The Impact of Deterministic Signal Loss on OCT Angiography Measurements
PURPOSE: In this study, we aimed at investigating the impact of deterministic signal loss on image quality and, thus, on optical coherence tomography angiography (OCTA) measurements performed by the RTVue-XR Avanti System. METHODS: Absorptive filters with different optical densities (ODs) were used...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7401964/ https://www.ncbi.nlm.nih.gov/pubmed/32821482 http://dx.doi.org/10.1167/tvst.9.5.10 |
Sumario: | PURPOSE: In this study, we aimed at investigating the impact of deterministic signal loss on image quality and, thus, on optical coherence tomography angiography (OCTA) measurements performed by the RTVue-XR Avanti System. METHODS: Absorptive filters with different optical densities (ODs) were used to simulate signal loss in a controlled way in 30 eyes from 15 healthy subjects. Scan Quality (SQ), provided by the AngioVue software, was applied as a measure of image quality. RESULTS: Assessing the effect of decreased light transmittance on SQ values as well as that of attenuated image quality on OCTA parameters revealed a strong systematic dependence between these quantities. Attenuated image quality was associated with significantly decreased macular and peripapillary vessel density (VD) values, and we calculated a correction factor of 2.27% to 3.97% for a one-unit change in SQ for the different VD parameters. CONCLUSIONS: Our findings suggest that the influence of systematic changes in image quality on OCTA parameters needs to be considered during patient follow-up in order to make valid assessment of progression. TRANSLATIONAL RELEVANCE: For accurate evaluation of longitudinal changes in OCTA parameters, equal scan quality or using a correction factor is suggested. |
---|