 fusion cases. Mean screw tip deviation was 2.1 mm (range 0.8–5.2 mm), mean tail deviation was 3.2 mm (range 0.9–5.4 mm), and mean angular offset was 2.4 degrees (range 0.7–3.8 degrees). All eight screws were accurately placed based on the Gertzbein-Robbins scale (88% Grade A and 12% Grade B). There were no cases of screw revision or new post-operative deficit. Both patients experienced improvement in Frankel grade and Karnofsky Performance Status (KPS) score by 6 weeks post-op. CONCLUSION: The ExcelsiusGPS® robot allows for precise execution of an intended pre-planned trajectory and accurate screw placement in the first patients to undergo robotic-assisted fusion with this technology. |
format | Online Article Text |
id | pubmed-7398380 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-73983802020-09-10 Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory Jiang, Bowen Karim Ahmed, A. Zygourakis, Corinna C. Kalb, Samuel Zhu, Alex M. Godzik, Jakub Molina, Camilo A. Blitz, Ari M. Bydon, Ali Crawford, Neil Theodore, Nicholas Chin Neurosurg J Research BACKGROUND: The ExcelsiusGPS® (Globus Medical, Inc., Audubon, PA) is a next-generation spine surgery robotic system recently approved for use in the United States. The objective of the current study is to assess pedicle screw accuracy and clinical outcomes among two of the first operative cases utilizing the ExcelsiusGPS® robotic system and describe a novel metric to quantify screw deviation. METHODS: Two patients who underwent lumbar fusion at a single institution with the ExcelsiusGPS® surgical robot were included. Pre-operative trajectory planning was performed from an intra-operative CT scan using the O-arm (Medtronic, Inc., Minneapolis, MN). After robotic-assisted screw implantation, a post-operative CT scan was obtained to confirm ideal screw placement and accuracy with the planned trajectory. A novel pedicle screw accuracy algorithm was devised to measure screw tip/tail deviation distance and angular offset on axial and sagittal planes. Screw accuracy was concurrently determined by a blinded neuroradiologist using the traditional Gertzbein-Robbins method. Clinical variables such as symptomatology, operative data, and post-operative follow-up were also collected. RESULTS: Eight pedicle screws were placed in two L4-L5 fusion cases. Mean screw tip deviation was 2.1 mm (range 0.8–5.2 mm), mean tail deviation was 3.2 mm (range 0.9–5.4 mm), and mean angular offset was 2.4 degrees (range 0.7–3.8 degrees). All eight screws were accurately placed based on the Gertzbein-Robbins scale (88% Grade A and 12% Grade B). There were no cases of screw revision or new post-operative deficit. Both patients experienced improvement in Frankel grade and Karnofsky Performance Status (KPS) score by 6 weeks post-op. CONCLUSION: The ExcelsiusGPS® robot allows for precise execution of an intended pre-planned trajectory and accurate screw placement in the first patients to undergo robotic-assisted fusion with this technology. BioMed Central 2018-09-03 /pmc/articles/PMC7398380/ /pubmed/32922884 http://dx.doi.org/10.1186/s41016-018-0131-x Text en © The Author(s) 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Jiang, Bowen Karim Ahmed, A. Zygourakis, Corinna C. Kalb, Samuel Zhu, Alex M. Godzik, Jakub Molina, Camilo A. Blitz, Ari M. Bydon, Ali Crawford, Neil Theodore, Nicholas Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
title | Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
title_full | Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
title_fullStr | Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
title_full_unstemmed | Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
title_short | Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
title_sort | pedicle screw accuracy assessment in excelsiusgps® robotic spine surgery: evaluation of deviation from pre-planned trajectory |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7398380/ https://www.ncbi.nlm.nih.gov/pubmed/32922884 http://dx.doi.org/10.1186/s41016-018-0131-x |
work_keys_str_mv | AT jiangbowen pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT karimahmeda pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT zygourakiscorinnac pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT kalbsamuel pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT zhualexm pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT godzikjakub pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT molinacamiloa pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT blitzarim pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT bydonali pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT crawfordneil pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory AT theodorenicholas pediclescrewaccuracyassessmentinexcelsiusgpsroboticspinesurgeryevaluationofdeviationfrompreplannedtrajectory |